Laboratory developed test (LDT): definition, characteristics and regulation
Short answer: A laboratory developed test (LDT) is a type of in vitro diagnostic (IVD) test that is designed, manufactured and performed within a single CLIA-certified laboratory. Often called in-house or “home brew” tests, LDTs are built by hospital, academic and commercial reference labs when no commercial test meets a clinical need. CLIA treats every LDT as high complexity.
Since the March 31, 2025 court ruling, FDA's 2024 LDT rule no longer applies. CLIA still requires the lab to validate each LDT before reporting patient results.
Last reviewed: November 19, 2026. Primary sources: FDA LDT FAQs, CMS LDT and CLIA FAQs, 42 CFR 493.1253, CRS: FDA Regulation of LDTs (IF11389) and AHA.

Definition: what a laboratory developed test is
FDA's LDT FAQs define LDTs as IVD products “intended for clinical use and designed, manufactured, and used within a single clinical laboratory that is certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA) and meets the regulatory requirements under CLIA to perform high complexity testing” (FDA). CMS uses the same definition and adds that LDTs “are also sometimes called in-house developed tests, or ‘home brew’ tests” (CMS LDT and CLIA FAQs). The Congressional Research Service describes LDTs as “a class of in vitro diagnostic (IVD) device that is designed, manufactured, and used within a single laboratory” (CRS IF11389).
Three conditions make a test an LDT: the lab designs it, the lab makes it (reagents, protocol, software, cutoffs) and the test is used only in that lab. A kit sold to other labs is a commercial IVD. For the full comparison with RUO, IUO and ASR products, see LDT vs IVD vs RUO.
Key characteristics of LDTs
In-house creation (“home brews”)
The test is created and run by one laboratory under one CLIA certificate. FDA's FAQ notes that a test designed, made or used across more than one laboratory is not an LDT, even within the same organization (FDA).
Who makes them
Hospital and health-system laboratories, academic medical centers, commercial reference laboratories and public health laboratories. AHA says hospital LDTs range from routine blood counts to molecular and genetic tests for cancer, heart disease and infectious or rare diseases (AHA).
Common uses
- Rare diseases: tests for small patient populations that no manufacturer serves. CSPI notes LDTs were historically used for uncommon tests serving small populations (CSPI).
- Specialized genomic and cancer profiling: CRS cites genomic cancer testing as a typical LDT area (CRS). See NGS and oncology LDTs.
- Toxicology: LC-MS/MS drug panels, a method CMS singled out in its 2026 CLIA request for information (91 FR 43586). See toxicology LDTs.
- Emerging public health threats: new strains of infectious disease (CRS). See infectious disease and public health LDTs.
Flexibility when no commercial IVD exists
Labs build LDTs when no FDA-cleared or approved test exists, when a cleared test does not cover the needed specimen type or population, or when science moves faster than commercial kits. CRS notes the COVID-19 pandemic showed “the pivotal role LDTs play in facilitating rapid access to tests” (CRS).
High complexity under CLIA
CMS states “LDTs are considered high complexity tests,” so the lab must meet every CLIA requirement for high complexity testing (CMS). See CLIA requirements for LDTs.
How LDTs differ from commercial IVD test kits
| Feature | Laboratory developed test (LDT) | Commercial IVD test kit |
|---|---|---|
| Who designs and makes it | The clinical laboratory that runs it | A device manufacturer |
| Where it is used | Only inside the single laboratory (one CLIA certificate) that developed it (FDA) | Sold or distributed to many laboratories |
| FDA premarket review | Generally none. FDA's 2024 LDT rule was vacated on March 31, 2025, and FDA reverted its regulation on September 19, 2025 | 510(k) clearance, De Novo authorization or premarket approval before marketing |
| What CLIA requires before patient testing | The lab establishes accuracy, precision, analytical sensitivity, analytical specificity, reportable range and reference intervals (493.1253(b)(2)) | The lab verifies the manufacturer's accuracy, precision, reportable range and reference intervals (493.1253(b)(1)) |
| CLIA complexity | High complexity by default (CMS) | Categorized by FDA as waived, moderate or high complexity |
| Clinical validity | Not addressed by CLIA (CMS); CAP and New York State review clinical claims | Reviewed by FDA as part of safety and effectiveness |
| Labeling and instructions | The lab's own written procedure (493.1251) and report statements | FDA-regulated labeling and package insert (21 CFR 809.10) |
| Changes | The lab controls changes and revalidates; a lab change to a kit turns it into an LDT-type test | Manufacturer controls changes; labs that alter the kit take on validation |
| Flexibility and speed | Can be built or updated quickly for new pathogens, rare diseases or new biomarkers (CRS) | Updates follow the manufacturer's regulatory timeline |
| Typical uses | Rare diseases, genomic and cancer profiling, toxicology, emerging infectious threats | High-volume routine tests across many labs |
| Reimbursement | CPT, PLA or unlisted codes; molecular tests in MolDX states need a DEX Z-Code and technical assessment | Established codes; FDA status can support coverage, for example under NCD 90.2 |
The biggest practical difference is who proves the test works. A manufacturer proves a kit's performance to FDA before sale, and each lab then verifies it. For an LDT, the lab proves performance itself, in its own setting, and CMS notes those findings “are not meaningful outside of the laboratory that did the analysis” (CMS). More detail: LDT vs IVD vs RUO and verification vs validation.
