Analytical Specificity and Interference Studies for LDTs
Short answer: Analytical specificity is how well an LDT measures only its target, without cross-reactivity or interference. CLIA requires labs to establish it, including interfering substances, under 42 CFR 493.1253(b)(2)(iv). CMS lists hemolysis, anticoagulants, lipemia, turbidity, disease states and medications as interferences to document from literature or the lab's own testing.
Last reviewed: November 10, 2026. Primary sources: 42 CFR 493.1253(b)(2)(iv), SOM Appendix C, CAP COM.40450/40500 and APHL toolkit.
What CMS requires
CMS: “The laboratory must determine the extent to which the method measures the analyte for which it is reporting results.” For interfering substances, “the laboratory must document information regarding interfering substances from product information, literature, or its own testing. These may include: specimen hemolysis, anticoagulant, lipemia, and turbidity; patients' clinical conditions, disease states, and medications” (SOM Appendix C).
CAP COM.40450 requires enough samples to establish specificity for LDTs, and COM.40500 requires the lab to know each test's interferences and have “an appropriate plan of action when they are present” (CAP excerpt).
How to design the study
- Cross-reactivity: test related organisms, analogs, metabolites or variants likely to be present or to cause similar symptoms. APHL suggests three to five samples for each potential cross-reactant, with results matching expectations at least 95% of the time (APHL).
- Interference: spike or pool samples with the interferent at clinically high levels and compare with unspiked aliquots, using an acceptance limit set in the plan.
- Inhibition: for amplification assays direct from specimens, consider an inhibition control (APHL; 493.1256(d)(3)(v)).
- If cross-reactivity cannot be removed, list it as a limitation of the assay (APHL).
Example interferent panels by method
| Method | What to challenge |
|---|---|
| Chemistry and immunoassay | Hemolysis, lipemia, icterus/bilirubin, turbidity, anticoagulants and tube additives, common drugs, heterophile antibodies where relevant |
| PCR and other NAATs | Genetically related organisms, normal flora, human DNA load, inhibitors such as blood or mucus, transport media |
| Mass spectrometry | Isobaric and structurally similar compounds, metabolites, ion suppression or enhancement, carry-over |
| NGS | Pseudogenes and homologous regions, contamination between samples, low-quality or degraded DNA |
CMS's 2026 CLIA request for information listed carry-over, internal standards and ionization among characteristics it asked labs about (91 FR 43586).
Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.
Where the results go
- The procedure manual must list “limitations in the test methodology, including interfering substances” (493.1251(b)(9)).
- On request, the lab must give clients information that may affect interpretation, “for example test interferences” (493.1291(e)).
- Specimen rejection criteria should reflect interference limits (493.1242(a)(7)).
How Sound Lab Consultants helps
- Build cross-reactivity and interferent panels for your method
- Set acceptance limits and write the limitations section
- Turn findings into specimen rejection rules
Talk through your test with a lab regulatory consultant.
Frequently asked questions
What is analytical specificity?
The extent to which a method measures only the analyte it reports, without cross-reactivity or interference, per CMS interpretive guidance for 493.1253(b)(2)(iv).
Do I have to test every interferent myself?
No. CMS allows interference information from product information, literature or the lab's own testing. For an LDT, literature must match your method and matrix closely enough to be meaningful.
How many samples per cross-reactant?
CLIA sets no number. APHL's toolkit recommends three to five samples for each potentially cross-reactive organism or substance.
What is the difference between analytical and clinical specificity?
Analytical specificity concerns cross-reactivity and interference. Clinical specificity is the share of people without the condition who test negative.
What if interference cannot be eliminated?
Document it as a limitation in the procedure, set specimen rejection or comment rules, and make the information available to clients as 493.1291(e) requires.
Sources
- 42 CFR 493.1253
- CMS SOM Appendix C (PDF)
- CAP Test Method Validation and Verification excerpt (PDF)
- APHL V&V Toolkit: Qualitative Assays (PDF)
- 42 CFR 493.1251
- 42 CFR 493.1291
- 42 CFR 493.1242
- 42 CFR 493.1256
- Federal Register 2026-14358
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.