New York State CLEP LDT Approval: Requirements and Process
Short answer: A laboratory anywhere in the world needs a New York State clinical laboratory permit before it tests specimens that originate in New York, and each laboratory developed test (LDT) it runs on those specimens needs approval from the Wadsworth Center's Clinical Laboratory Evaluation Program (CLEP) or FDA clearance. CLEP reviews LDTs only as part of a permit application or an existing permit: the lab files a validation package with a Risk Attestation Form, and CLEP assigns a high, moderate or low risk tier (or a clinical trial or not evaluated class) that decides whether the test can be offered before review ends.
Last reviewed: November 15, 2026. Primary sources: Wadsworth Center Test Approval page, the Tiered Evaluation of LDTs Policy (PDF), the CLEP LDT FAQ, Apply for a Permit, Certificate of Qualification requirements, the CLEP Program Guide, March 2026 (PDF), 10 NYCRR Subpart 58-1 and Subpart 58-6 (effective September 2, 2026), and Public Health Law Article 5, Title V (sections 574, 575, 576 and 578).
What New York State CLEP LDT approval is
CLEP LDT approval is New York's state-level review of tests a laboratory builds or modifies itself, and it is required before those tests are used on specimens from New York. The Clinical Laboratory Evaluation Program, part of the Department of Health's Wadsworth Center, regulates every clinical laboratory that tests specimens originating in the state, wherever the lab is located. In its own words, CLEP is currently the only regulator in the United States that formally reviews LDTs, and it has done so since 1991 (CLEP home; LDT review data).
The legal basis is Public Health Law section 576(3), which authorizes the Department to examine and approve a lab's own or altered test methods before New York patients are offered them and requires state standards at least as stringent as CLIA. 10 NYCRR 58-1.1 then says a permit is issued only when every test performed on New York specimens is either FDA approved, cleared or exempt, or approved or authorized by the Department. Wadsworth defines an LDT as a test that FDA has not cleared or approved and that the offering lab developed, including tests built from ASR, RUO or IUO reagents not covered by an FDA investigational device exemption (Test Approval).
In general, CLEP evaluates analytical validity, clinical validity and certain additional parameters (LDT FAQ). A 2025 article in The Journal of Molecular Diagnostics, coauthored with a Wadsworth Center scientist, summarizes the program as three principles: analytical validation, clinical validation and approval before the test is used (Galbo et al., 2025).
Key Approval Requirements & Process
LDT approval sits inside the permit process, so a lab must satisfy four groups of requirements: a permit, qualified directors, validation data, and inspections with proficiency testing. Wadsworth's Apply for a Permit page lists the conditions for issuing a permit, and LDT approval ("Test Approval, as applicable") is one of them.
- Directors obtain CQs first. The laboratory director and each technical director need Certificates of Qualification covering each category the lab will offer to New York patients.
- File the permit application with $1,100, the Clinical Laboratory Permit Application and the ownership disclosure statement. Corporations formed outside New York also need authorization from the New York Department of State to do business in the state.
- Return the Test Menu CLEP sends after reviewing the application, with a category selected for each test.
- Submit LDT packages using the category checklist plus a Risk Attestation Form for each test. While the permit is pending, each package is classified as high risk unless it qualifies as clinical trial or not evaluated.
- Enroll in proficiency testing and authorize the PT provider to release results to New York.
- Host the on-site survey and correct any deficiencies; out-of-state labs pay survey travel.
- Start New York testing only after the permit is issued and each LDT has explicit approval, and after the director's Health Commerce System (HCS) account is affiliated with the lab and fees are paid.
