LDT Design Control and Change Control
Short answer: LDT design control is the documented way a lab turns an intended use into a locked test design: design inputs, outputs, reviews, verification, validation and transfer to routine use. Change control then assesses every later change for impact and triggers revalidation under 42 CFR 493.1253(b)(2) when performance could be affected. CLIA requires the documentation.
Last reviewed: November 18, 2026. Primary sources: 42 CFR 493.1253, CMS SOM Appendix C, NIH SEED LDT guide, FDA QMSR and ISO 15189:2022.

Why design control matters for LDTs after 2025
FDA's device quality system rules, now the Quality Management System Regulation that incorporates ISO 13485 and took effect February 2, 2026, apply to device manufacturers (FDA QMSR). After the March 31, 2025 vacatur of FDA's LDT rule, most LDTs are outside FDA's purview (CRS), so labs are not required to follow QMSR design controls for LDTs. CLIA still requires documented establishment of performance specifications for any new or modified test (493.1253(b)(2)), and CAP requires a quality management system (GEN.13806). Design-control practices are the cleanest way to produce that evidence. Labs planning an FDA submission or selling kits should follow QMSR.
Design-control elements adapted for a CLIA lab
| Element | What it means for an LDT | Record |
|---|---|---|
| Design plan | Who does what, gates and approvals | Project plan signed by director or designee |
| Design inputs | Intended use, specimen types, population, analytical goals, risk level (NIH SEED) | Intended-use and requirements document |
| Risk assessment | Failure modes in pre-analytic, analytic and post-analytic steps | Risk table; mitigations |
| Design outputs | Reagent formulations, primer/probe or antibody specs, cutoffs, software and pipeline versions, procedure draft | Locked specification |
| Design review | Formal check that outputs meet inputs before validation | Signed review minutes |
| Design verification / analytical validation | Studies under 493.1253(b)(2) | Validation report |
| Clinical claims | Literature or studies for any clinical claim (CAP COM.40640) | Clinical validity file |
| Design transfer | Procedure, QC plan, training, competency, LIS build, report wording | 493.1251 procedure; training records |
| Design changes | Change requests, impact assessment, revalidation | Change records |
Change control: how to classify a change
| Change type | Examples | Typical action |
|---|---|---|
| No impact on performance | Supplier name change for identical catalog item, SOP formatting | Document assessment; no study |
| Verify | New lot of a critical reagent, instrument repair, software patch with no analytic change | Lot-to-lot or comparison check against acceptance criteria |
| Revalidate (partial) | New specimen type or collection device, new extraction kit, cutoff change | Repeat affected 493.1253(b)(2) studies |
| Revalidate (full) / new test | New target, new technology, new intended use or claim | Treat as a new LDT |
CMS guidance treats changes such as new specimen types or modified procedures as modifications requiring establishment of performance specifications (SOM Appendix C). Detailed triggers: when to revalidate an LDT; QMS structure: LDT quality management system.
What goes in an LDT design history file
- Intended use, design inputs and risk assessment.
- Optimization summaries explaining design choices.
- Locked design outputs with version numbers.
- Design review sign-offs.
- Validation plan, raw data, analysis and director-signed report (template).
- Transfer records: procedure, QC plan, training and competency.
- Change log with impact assessments and any revalidation.
Building, validating or fixing an LDT? We help CLIA labs plan and document LDT development and validation and prepare for CLIA, CAP and New York State review.
Where ISO 15189 fits
ISO 15189:2022 sets requirements for quality and competence in medical laboratories, including selection, verification and validation of examination methods (ISO). Labs accredited to ISO 15189, including through CAP's ISO 15189 program, apply its document and change control to LDTs on top of CLIA.
Software, algorithms and bioinformatics
CRS notes LDTs increasingly use AI software (CRS). ADLM's 2026 comments to CMS listed software version control, reproducibility, decision thresholds, change control and post-update monitoring as relevant characteristics (ADLM RFI response). Version-lock pipelines and algorithms and revalidate after updates; see bioinformatics pipeline validation.
How Sound Lab Consultants helps
- Set up a right-sized design-control and change-control process for your LDT menu.
- Build the design history file and change log templates.
- Run change impact assessments and decide verify vs revalidate.
- Prepare for CAP or ISO 15189 assessment of LDT records.
Talk through your test with an LDT regulatory consultant. Bring your intended use, platform and target launch date.
Frequently asked questions
Do LDTs require FDA design controls?
Not for most LDTs after the March 31, 2025 vacatur of FDA's LDT rule. FDA's QMSR design controls apply to device manufacturers. CLIA still requires documented validation, and many labs use design-control practices to meet it.
What is LDT change control?
A documented process to assess every change to an LDT, such as reagents, specimen types, instruments, software or cutoffs, decide whether it affects performance, and verify or revalidate before reporting results.
When does a change require revalidation of an LDT?
When it could affect performance: new specimen types or collection devices, new extraction methods, cutoff changes, new targets or a new intended use. Minor changes may need only verification.
What is a design history file for an LDT?
The collected record of the test's design: intended use, inputs, risk assessment, outputs, reviews, validation, transfer and changes. CLIA does not use the term, but the records support 493.1253 compliance.
Does ISO 15189 apply to LDTs?
For labs accredited to ISO 15189, yes. It covers selection, verification and validation of examination methods along with document and change control.
Sources
- 42 CFR 493.1253
- 42 CFR 493.1251
- CMS SOM Appendix C (PDF)
- NIH SEED regulatory knowledge guide (PDF)
- FDA Quality Management System Regulation
- CRS In Focus IF11389
- ISO 15189:2022
- ADLM response to CMS CLIA RFI (2026, PDF)
- CAP test method validation excerpt (PDF)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.