How to Develop a Laboratory Developed Test (LDT), Step by Step
Short answer: To develop a laboratory developed test (LDT), a high complexity CLIA lab defines the intended use, sets design inputs and performance goals, optimizes the assay, validates it under 42 CFR 493.1253(b)(2) against acceptance criteria set in advance, writes the procedure and QC plan, gets director approval, then monitors and controls changes after launch.
Last reviewed: November 18, 2026. Primary sources: 42 CFR 493.1253, NIH SEED LDT guide, APHL verification and validation toolkit and CAP validation checklist excerpt.

The 9-step LDT development lifecycle
| Step | What you produce | Rule or guide | SLC guide |
|---|---|---|---|
| 1. Define intended use | Analyte, specimen types, population, clinical purpose, claims | NIH SEED: intended use, population and claims drive the test profile | Clinical validity |
| 2. Check the pathway | LDT, modified FDA test or verification; certificate and specialty; states and payers | 493.1253(b)(1) vs (b)(2) | LDT vs IVD vs RUO |
| 3. Set design inputs and risk level | Performance goals (detection capability, bias, precision, measuring interval), risk assessment | NIH SEED sections 2.2–2.3 | Design controls |
| 4. Optimize (feasibility) | Reagent, cutoff, extraction and instrument settings fixed; design frozen | Lab practice; AMP/CAP NGS guideline separates optimization from validation | Molecular LDTs |
| 5. Write the validation plan | Studies, sample numbers, statistics and acceptance criteria set in advance | APHL toolkit; CAP COM.40350 | Plan template |
| 6. Run analytical validation | Accuracy, precision, analytical sensitivity, specificity with interferences, reportable range, reference intervals, other characteristics | 493.1253(b)(2) | LDT validation |
| 7. Build QC, calibration and the procedure | QC frequency and levels, calibration, written procedure, report wording | 493.1253(b)(3); 493.1251; 493.1256 | Report statement |
| 8. Approve and launch | Validation report, director sign-off, training and competency, LIS build, LDT list | 493.1445; CAP COM.40475 and COM.40200 | CAP LDT requirements |
| 9. Monitor and control change | QC review, PT or alternative assessment twice a year, change control and revalidation | 493.1236(c); 493.1253 | Revalidation |
Step 1–3: intended use, pathway and design inputs
The NIH SEED guide says several factors decide whether a new product is an LDT, “including the test's intended use, patient population, claims, manufacturing, labeling, and marketing plans,” and that the developer should set “formal specifications for the test, including the desired performance criteria required to meet defined clinical needs” (NIH SEED). It then lists analytical performance specifications to set: analytical specificity, detection capability, bias, precision and measuring interval.
SEED also recommends establishing a risk level (low, moderate or high) based on the potential harm of an incorrect result, separate from CLIA complexity. Write the intended use and risk level down first; every later study is judged against them.
Decide early whether you are building a true in-house test, modifying an FDA-cleared kit (see modified FDA-cleared tests) or simply verifying an unmodified kit (see verification vs validation). Confirm your CLIA certificate covers high complexity testing in the right specialty (CLIA requirements for LDTs).
Step 4: optimization before validation
Optimization is where you change things: primer and probe sets, extraction chemistry, cutoffs, instrument methods, bioinformatics settings. Validation is where you stop changing them and prove performance. The AMP/CAP NGS oncology guideline recommends optimization and familiarization before formal validation (AMP/CAP). Keep optimization records: they explain why the final design looks the way it does, and inspectors and New York reviewers often ask.