Mandatory Permit
Owning or running a lab that sits in New York, or that takes in New York specimens, requires a state permit, and CLEP reviews LDTs only for labs that hold or have applied for one. Public Health Law section 574 sets the permit requirement and calls for a separate permit for each testing facility. Wadsworth adds that LDT review begins once a permit application is on file and cannot be requested on its own (Test Approval; LDT FAQ).
| Permit rule | What it says | Source |
|---|---|---|
| Who applies | The owner and the director apply together; the permit is issued to both, and they are jointly responsible | PHL 575(1), (3) |
| Permit year | Permits run July 1 through June 30; initial permits can be issued at any point in the cycle | Apply for a Permit |
| Conditions | Director CQ (and state licensure where required), technical director CQ in each category, inspection with deficiencies corrected, successful PT, and only FDA or Department approved tests | 10 NYCRR 58-1.1(a)(1) |
| Scope | Only tests within the permit categories; other specimens must be referred to a lab holding the category and test-specific approval | 58-1.1(b) |
| Changes | A permit is void on a change of director, owner or location; the Department may stay the voiding if it is notified within 60 calendar days, fees are paid and the lab is in good standing (or shows a compelling reason) | PHL 575(6); 58-1.1(e) |
| Single use permit | Patient- or test-specific authorization for a lab lacking test approval or a permit, based on medical necessity or similar grounds; the testing lab applies in eCLEP-SUP before testing | 58-1.1(f); Single Use Permits |
Inside New York, CLEP operates as a CLIA-exempt state program, so in-state labs receive a CLIA number on the state permit. Labs outside New York get their CLIA certificate through the federal program and still need the New York permit (Apply for a Permit; NYSDOH, Oversight of Clinical Laboratories in NYS). More on state licensure: New York clinical lab license requirements.
Director Qualifications: Certificate of Qualification (CQ)
Anyone directing a lab that is in New York or accepts New York specimens must hold a New York Certificate of Qualification, and since September 2, 2026 there are two kinds: one for the laboratory director and a separate technical director certificate for each permit category a person oversees. Public Health Law section 572 sets the requirement, and Subpart 58-6 replaced the old Part 19 rules. A Laboratory Director CQ alone does not authorize oversight of a permit category (CQ requirements).
| Item | Laboratory Director CQ | Technical Director CQ |
|---|---|---|
| Role | Overall direction and operation of the lab | Technical and scientific operations for one or more categories |
| Valid for | Two years | Two years |
| Renew by | No later than 60 days before expiration | No later than 45 days before expiration |
| Core qualifications | Board-certified pathologist; or physician with two years supervising high complexity testing plus 20 credit hours of lab operations training; or earned doctorate with a listed board certification, two years of experience and 20 credit hours (separate route for parentage and forensic identity) | Category-specific requirements in 58-6.4; each category applied for needs experience from the last 10 years |
| Source | 10 NYCRR 58-6.2 | 10 NYCRR 58-6.4 |
- Transition. A current director who does not meet Subpart 58-6 may stay in place only until January 1, 2027, when the lab must have a qualified Laboratory Director (CQ requirements; 58-6.2(e)).
- Limits. One person may direct no more than five laboratories, and the director must spend regular part-time hours on site, which the rule sets at eight or more hours a week unless the Department approves less (58-1.2(b), (c)).
- Fee. The statute sets a $150 registration fee for each original and renewal CQ application (PHL 573(3)).
See also laboratory director services.
Validation Data
A full LDT submission is a method validation package built to a CLEP checklist, and it must show both analytical and clinical performance. Labs use the checklist for their category, or the General Checklist (February 2026) where no category checklist exists, and add a stand-alone Risk Attestation Form (Test Approval).