How LDTs are regulated in 2026
CLIA (CMS): the laboratory
CLIA regulates the laboratory, not the test as a product. Before reporting results the lab must establish performance specifications under 42 CFR 493.1253(b)(2), set calibration and QC from that data, and document it all. Surveyors review LDT validation during the routine biennial survey (CMS). Details: LDT validation requirements.
CAP (if accredited)
CAP adds checklist items, including validation of clinical claims, director approval before clinical use, a current list of LDTs and a report statement. See CAP LDT requirements.
New York State (NYS CLEP)
Labs testing New York specimens need a New York permit, and New York reviews LDTs through its own test approval process (Wadsworth Center). See NYS CLEP LDT approval.
FDA: where the rule stands
FDA issued a final rule on May 6, 2024 to phase LDTs into device requirements. On March 31, 2025, a federal district court vacated it, and on September 19, 2025 FDA reverted 21 CFR 809.3(a) to its earlier text (FDA; 90 FR 45134). CRS says the ruling, which FDA did not appeal, “has the practical effect of removing most if not all LDTs from FDA's regulatory purview” (CRS). Full history: FDA LDT rule status.
| Date | Event | Source |
|---|---|---|
| October 2014 | FDA issues draft guidance proposing a risk-based framework for LDT oversight; finalization is later delayed | CRS IF11389 |
| January 2017 | FDA publishes a discussion paper instead of final guidance | CRS IF11389 |
| August 2020 / November 2021 | HHS says FDA cannot require LDT premarket review without rulemaking; that policy is rescinded in November 2021 | CRS IF11389 |
| May 6, 2024 | FDA final rule adds “including when the manufacturer of these products is a laboratory” to 21 CFR 809.3(a) and announces a roughly four-year phaseout of enforcement discretion | 89 FR 37286 |
| March 31, 2025 | U.S. District Court for the Eastern District of Texas vacates the rule in the consolidated ACLA v. FDA and AMP v. FDA cases, holding LDTs are not “devices” under the FD&C Act | Memorandum opinion, No. 4:24-cv-479 (E.D. Tex.); 90 FR 45134 (background); CRS |
| May 30, 2025 | Deadline to appeal the vacatur passes with no appeal filed; the vacatur stands | CRS IF11389 |
| September 19, 2025 | FDA final rule reverts 21 CFR 809.3(a) to its pre-2024 text, effective the same day | 90 FR 45134; FDA LDT page |
| December 19, 2025 | CRS updates its In Focus report: the ruling removes “most if not all LDTs” from FDA's regulatory purview | CRS IF11389 |
| May 19, 2026 | Enhancing CLIA Act of 2026 (H.R. 8890) introduced in the House; it would reaffirm that LDTs are regulated by CMS under CLIA and create a public LDT database. Latest action: referred to committee (May 19, 2026). A bill, not law | Congress.gov H.R. 8890 |
| July 16, 2026 | CMS and CDC publish a CLIA request for information (CMS-3485-NC) asking how labs establish performance specifications for non-FDA-cleared tests, naming toxicology and NGS; comments closed September 14, 2026 | 91 FR 43586 |
Building, validating or fixing an LDT? We help CLIA labs plan and document LDT development and validation and prepare for CLIA, CAP and New York State review.
Examples of laboratory developed tests by specialty
| Specialty | Typical LDTs | Guide |
|---|---|---|
| Oncology and genomics | NGS tumor panels, hematologic malignancy panels, comprehensive genomic profiling | NGS and oncology LDTs |
| Rare and inherited disease | Single-gene and exome sequencing, targeted variant tests for small patient groups | LDT examples |
| Toxicology | LC-MS/MS drug confirmation and quantitation panels | Toxicology LDTs |
| Infectious disease and public health | PCR for emerging pathogens, multiplex syndromic panels, resistance markers, outbreak sequencing | Infectious disease LDTs |
| Molecular (PCR) | In-house primer and probe assays, ASR-based assays | Molecular and PCR LDTs |
| Chemistry and endocrinology | Mass spectrometry hormone and vitamin assays; FDA-cleared assays run on unapproved specimen types | Modified FDA-cleared tests |
| Pathology and immunology | Immunohistochemistry stains, flow cytometry panels, assays assembled from components | LDT examples |
Named examples and more categories: laboratory developed test examples.
How to develop and validate an LDT
- Define the intended use: analyte, specimen types, patient population, clinical purpose and every claim you plan to make.
