CAP Checklist Requirements for Laboratory Developed Tests (LDTs)
Short answer: A CAP-accredited lab running a laboratory developed test (LDT) must meet CLIA's 42 CFR 493.1253(b)(2) validation plus CAP checklist items: a validation study (COM.40350), validation of clinical claims (COM.40640), written assessment and director approval (COM.40475), a current LDT list (COM.40200), an LDT report statement (COM.40630) and alternative assessment when no PT exists.
Last reviewed: November 18, 2026. Primary sources: CAP test method validation and verification excerpt, CAP validation and verification decision tree, 42 CFR 493.1253 and CAP Laboratory Accreditation Program.

CAP vs CLIA for LDTs at a glance
| Topic | Federal CLIA | CAP checklist item |
|---|---|---|
| Performance specifications | Establish accuracy, precision, analytical sensitivity, specificity with interferences, reportable range and reference intervals (493.1253(b)(2)) | COM.40350 validation study; COM.40400 sensitivity, COM.40450 specificity, COM.40500 interferences, COM.40600 reportable range |
| Clinical validity | Not addressed by CLIA (CMS) | COM.40640: validate any clinical claim |
| Sign-off | Document all 493.1253 work (493.1253(c)) | COM.40475: written assessment and director approval before clinical use |
| Calibration and QC | From the specifications (493.1253(b)(3)) | COM.40610: define calibration and QC frequency, number and levels |
| No PT available | Verify accuracy twice a year (493.1236(c)) | COM.01500 alternative performance assessment at least semiannually; COM.01700 review |
| Test inventory | No specific list | COM.40200: list of LDTs and modified FDA tests, including ASR-based tests |
| Report wording | Required for ASR-based tests (21 CFR 809.30(e)) | COM.40630: LDT statement on patient reports |
| Multiple instruments | Twice-yearly comparison (493.1281) | COM.04250 instrument/method comparability |
| Records | While in use and at least 2 years (493.1105) | While in use plus 2 years after retirement |
CAP renumbers and rewords items between checklist editions. The IDs here come from CAP's published validation excerpt and inspection decision tree. Confirm each against the edition you are inspected to.
What CAP counts as an LDT
CAP treats a test as an LDT when the laboratory performing it developed it, in whole or in part, and it is not FDA-cleared or approved. A commercial kit your lab changed falls in the same group as a fully in-house assay (CAP; modified FDA-cleared tests).
Validation study (COM.40350) and sample numbers
- Cover every characteristic that applies: accuracy, precision, analytical sensitivity, analytical specificity and interferences, reportable range, and others such as stability, carryover and cross-contamination.
- Quantitative tests: CAP's guidance calls for at least 20 samples across the measuring range.
- Qualitative tests: at least 20 samples including positives and negatives, with low positives near the detection limit.
- Fewer samples are acceptable when the director documents the reason and the acceptance criteria.
- Run the study on the instruments that will report results. Study design: LDT validation and accuracy and precision.
Clinical claims (COM.40640)
If your lab claims a result means something clinically, such as a marker predicting a diagnosis or treatment response, CAP requires support from peer-reviewed literature or a clinical study (generally at least 20 positive and negative samples unless the director documents why fewer suffice) (CAP). ADLM notes CAP and The Joint Commission require labs to document clinical validity even though CLIA does not (ADLM). See clinical validity for LDTs.
Building, validating or fixing an LDT? We help CLIA labs plan and document LDT development and validation and prepare for CLIA, CAP and New York State review.
Director approval (COM.40475)
- A written assessment of the study data against each acceptance criterion.
- Investigation and outcome of any discordant results.
- A signed statement approving the test for clinical use before the first patient report.
- Dates proving validation came before clinical use. CAP's decision tree has inspectors check this for each test added since the last inspection (CAP decision tree).
LDT list, report statement and clients
- LDT list (COM.40200): keep a current list of every LDT and modified FDA test, including ASR-based tests.
- Report statement (COM.40630): patient reports for LDTs carry a statement that the test was developed and its performance characteristics determined by the laboratory. Wording options: LDT disclaimer statement.
- Clients: be ready to share performance information with ordering clinicians on request.
What the CAP inspector reviews
- The LDT list against the test menu.
- For each new LDT: plan, data, written assessment and director approval dated before go-live.
- Clinical claims files for tests that make claims.
- Calibration and QC definitions per test.
- Alternative assessment records twice a year for tests without PT.
- Change control and revalidation records (when to revalidate).
How Sound Lab Consultants helps
- Map every LDT to the CAP items above and close gaps before the inspection window.
- Build the LDT list, clinical claims files and director approval records.
- Review report statements and client-facing claims.
- Respond to CAP deficiencies with documented corrective action.
Talk through your test with an LDT regulatory consultant. Bring your intended use, platform and target launch date.
Frequently asked questions
What CAP checklist items apply to laboratory developed tests?
Key items include COM.40350 (validation study), COM.40640 (clinical claims), COM.40475 (written assessment and director approval), COM.40200 (LDT list), COM.40630 (LDT report statement), COM.40610 (calibration and QC) and COM.01500 (alternative performance assessment). Item numbers can change between editions.
How many samples does CAP require to validate an LDT?
CAP guidance generally calls for at least 20 samples for accuracy, spread across the measuring range for quantitative tests or including positives, negatives and low positives for qualitative tests, unless the director documents why fewer are appropriate.
Does CAP require clinical validation of LDTs?
CAP requires the lab to validate any clinical claim it makes for an LDT (COM.40640), using literature or a clinical study. CLIA itself does not address clinical validity.
Does CAP require an LDT disclaimer on reports?
Yes. COM.40630 requires a statement on patient reports for LDTs indicating the test was developed and its performance characteristics determined by the laboratory.
Did the 2025 FDA LDT ruling change CAP requirements?
No. The ruling vacated FDA's 2024 LDT rule. CAP's checklist requirements for LDTs continue to apply to accredited labs.
Sources
- CAP test method validation and verification excerpt (PDF)
- CAP inspecting test method verification and validation decision tree (PDF)
- CAP Laboratory Accreditation Program
- 42 CFR 493.1253
- 42 CFR 493.1236
- 42 CFR 493.1281
- 42 CFR 493.1105
- 21 CFR 809.30
- CMS LDT and CLIA FAQs (PDF)
- ADLM position statement on LDTs (2024)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.