| Checklist section | What CLEP asks for |
|---|---|
| 1. General information | Test name and codes, test type (LDA, IUO, ASR or RUO), target population, method, analytes, validated specimen types, intended use and permit category. Panels and scores must be validated as a whole as well as assay by assay |
| 2.1 Procedure and controls | Procedure meeting standard TPC S1, including clinician and patient education materials, indications, specimen handling and rejection, assay principle and clinical validity, reagents (primer lists for amplification tests), calibrators, QC placement, calculations and limitations |
| 2.2 Requisition and reports | A sample requisition (TR S3) and reports for every possible result (REP S2), with required disclaimers such as ASR statements |
| 2.3 References | Primary literature supporting the scientific basis and clinical validity; package inserts |
| 2.4 Validation | Narrative summary, protocols and raw data: specimen and analyte stability, transport and storage, accuracy, within- and between-run precision, reportable range, analytical specificity and interference, clinical sensitivity and specificity (including blinding and discrepant resolution), algorithm validation, reference intervals or cutoffs, one representative run, and QC data for 20 to 30 consecutive runs |
| 2.5 Quality assurance | A PT plan, or an alternative-assessment plan under PT S3; agreement across instruments; a QC plan for critical steps |
| 2.6 Test design | Disclosure of any software, statistical model or AI method; algorithms must be locked, and in-house algorithms need a training and validation summary |
Category checklists and guidelines exist for molecular genetic testing, germline NGS, somatic NGS, molecular oncology, infectious agent NGS, molecular microbiology, MALDI-TOF, FISH cytogenetics, cellular and diagnostic immunology, forensic identity, toxicology and trace elements. CLEP reviews software only within the specific LDT and does not approve it for other tests (LDT FAQ). For the CLIA side of the same studies, see LDT validation requirements, LDT clinical validity and NGS assay validation.
Inspections & Proficiency Testing
A lab must pass an on-site survey and take part in proficiency testing before New York issues a permit, and both continue for as long as the permit is held. Out-of-state surveys are announced; routine surveys of in-state labs after the first are unannounced. A survey is not scheduled until the lab has taken part in any required PT and has submitted any required validation (On-Site Survey).
- Survey scope. Premises, practice, equipment, personnel and record-keeping are checked against Article 5, Title V of the Public Health Law, 10 NYCRR and the Clinical Laboratory Standards of Practice. Labs complete a Facility Personnel Form beforehand (On-Site Survey).
- Postponements. Filing an application signals that the lab is prepared for a survey; postponing it three or more times closes the application (CLEP Program Guide).
- Travel costs. Out-of-state labs pay the survey's travel expenses, based on the number of surveyors, days and facilities, which Wadsworth bills after the survey and which must be paid before a permit is issued; unpaid travel fees can lead to suspension or nonrenewal (Laboratory Fees; 10 NYCRR 58-3.8).
- PT. Labs enroll with a CMS-approved provider for every test New York mandates and designate the provider each year in eCLEP. For tests without mandated PT, accuracy must be checked another way at least twice a year under PT S3. Unsatisfactory results require investigation and corrective action (Proficiency Testing). See LDT proficiency testing alternatives.
- FDA tests later. Labs in good standing can add FDA approved, cleared or exempt tests to an existing category; the surveyor reviews that validation at the next survey (CLEP Program Guide).
Adding New York to your test menu? Sound Lab Consultants can map your permit categories, CQs and validation files to the CLEP checklists before you file.
Risk-Tiered Approval Policy
CLEP sorts every LDT submission into a risk class, and the class decides whether the test can be offered before technical review is complete. The Wadsworth Center adopted the three-tier model on November 14, 2016 and revised it effective November 7, 2023 to add a class for tests the Department chooses not to evaluate. Clinical Laboratory Reference System (CLRS) staff assign the class from the lab's Risk Attestation Form (Tiered Evaluation Policy).
| Class | Submission required | Initial approval | Technical review | Review priority |
|---|---|---|---|---|
| High | Yes | None | Yes | High |
| Moderate | Yes | Conditional (if the lab holds the category) | Yes | Medium |
| Low | Yes | Full (if the lab holds the category) | No | Not applicable |
| Clinical trial | Yes | Not applicable | No | Not applicable |
| Not evaluated | Yes | Not applicable | Not applicable | Not applicable |
Footnotes to the table: a lab still awaiting its permit or the relevant category gets the high risk tier by default, except for clinical trial or not evaluated packages, and the Department may withhold approval or review any application at its discretion (Test Approval; policy). The Test Approval page says labs hear their class within three weeks of a complete package; the policy document says generally three to six weeks. CLEP does not make risk determinations before a submission; questions about the form go to clepval@health.ny.gov.