Steps 5–6: validation plan and analytical studies
CLIA requires the lab to establish, before reporting patient results, “accuracy; precision; analytical sensitivity; analytical specificity to include interfering substances; reportable range of test results for the test system; reference intervals (normal values); and any other performance characteristic required for test performance” (42 CFR 493.1253(b)(2)). CLIA sets no sample counts; CAP's excerpt gives sample-number guidance such as at least 20 samples for accuracy unless the director documents a reason (CAP).
| Characteristic | CFR | Guide |
|---|---|---|
| Accuracy and precision | 493.1253(b)(2)(i)–(ii) | Accuracy and precision |
| Analytical sensitivity (LoD) | (iii) | Limit of detection |
| Analytical specificity and interference | (iv) | Specificity |
| Reportable range | (v) | Reportable range |
| Reference intervals | (vi) | Reference intervals |
| Other: stability, matrix, carryover, bioinformatics | (vii) | Specimen and stability |
Building, validating or fixing an LDT? We help CLIA labs plan and document LDT development and validation and prepare for CLIA, CAP and New York State review.
Steps 7–8: QC, procedure, approval and launch
- Base calibration and control procedures on the validation data (493.1253(b)(3)).
- Write the procedure manual entry and have it approved, signed and dated by the director before use (493.1251).
- Document all validation activities and keep them while the test is in use and at least 2 years after (493.1105(a)(3)(i)).
- For CAP-accredited labs: director approval of the validation (COM.40475), an entry on the LDT list (COM.40200) and the LDT report statement (COM.40630). See CAP LDT requirements.
- For New York specimens, plan the NYS CLEP test approval before launch (NYS CLEP).
- For Medicare billing of molecular tests in MolDX states, register in DEX and plan the technical assessment (LDT reimbursement).
Step 9: monitoring and change control
After go-live, run QC under 493.1256, enroll in PT where it exists, and where it does not, verify accuracy at least twice a year under 493.1236(c) (alternative assessment). Every change to reagents, specimen types, instruments, software or cutoffs needs an impact assessment and, where performance could shift, revalidation (when to revalidate; design and change control).
How long LDT development takes
There is no regulatory timeline; it depends on the method, sample availability (especially rare positives) and how many specimen types and claims you add. Practical drivers we see in every project: sourcing positive and interfering samples, multi-day precision runs, bioinformatics locking for NGS, and external reviews such as NYS CLEP or MolDX that run after validation is complete.
How Sound Lab Consultants helps
- Turn your idea into a written intended use, design inputs and risk assessment.
- Build the validation plan and acceptance criteria before any data are collected.
- Review optimization and validation data and draft the report for director sign-off.
- Prepare the CAP, NYS CLEP and MolDX pieces in parallel so launch is not held up.
Talk through your test with an LDT regulatory consultant. Bring your intended use, platform and target launch date.
Frequently asked questions
What are the steps to develop an LDT?
Define intended use, confirm the regulatory pathway, set design inputs and risk level, optimize the assay, write a validation plan, run analytical validation under 42 CFR 493.1253(b)(2), build QC and the procedure, obtain director approval and launch, then monitor and control changes.
What is the difference between LDT development and validation?
Development (design and optimization) is where the lab changes the assay until it meets its goals. Validation starts once the design is locked and proves performance against acceptance criteria set in advance.
Who can develop an LDT?
A laboratory certified under CLIA to perform high complexity testing. The laboratory director is responsible for approving the validation and the procedure before patient testing.
Do I need design controls for an LDT?
CLIA does not use the term design controls, but it requires documented validation and procedures. Many labs apply design-control practices such as design inputs, reviews and change control because they make validation and inspections easier.
How many samples are needed to validate an LDT?
CLIA sets no number. CAP guidance generally calls for at least 20 samples for accuracy, including positives, negatives and low positives for qualitative tests, unless the director documents why fewer are adequate.
Do I need FDA clearance to launch an LDT?
For most LDTs, no, after the March 31, 2025 court ruling vacating FDA's LDT rule. CLIA validation, accreditor requirements and New York approval still apply.
Sources
- 42 CFR 493.1253
- 42 CFR 493.1251
- 42 CFR 493.1256
- 42 CFR 493.1236
- 42 CFR 493.1105
- NIH SEED regulatory knowledge guide for LDTs (PDF)
- APHL verification and validation toolkit
- CAP test method validation excerpt (PDF)
- AMP/CAP NGS oncology panel validation guideline (2017)
- CMS LDT and CLIA FAQs (PDF)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.