LDT Accuracy and Precision Studies

Short answer: For an LDT, accuracy shows results agree with a reference method or reference materials, and precision shows results repeat within a run, between runs, across days and across operators. CLIA requires both before patient testing under 42 CFR 493.1253(b)(2) but sets no sample count; CAP's published checklist excerpt sets 20-sample minimums.

Last reviewed: November 10, 2026. Primary sources: 42 CFR 493.1253(b)(2)(i)–(ii), SOM Appendix C, CAP COM.40300–40350 and the APHL Verification and Validation Toolkit.

Book a free consultation

What CMS expects

Accuracy. CMS guidance says a lab may establish accuracy by “testing reference materials or comparing results of tests performed using an established reference method; or comparing split sample results with results obtained from another method, which has already been shown to provide accurate results.” For qualitative methods the lab must show the method identifies presence or absence of the analyte. For a new analyte, “research results may be used to document the accuracy of the test by correlation with the clinical presentation” (SOM Appendix C).

Precision. The lab assesses “day-to-day, run-to-run, and within-run variation, as well as operator variance,” by repeat testing of known patient samples, QC material in duplicate over time, or calibration materials over time. Operator variance need not be evaluated for fully automated systems that are not user dependent (SOM).

How many samples?

CLIA itself sets no number. Accreditor and professional guidance fills the gap:

SourceAccuracyPrecision
CAP COM.40350 (public 2017 excerpt)Quantitative: at least 20 samples spread across the analytical measurement range. Qualitative: at least 20, including positive, negative and low positive samples near the detection limit; no equivocal samples. Multi-analyte methods (NGS, FISH, HPLC, GC-MS, MALDI-TOF) may be validated per method. Fewer samples require the director to record the rationale.COM.40310: repeat measurement at varying concentrations within-run and between-run over time
APHL toolkit, qualitative assaysIf no other guidance, 40 samples (20 positive, 20 negative) per CLSI EP09c, over at least 5 days; expected total agreement ≥95% with the reference methodAt least one positive and one negative near the cutoff, 10–30 total measurements, multiple days, more than one laboratorian
APHL toolkit, quantitative assaysSame approach, spanning the AMR, using reference materials, PT materials or characterized patient samplesPer CLSI EP05-A3: two runs a day at least 2 hours apart, duplicates, for 20 days

CAP's current checklists are available to subscribers. The excerpt cited here is CAP's publicly posted August 2017 version; confirm the current requirement text in your checklist edition.

Choosing samples

  • Use characterized patient specimens in the test's own matrix where possible; contrived (spiked) samples are acceptable when real ones are scarce (APHL).
  • Do not use routine QC materials or the calibrators used to calibrate the method to establish accuracy (CAP COM.40300).
  • If calibration material is used, show minimal matrix effect (SOM).
  • Include values near medical decision points and cutoffs, where errors matter most.

Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.

Book a free consultation

Statistics to report

Assay typeAccuracy statisticsPrecision statistics
Qualitative2×2 table: positive and negative percent agreement, overall agreement, kappa when the comparator is imperfect (APHL)Agreement of replicates near the cutoff; hit rate
QuantitativeMethod comparison regression, bias at decision levels, difference plotsSD and CV within-run, between-run, between-day and total

Write acceptance criteria into the plan before testing, approved by the laboratory director. See plan and report template.

After go-live

Accuracy keeps being checked through PT or twice-yearly alternative assessment (493.1236(c); guide), and through comparisons twice a year when the same test runs on more than one instrument, method or site (493.1281(a)). Precision is watched through daily QC (493.1256).

How Sound Lab Consultants helps

  • Size accuracy and precision studies to your test's risk and accreditor
  • Pick comparators and characterized samples
  • Run the statistics and write the summary for director sign-off

Talk through your test with a lab regulatory consultant.

Book a free consultation

Frequently asked questions

How many samples does CLIA require for LDT accuracy?

None specified. 42 CFR 493.1253 lists accuracy as a required characteristic but sets no number. CAP's published excerpt calls for at least 20 samples, and APHL cites CLSI EP09c's 40 for qualitative assays.

What is the difference between accuracy and precision?

Accuracy is closeness to the true value, shown against reference materials or a reference method. Precision is how closely repeated results agree, across runs, days and operators.

Can QC material be used for an LDT accuracy study?

CAP's checklist excerpt says routine QC materials and the calibrators used to calibrate the method are not appropriate for accuracy. QC materials can be used for precision.

How long should a precision study run?

CLIA sets no duration. APHL's toolkit cites CLSI EP05-A3's 20-day design for quantitative assays and multiple days with more than one laboratorian for qualitative assays.

Does operator variance have to be tested?

Yes, unless the system is fully automated and not user dependent, per CMS interpretive guidance for 493.1253.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.