High Risk
A high risk LDT gets no conditional approval: it cannot be used on New York specimens until CLRS finishes its review and grants full approval. Under the policy, a test is high risk when the lab has not yet shown its method to be well established in-house and a wrong result could steer a patient toward a mistaken diagnosis or treatment likely to cause serious harm or death (high impact). That holds whether or not the result is a key determinant. Every package from a lab still pending a permit or the needed category is also treated as high risk, including in distributive testing arrangements where either lab is pending.
High risk reviews are prioritized, but the policy states that no timeline can be given. Labs must answer CLRS questions within 60 business days, with 60-day extensions available by request, or the application is inactivated (Tiered Evaluation Policy).
Moderate Risk
A moderate risk LDT receives conditional approval, so a lab that already holds the right permit category may offer it on New York specimens as soon as CLEP gives notice. The policy places three combinations here: a well-established method with a high-impact result that is a key determinant; a well-established method with a high-impact result that is not a key determinant; and a method not yet well established in the lab whose result is neither a key determinant nor high impact.
The Department may withhold or withdraw conditional approval. CLRS can defer review of a moderate risk test indefinitely when the lab's method is well established, the result is not a key determinant, or a wrong result would carry little patient-safety risk, and only tests that are actually reviewed can earn full approval. If a review is opened, the lab has 60 business days to respond (extensions available) or conditional approval is rescinded (Tiered Evaluation Policy).
Low Risk
A low risk LDT receives full approval without CLRS technical review, provided the lab holds a permit in the right category, and the lab may offer it once notified. The policy reserves low risk for a well-established method whose result is not a key determinant and whose inaccurate result is unlikely to support a wrong diagnosis or treatment (low impact). Tests filed under an approved methodology exemption can land in the low tier as well. The Department keeps the right to withhold approval or require review of any LDT (Tiered Evaluation Policy).
Clinical Trial and Not Evaluated Classes
- Clinical trial. For an LDT used solely in a trial approved by NIH or another independent IRB, CLEP sends notice that it may be performed on participant specimens. The class does not apply if the test is also used diagnostically, and later diagnostic use needs a new full package.
- Not evaluated. CLEP may decline to evaluate analytical and clinical validity when the intended use makes no clinical claim about a recognized disease, or relates to general wellness or a healthy lifestyle, and results could not support a treatment likely to cause significant harm. The test may then be offered with a disclaimer, on both marketing materials and reports, that the New York State Department of Health has not evaluated it. An authorized ordering source is still required, and documents may be shared with CMS for a CLIA determination (policy; Test Approval).
How CLEP Assigns the Tier: the Risk Attestation Form
The tier comes from three factors defined in the policy: whether the method is well established in the lab, whether the result is a key determinant, and how much harm an inaccurate result could cause. "Well established" means the lab has a consistent record of complete, organized submissions using the same or similar technology, and the method has been approved by FDA or New York or has produced meaningful clinical impact in multiple peer-reviewed publications. A "key determinant" result gives critical information to diagnose, predict a greater likelihood of disease, establish treatment eligibility, give prognosis that changes management, or show treatment adherence or drug use.
The Risk Attestation Form (January 2024) asks for a summary of up to 500 words (intended use and population, method, specimen types) and answers on: clinical claims; clinical-trial-only use and IRB status; whether the test modifies an FDA device or a CLEP-approved LDT, and what changed; prior fully approved LDTs using the method; whether the method is well established and covered by an exemption; how the clinical claim was established; key determinants; and the impact of an inaccurate result. Most narrative answers are limited to 200 words. Vague or incomplete answers can delay review (policy).
