CLIA Requirements for Laboratory Developed Tests (LDTs)
Short answer: Under CLIA, a laboratory developed test (LDT) is high complexity. The lab needs a CLIA certificate covering high complexity testing, qualified high complexity personnel, and documented establishment of accuracy, precision, analytical sensitivity, specificity, reportable range and reference intervals under 42 CFR 493.1253(b)(2) before reporting results, plus QC, PT or alternative assessment and records.
Last reviewed: November 18, 2026. Primary sources: CMS LDT and CLIA FAQs, 42 CFR 493.1253, 42 CFR 493.25 and CMS State Operations Manual Appendix C.

CLIA requirements for LDTs at a glance
| Requirement | What CLIA requires | Citation |
|---|---|---|
| Certificate | Certificate of compliance or accreditation covering high complexity testing in the specialty | CMS LDT FAQs; 493.25 |
| Complexity | LDTs are considered high complexity tests | CMS LDT FAQs |
| Performance specifications | Establish accuracy, precision, analytical sensitivity, analytical specificity incl. interfering substances, reportable range, reference intervals and other needed characteristics before reporting | 493.1253(b)(2) |
| Calibration and QC | Calibration and control procedures based on the established specifications; QC under 493.1256 | 493.1253(b)(3); 493.1256 |
| Documentation | Document all activities under 493.1253; keep records while the test is in use and at least 2 years | 493.1253(c); 493.1105(a)(3)(i) |
| Procedure manual | Written procedure approved, signed and dated by the director before use | 493.1251 |
| Proficiency testing | Enroll where CMS-approved PT exists; otherwise verify accuracy at least twice annually | 493.1236(c) |
| Calibration verification | At least every 6 months and when specified events occur | 493.1255 |
| Comparison of results | When a test is run on more than one method or site, evaluate the relationship twice a year | 493.1281 |
| Personnel | High complexity director, technical supervisor, clinical consultant, general supervisor and testing personnel | Subpart M (493.1441–493.1495) |
| Test reports | Required report elements | 493.1291 |
| Survey | Analytical validation is reviewed during the routine biennial survey, after testing has started | CMS LDT FAQs |
Why every LDT is high complexity
CMS states that “LDTs are considered high complexity tests. Therefore, the laboratory must meet all applicable CLIA requirements for high complexity testing” (CMS). The high complexity laboratory standard in 42 CFR 493.25 pulls in the facility, PT, QC, personnel and quality system subparts. CMS guidance also says a modified moderate complexity test becomes high complexity (SOM Appendix C; modified FDA-cleared tests).
Establishing performance specifications (493.1253(b)(2))
CMS says CLIA “prohibits the release of any test results prior to the laboratory establishing certain performance characteristics relating to analytical validity” and that this validation is “limited to the specific conditions, staff, equipment and patient population of the particular laboratory” (CMS). The rule applies when a lab introduces a test it developed, modifies an FDA-cleared or approved test, or uses a test system with no manufacturer performance specifications (493.1253(b)(2)). Study-by-study guidance: LDT validation requirements.
What CLIA does not cover: clinical validity
CMS states its CLIA program “does not address the clinical validity of any test” (CMS). CAP and New York do look at clinical claims, so plan for them if you are accredited or test New York specimens (clinical validity; CAP; NYS CLEP).
Building, validating or fixing an LDT? We help CLIA labs plan and document LDT development and validation and prepare for CLIA, CAP and New York State review.
PT and alternative assessment for LDTs
Most LDTs have no CMS-approved PT program. In that case the lab must verify the accuracy of the test “at least twice annually” (493.1236(c)). Options include split-sample comparison with another lab, testing assayed material, or chart review where appropriate. See LDT PT and alternative assessment.
What surveyors ask for
- The LDT validation plan, raw data, statistics and the signed summary.
- Evidence the study came before the first patient report.
- QC and calibration procedures derived from the data.
- The approved procedure manual entry.
- Alternative assessment records twice a year.
- Change records and revalidation for modifications since the last survey (revalidation).
Could CLIA's LDT rules change?
On July 16, 2026, CMS and CDC published a CLIA request for information (CMS-3485-NC) asking how labs establish performance specifications for tests that are not FDA-cleared or approved, which methods (for example toxicology and NGS) have characteristics not addressed in current rules, and how labs prepare in-house reagents. Comments closed September 14, 2026 (91 FR 43586). An RFI is not a rule; current 493.1253 requirements apply until a final rule changes them.
How Sound Lab Consultants helps
- Gap review of your certificate, specialties and personnel for high complexity LDT testing.
- Validation plan and report review against 493.1253(b)(2).
- Alternative assessment program design under 493.1236(c).
- Mock LDT file review before a CLIA survey.
Talk through your test with an LDT regulatory consultant. Bring your intended use, platform and target launch date.
Frequently asked questions
What CLIA certificate do I need to run an LDT?
A CLIA certificate of compliance or certificate of accreditation that covers high complexity testing in the relevant specialty. A certificate of waiver or PPM certificate does not allow LDT testing.
What does CLIA require to validate an LDT?
Before reporting patient results, the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals and any other needed performance characteristic under 42 CFR 493.1253(b)(2).
Does CLIA require clinical validation of LDTs?
No. CMS states its CLIA program does not address clinical validity. CAP and New York State review clinical claims for labs they oversee.
How long must LDT validation records be kept?
While the test is in use and for at least 2 years, under 42 CFR 493.1105(a)(3)(i). CAP-accredited labs keep them while in use plus 2 years after retirement.
What if there is no proficiency testing for my LDT?
Verify the accuracy of the test at least twice annually under 42 CFR 493.1236(c), using methods such as split-sample exchange with another laboratory.
Did the 2025 FDA ruling change CLIA requirements for LDTs?
No. The March 31, 2025 ruling vacated FDA's LDT rule. CLIA requirements in 42 CFR Part 493 were not changed.
Sources
- CMS: LDTs and CLIA FAQs (PDF)
- 42 CFR 493.25
- 42 CFR 493.1253
- 42 CFR 493.1256
- 42 CFR 493.1255
- 42 CFR 493.1251
- 42 CFR 493.1236
- 42 CFR 493.1281
- 42 CFR 493.1291
- 42 CFR 493.1105
- 42 CFR Part 493 Subpart M
- CMS SOM Appendix C (PDF)
- 91 FR 43586 CLIA RFI
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.