Analytical Sensitivity and Limit of Detection for LDTs
Short answer: Analytical sensitivity is the lowest amount of analyte an LDT can reliably measure or tell apart from a blank, usually reported as the limit of detection (LoD). CLIA requires labs to establish it for LDTs under 42 CFR 493.1253(b)(2)(iii). Many labs follow CLSI EP17 designs and set LoD at 95% detection.
Last reviewed: November 10, 2026. Primary sources: 42 CFR 493.1253(b)(2)(iii), SOM Appendix C, CAP COM.40400 and APHL toolkit.
What CMS means by analytical sensitivity
CMS: “The laboratory is responsible for determining the lowest concentration or amount of the analyte or substance that can be measured or distinguished from a blank, i.e., minimum detection limits or how much of the analyte must be present to be measured.” For modified test systems, the lab may use the manufacturer's lower limit “if it has demonstrated that the modification has not affected the lower limit” (SOM Appendix C). CAP COM.40400 requires LDTs and modified tests to establish the lower detection limit, as applicable (CAP excerpt).
A typical LoD study design
- Prepare a sample with known analyte content in the test's real matrix, then make serial dilutions spanning the expected LoD.
- Run replicates over several days. APHL, citing CLSI EP17-A2, describes duplicate or triplicate runs over three days, with about 20 measurements per concentration to verify a manufacturer's claim and 60 to establish an LoD (APHL).
- Calculate the hit rate (positives ÷ replicates) at each level.
- Set the LoD at the lowest level detected 95% of the time; probit regression is a common way to estimate it.
- Confirm the claimed LoD with a fresh set of replicates and record it in the procedure.
APHL notes the exact number of samples should be set case by case with the laboratory director.
LoB, LoD and LoQ
| Term | Meaning | Where it is used |
|---|---|---|
| Limit of blank (LoB) | Highest result expected from samples with no analyte | Quantitative assays, to separate signal from noise |
| Limit of detection (LoD) | Lowest amount reliably distinguished from the blank; for qualitative tests often the 95% hit-rate level | All LDTs (493.1253(b)(2)(iii)) |
| Lower limit of quantitation | Lowest amount measured with acceptable precision and bias | Sets the low end of the reportable range for quantitative tests |
The low end of the reportable range should not fall below what the data support. Results below it are reported as “less than”; CMS notes detection limits “must be established through method verification” (SOM). See reportable range.
Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.
Molecular and NGS specifics
For PCR, express LoD in copies or organisms per mL of the specimen type, carried through extraction. For NGS, LoD is often defined as the lowest variant allele fraction detected reliably at a given depth. See molecular LDT validation and NGS validation.
Analytical vs clinical sensitivity
Analytical sensitivity is about detecting small amounts of analyte. Clinical (diagnostic) sensitivity is the share of patients with the condition who test positive. CLIA requires the first; clinical sensitivity belongs to clinical validity.
How Sound Lab Consultants helps
- Design LoD studies in the correct matrix
- Run probit or hit-rate analysis
- Tie the LoD to reportable range, QC levels and report wording
Talk through your test with a lab regulatory consultant.
Frequently asked questions
What is analytical sensitivity in CLIA?
CMS defines it as the lowest concentration or amount of analyte that can be measured or distinguished from a blank, that is, the minimum detection limit.
How many replicates are needed to establish an LoD?
CLIA sets no number. APHL's toolkit, citing CLSI EP17-A2, describes about 60 measurements per concentration to establish an LoD and 20 to verify a manufacturer's claim, over three days.
Can I use the manufacturer's LoD for a modified test?
Yes, if you show the modification did not affect the lower limit, per CMS interpretive guidance.
Is analytical sensitivity the same as clinical sensitivity?
No. Analytical sensitivity is the smallest amount detected. Clinical sensitivity is the percentage of people with the condition who test positive.
What hit rate defines the LoD?
Commonly 95%: the lowest concentration detected in at least 95% of replicates, as described in APHL's toolkit.
Sources
- 42 CFR 493.1253
- CMS SOM Appendix C (PDF)
- CAP Test Method Validation and Verification excerpt (PDF)
- APHL V&V Toolkit: Qualitative Assays (PDF)
- APHL V&V Toolkit: Quantitative Assays (PDF)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.