Which Tests Need a CLEP Submission
LDTs, modified FDA tests, and kits sold without FDA clearance all need a submission, while unmodified FDA tests, standard methods and legacy assays do not. The Test Approval page sorts tests as follows.
| Test type | What to file |
|---|---|
| Unmodified FDA cleared (510(k)), approved (PMA), exempt or EUA test; IUO only under an FDA IDE | No validation submission. If the lab holds the category, it can offer the test; otherwise it requests the category in eCLEP (or completes the Test Menu as a new applicant) |
| Standard method (for example IHC and classic ISH under pathologist-guided processes, culture-based microbial ID, STR parentage testing, protein electrophoresis, metaphase karyotyping, manual differentials) | No submission; same category rules |
| Legacy assay in use before April 24, 2003 | Performance specifications still verified or established, but procedure and validation submitted only if requested |
| Modification changing intended use (specimen type, qualitative vs quantitative, purpose, population) or procedure | Assay Modification Form plus Risk Attestation Form; explicit approval before patient testing. A standard method moved onto newer technology is a new LDT needing full validation |
| Commercial kits labeled RUO or IUO, or unlabeled | Category checklist plus Risk Attestation Form |
| Laboratory developed test | Category checklist plus Risk Attestation Form; applicants also need the permit or amended permit |
| Digital pathology slide imaging, including AI/ML | Not an LDT submission; a licensed pathologist must still sign out cases, and FDA approval may apply |
In toxicology, adding drug targets to an approved assay using the same matrix and extraction counts as a modification, but a new specimen matrix needs a full LDT submission. Background: when a modified FDA-cleared test becomes an LDT and LDT vs IVD vs RUO.
Submitting and Tracking a CLEP Package
CLEP wants one paper copy and one identical PDF copy on a CD or flash drive, with the Risk Attestation Form as its own file, mailed to the address on the checklist. The General Checklist addresses packages to the Clinical Laboratory Evaluation Program, Biggs Laboratory, Wadsworth Center, Empire State Plaza, Albany, NY 12237, Attn: Assay Validation Review (General Checklist).
- Digital rules. OCR any scanned PDF; no single PDF over 50 MB; one PDF covering all checklist elements; raw data in up to five separate files; no unrequested material.
- Confidentiality. Mark the cover of each unit "proprietary" or "confidential" so the lab is told of any Freedom of Information Law request and can try to block release.
- Project ID. CLEP assigns a PID, adds it to the file names and emails it to confirm receipt.
- Tracking. Check the Test Approval Status link on the eCLEP home screen inside the Health Commerce System, or email the CLEP Validation Unit with the four-digit PFI and the PID in the subject line.
- Responding. Answer requests for more information within 60 days (the policy says 60 business days), in paper and digital form; extensions can be requested, and silence inactivates the request.
- Discontinued tests. Email the Validation Unit to remove an LDT from the approval list, because approved methods are shared with referring labs, consumers and providers on request.
CLEP's published data show that 46% of LDTs approved by September 30, 2021 needed no request for more information, 33% needed one, 14% two, 5% three and 2% more than three (LDT review data).
Exemptions From Comprehensive Submissions
Once New York has approved a representative set of a lab's tests on one methodology, such as FISH, Sanger sequencing or LC-MS/MS, the lab can request an exemption so later tests on that method need only a short submission. The request includes the Exemption Request Form, a standardized validation protocol and the lab's ongoing validation, QC and QA protocols. Approval comes by letter with a PID, and the Department can rescind it at any time (Test Approval).
- Five-year term. Exemptions last five years, including those approved before 2024. Renewal needs a full package for a representative assay not submitted before; do not file more than one month before expiration.
- Adding a test. File the Add Under Exemption Form with the exemption PID, a validation summary, sample reports for every outcome and a Risk Attestation Form, plus informed consent education materials for molecular genetic testing and restricted or prenatal cytogenetics. The lab must wait for written confirmation before testing (LDT FAQ).