- Design and optimize the assay (reagents, controls, calibrators, cutoffs, software) under design and change control. See LDT design controls.
- Write a validation plan with acceptance criteria set before testing. Template: plan and report template.
- Run the analytical studies in 493.1253(b)(2): accuracy and precision, analytical sensitivity, specificity and interference, reportable range and reference intervals, and other characteristics such as specimen stability.
- Support clinical claims with literature or studies (clinical validity).
- Set QC and calibration from the data, write the procedure, train staff and get director sign-off before the first patient report.
- Monitor after launch: QC, PT or twice-yearly alternative assessment, and revalidation after changes.
The full step-by-step process is in how to develop a laboratory developed test.
How LDTs are billed and reimbursed
LDTs are billed with CPT codes when one describes the test, with an AMA Proprietary Laboratory Analyses (PLA) code for a single lab's test (AMA), or with an unlisted code. Medicare pays most lab tests under the Clinical Laboratory Fee Schedule, and a single-lab test can qualify as an Advanced Diagnostic Laboratory Test (ADLT) if it meets CMS criteria (CMS). In MolDX jurisdictions, molecular LDTs need a DEX Z-Code and a completed technical assessment before Medicare pays (LCD L35025). See LDT reimbursement, DEX Z-Code registration and MolDX technical assessment.
How Sound Lab Consultants helps
- Decide whether your test is an LDT, a modified FDA-cleared test or a verification of an unmodified kit.
- Plan the development lifecycle and write the validation plan and acceptance criteria.
- Review validation data and the director-signed report before go-live.
- Prepare for CLIA surveys, CAP inspections and NYS CLEP submissions.
- Map reimbursement steps: coding, DEX registration and MolDX evidence.
- Start with the LDT consulting hub.
Talk through your test with an LDT regulatory consultant. Bring your intended use, platform and target launch date.
Frequently asked questions
What is a laboratory developed test (LDT)?
An LDT is an in vitro diagnostic test that is designed, manufactured and used within a single CLIA-certified laboratory that meets CLIA requirements for high complexity testing. LDTs are also called in-house or home brew tests.
What does LDT stand for in a lab?
LDT stands for laboratory developed test (also written lab-developed test). It means a test the laboratory built and runs itself rather than a commercial kit bought from a manufacturer.
Who makes laboratory developed tests?
Hospital and health-system laboratories, academic medical centers, commercial reference laboratories and public health laboratories. Each LDT must stay within the one laboratory that developed it.
What are LDTs commonly used for?
Rare diseases, specialized genomic and cancer profiling, toxicology, emerging infectious diseases and other needs where no FDA-cleared or approved test exists or an existing kit does not fit the specimen or population.
How is an LDT different from a commercial IVD test kit?
A commercial IVD kit is made by a manufacturer, reviewed by FDA before sale and verified by each lab that uses it. An LDT is made and used in one lab, which must establish all of its performance specifications under 42 CFR 493.1253(b)(2).
Are LDTs regulated by the FDA in 2026?
FDA's May 6, 2024 LDT rule was vacated by a federal court on March 31, 2025, and FDA reverted its regulation on September 19, 2025. Most LDTs are now outside FDA's regulatory purview, according to CRS. CLIA, accreditor and state rules still apply.
Are LDTs high complexity under CLIA?
Yes. CMS states that LDTs are considered high complexity tests, so the laboratory must meet all CLIA requirements for high complexity testing, including personnel, quality control and proficiency testing or alternative assessment.
Does an LDT need FDA approval?
Not under current federal rules for most LDTs. The lab must still validate the test under CLIA before reporting results, and New York requires its own approval for LDTs used on New York specimens.
Is a modified FDA-cleared test an LDT?
CLIA treats it the same way. When a lab modifies an FDA-cleared or approved test, it must establish performance specifications under 42 CFR 493.1253(b)(2), just as for an in-house test.
Can an LDT be billed to Medicare?
Yes, if it is a covered service. It is billed with a CPT, PLA or unlisted code. In MolDX states, molecular LDTs also need a DEX Z-Code and a completed technical assessment.
Sources
- FDA: Definitions and general oversight, LDT FAQs
- FDA: Laboratory developed tests
- CMS: LDTs and CLIA FAQs (PDF)
- Congress.gov CRS In Focus IF11389: FDA Regulation of LDTs (Dec. 19, 2025)
- American Hospital Association: Laboratory developed tests
- Center for Science in the Public Interest: Laboratory-developed tests
- 42 CFR 493.1253
- 89 FR 37286 (May 6, 2024)
- 90 FR 45134 (Sept. 19, 2025)
- 91 FR 43586 CLIA RFI (July 16, 2026)
- NYS Wadsworth Center CLEP test approval
- CAP test method validation and verification excerpt (PDF)
- NIH SEED regulatory knowledge guide for LDTs (PDF)
- AMA CPT PLA codes
- LCD L35025 MolDX
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.