Clinical Trials, Research, Wellness and At-Home Collection
- Clinical trials. A permit is needed when trial results are reported and used for decisions such as enrollment, arm assignment, safety, toxicity or dosing. Submit a Risk Attestation Form, a narrative of the testing purpose, a validation summary, sample reports, and one of: an IRB letter stating no significant risk, the clinicaltrials.gov identifier, or an FDA no-significant-risk letter. Tests under an FDA IDE count as FDA approved.
- Research. Research testing is outside the permit requirement under Public Health Law section 580(2) when no patient-specific result is reported; reporting results requires a permit and test approval first.
- Wellness tests. These must still be submitted as a full package with a Risk Attestation Form; CLEP may then classify them as not evaluated (LDT FAQ).
- At-home collection. The lab needs the permit category plus FDA or New York approval of the test. New York does not review collection kits; a switch to home collection of the same specimen type needs no review, but a change in the specimen itself (for example whole blood to dried blood spots) must be submitted. A licensed provider must order the test.
Source: Test Approval. Labs run by licensed practitioners solely for their own patients are outside the permit law (PHL 579), and a New York physician office lab holding a CLIA certificate through POLEP or accreditation is not eligible for CLEP LDT review unless it voluntarily obtains a CLEP permit (LDT FAQ).
Fees and Costs
CLEP charges no separate fee for LDT review; the cost is built into the permit and reference fees every permitted lab pays. Those fees are set under Public Health Law section 576 and Subpart 58-3, and the process is the same for labs inside and outside New York (LDT FAQ).
| Cost | Amount or basis | Source |
|---|---|---|
| Initial permit application | $1,100 by check: $100 non-refundable application fee plus $1,000 inspection and reference fee; no credit cards or non-U.S. bank checks | Apply for a Permit; PHL 575(4) |
| Annual permit and reference fee | Program operating expenses multiplied by the lab's share of total gross annual receipts (GAR); out-of-state labs report revenue from New York specimens only; invoices in June, quarterly payments allowed | Laboratory Fees; PHL 576(4) |
| Out-of-state survey travel | Travel or mileage plus per diem for the surveyors, split among labs on the same trip. The fee page says the bill comes after the survey; 58-3.8 describes an estimate paid before the inspection and adjusted to actual cost | Laboratory Fees; 58-3.8 |
| Certificate of Qualification | $150 per original or renewal application | PHL 573(3); CQ page |
Penalties, CLIA and FDA
- Penalties. Owning or operating a lab without a valid permit is a misdemeanor punishable by up to one year of imprisonment and a fine of up to $2,000, or both; acting as a director without a CQ carries up to one year and up to $500 (PHL 578). Running a test without the needed permit category counts as testing without approval (CLEP Program Guide), and section 577 allows permits and CQs to be suspended, limited or revoked.
- CLIA. CMS granted New York exempt status in 1995 because its requirements are equal to or stricter than CLIA. A CLIA certificate or accreditation does not replace the New York permit for labs outside the state (CLEP Program Guide).
- FDA. FDA's 2024 LDT rule was vacated by a federal district court on March 31, 2025, and FDA reverted the regulation on September 19, 2025 (FDA). New York's review rests on state law, so it continues. Wadsworth's FAQ adds that FDA could still require its own clearance for a given test. More: LDT regulatory oversight.
Searching Approved LDTs and Permitted Labs
Wadsworth publishes a free search of approved LDTs that anyone can filter by analyte or test name, specimen type, method, permit category, facility name, PFI, state and approval status. On November 15, 2026 the Search Approved Laboratory Developed Tests tool showed 11,243 records. A listing does not prove the lab still offers the test, standard methods are not included, and tests marked conditionally approved have not been reviewed but are allowed on New York samples. The companion approved laboratories search listed 1,137 facilities by permit category the same day.
How Sound Lab Consultants Helps
- Mapping each test to its permit category and to the right CLEP checklist
- Planning CQ coverage for the laboratory director and each category under Subpart 58-6
- Building validation packages, procedures, sample reports and Risk Attestation answers
- Preparing for the on-site survey and PT designations
- Responding to CLEP requests for more information and planning exemption requests
Planning a New York permit or LDT submission? Review your plan and gaps with Sound Lab Consultants.
Frequently Asked Questions
What is NYS CLEP LDT approval?
It is the New York State Department of Health's review of laboratory developed tests, carried out by the Wadsworth Center's Clinical Laboratory Evaluation Program (CLEP) under Public Health Law section 576 and 10 NYCRR Subpart 58-1. A lab that tests specimens from New York may use an LDT on those specimens only after CLEP approves it, at the level its risk classification allows, or after FDA clears or approves it.
Does an out-of-state lab need New York approval for its LDTs?
Yes. The permit requirement covers labs located in New York and labs that accept specimens from New York, and Wadsworth says the permit and LDT approval process is the same inside and outside the state. Out-of-state labs also pay the travel costs of their on-site survey, and CLEP's 2018 to 2020 data show 77% of LDT submissions came from labs outside New York.
Can CLEP review an LDT without a New York permit?
No. Wadsworth treats LDT review as one piece of permitting: CLEP does not sell it separately, and it starts only once the lab has filed its permit application. A lab still waiting on its permit, or on the category the test falls in, has every package placed in the high risk tier, except packages that fit the clinical trial or not evaluated classes.
What are the CLEP risk tiers for LDTs?
Under the Tiered Evaluation of Laboratory Developed Tests Policy (revised November 2023), CLEP classifies each LDT as high, moderate or low risk, or as clinical trial or not evaluated. High risk tests need full approval before use; moderate risk tests get conditional approval and can be offered once CLEP gives notice; low risk tests get full approval without technical review. The tier depends on whether the method is well established in the lab, whether the result is a key determinant, and the impact of an inaccurate result.
Can a moderate risk LDT be offered before CLEP finishes its review?
Yes, if the lab already holds a permit in the right category. CLEP grants conditional approval to moderate risk tests, and the lab may start testing New York specimens once it is notified of that classification. The Department can withhold or withdraw conditional approval, may defer the technical review indefinitely, and only reviewed tests can receive full approval.
How long does NYS CLEP LDT approval take?
Wadsworth does not publish a review timeline; its policy says one cannot be given because of the many variables. The Test Approval page says labs are told their risk class within three weeks of a complete package, while the policy document says generally three to six weeks. In CLEP's published data, 46% of approved LDTs needed no request for more information and 54% needed one or more.
Is there a fee for CLEP LDT review?
There is no fee specific to LDT review. Wadsworth says the cost is built into permit and reference fees, which are set under Public Health Law section 576 and Subpart 58-3. A new permit application is accompanied by $1,100 ($100 for the non-refundable application plus $1,000 toward inspection and reference costs), the annual fee is based on the lab's gross annual receipts, and out-of-state labs are billed for survey travel.
Who needs a New York Certificate of Qualification?
Every laboratory director must hold a Laboratory Director CQ, and each permit category must be overseen by someone holding a Technical Director CQ in that category, under Subpart 58-6 adopted September 2, 2026. Both certificates are valid for two years. A director who does not meet the new Subpart 58-6 requirements must be replaced by a qualified Laboratory Director by January 1, 2027.
Do FDA-cleared or FDA-approved tests need CLEP approval?
An unmodified FDA-cleared, approved, exempt or EUA test does not need a CLEP validation submission, and Wadsworth's FAQ says an FDA-cleared LDT does not need CLEP review. The lab still needs a New York permit in the right category before testing New York specimens. A modification that changes the intended use or procedure needs an Assay Modification Form and a Risk Attestation Form.
What goes in a CLEP LDT validation package?
The category checklist (or the General Checklist) lists the contents: test information and intended use; the full procedure with controls; a sample requisition and reports for every possible outcome; references; a narrative validation summary with protocols and raw data for stability, accuracy, precision, reportable range, analytical specificity, clinical sensitivity and specificity and cutoffs; one representative run; QC data for 20 to 30 consecutive runs; a PT or alternative assessment plan; and answers on any software or algorithm, which must be locked. A Risk Attestation Form goes with every package.
How do I track a CLEP LDT submission?
Each logged submission gets a Project ID (PID), which CLEP emails to the lab and adds to the file names. Labs can follow progress in eCLEP on the Health Commerce System, where the home page has a Test Approval Status link, and can email the CLEP Validation Unit with the lab's four-digit PFI and the PID in the subject line. Requests for more information must be answered within 60 days unless an extension is granted.
What happens if a lab tests New York specimens without a permit or test approval?
Public Health Law section 578 makes owning or operating a clinical laboratory without a valid permit a misdemeanor punishable by up to one year in jail, a fine of up to $2,000, or both. The CLEP Program Guide says testing without the required permit category counts as testing without Department approval and can lead to administrative action, and section 577 allows permits and certificates to be suspended, limited or revoked.
How can I check whether an LDT is approved in New York?
Use the Wadsworth Search Approved Laboratory Developed Tests tool, which can be filtered by analyte or test name, specimen type, method, permit category, facility, PFI, state and status. On November 15, 2026 it listed 11,243 records. Tests shown as conditionally approved have not yet been reviewed but may be performed on New York samples. A separate tool lists permitted laboratories by category.
Did the FDA LDT rule change New York's requirements?
No. FDA's 2024 LDT rule was vacated by a federal court on March 31, 2025, and FDA restored the earlier regulation text on September 19, 2025. New York's LDT review rests on state law (Public Health Law section 576 and 10 NYCRR 58-1.1), which was not affected, so the permit and approval requirements still apply.
Is Sound Lab Consultants affiliated with the New York State Department of Health?
No. Sound Lab Consultants is an independent consulting firm and is not affiliated with or endorsed by the New York State Department of Health, the Wadsworth Center, CLEP, CMS or FDA. Submissions to CLEP are made by your laboratory.
Sources
- Wadsworth Center CLEP: Test Approval, Clinical Laboratory Evaluation Program, Frequently Asked Questions about LDTs, Search Approved LDTs, Search Approved Laboratories, Apply for a Permit, Certificate of Qualification, Laboratory Fees, On-Site Survey, Proficiency Testing, Single Use Permits, Laboratory Standards, Health Commerce System, Laws and Regulations, CLEP Releases LDT Review Data
- Wadsworth forms and policies (PDF): Tiered Evaluation of LDTs Policy (rev. Nov 2023), Risk Attestation Form (Jan 2024), General Checklist (Feb 2026), Assay Modification Form, Add Under Exemption Form, Somatic NGS Guidelines (Oct 2024), Germline NGS Guidelines (Oct 2024), CLEP Program Guide (Mar 2026)
- 10 NYCRR: Subpart 58-1, 58-1.1, 58-1.2, Subpart 58-3, 58-3.8, Subpart 58-6, 58-6.2, 58-6.4
- NY Public Health Law Article 5, Title V: 572, 573, 574, 575, 576, 577, 578, 579, 580
- NYSDOH: Oversight of Clinical Laboratories in New York State (2018, PDF)
- Galbo PM Jr et al. A Comprehensive Guide to Achieving New York State Clinical Laboratory Evaluation Program Approval for Next-Generation Sequencing Assays, J Mol Diagn 2025;27(6):485-501 (open access)
- FDA: Laboratory Developed Tests
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for the New York State Department of Health, the Wadsworth Center, CLEP, CMS, CDC or FDA. This page is general regulatory information, not legal advice. Confirm current requirements with the Wadsworth Center before filing. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.