CLIA Requirements for Laboratory Developed Tests (LDTs)

Under CLIA, every laboratory developed test is high complexity. To run one, a lab needs a CLIA certificate that covers high complexity testing and qualified high complexity personnel. Before it reports any result, it must also document that it has established the test's performance specifications under 42 CFR 493.1253(b)(2). After that come QC, proficiency testing or an alternative assessment, and records.

The CLIA checklist for an LDT

1. Certificate. A certificate of compliance or certificate of accreditation that covers high complexity testing in the specialty. A certificate of waiver or a PPM certificate doesn't allow LDTs.

2. Personnel. A laboratory director qualified under 493.1443, a technical supervisor under 493.1449, a clinical consultant under 493.1455, general supervisors, and testing personnel under 493.1489.

3. Establishing performance specifications (493.1253(b)(2)). Accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals and any other characteristic the test needs. All of it is documented before patient testing.

4. Calibration and QC. Calibration and control procedures based on the validation data (493.1253(b)(3)), plus QC under 493.1256.

5. Proficiency testing or an alternative assessment. Where no PT program exists, an alternative assessment at least twice a year (493.1236(c)).

6. Records. Keep the performance specification records for as long as the test system is in use, and no less than 2 years (493.1105(a)(3)(i)).

7. Survey. CMS surveyors review LDT validation during the routine biennial survey, which happens after testing has started.

What CLIA does not cover

CLIA does not address clinical validity, meaning how well a test identifies a clinical condition (CMS LDT and CLIA FAQs). CAP-accredited labs must also meet CAP's LDT checklist items, and labs testing New York specimens need NYS CLEP test approval.

Could CLIA's LDT rules change?

CMS and CDC published a CLIA request for information on July 16, 2026 (91 FR 43586) asking how labs establish performance specifications for tests that aren't FDA-cleared. Comments closed September 14, 2026. An RFI is not a rule, so the current 493.1253 requirements apply until a final rule changes them.

Frequently asked questions

What CLIA certificate do I need to run an LDT?

A certificate of compliance or certificate of accreditation that covers high complexity testing in the relevant specialty.

What does CLIA require to validate an LDT?

Before reporting patient results, the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals and any other needed characteristic under 42 CFR 493.1253(b)(2).

Does CLIA require clinical validation of LDTs?

No. CMS states that its CLIA program does not address clinical validity. CAP and New York State review clinical claims for the labs they oversee.

Sources: 42 CFR 493.1253 (https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-K/section-493.1253) · 42 CFR 493.1236 (https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-K/section-493.1236) · 42 CFR 493.1105 (https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-J/section-493.1105) · CMS LDT and CLIA FAQs (https://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/ldt-and-clia_faqs.pdf) · 91 FR 43586 (https://www.federalregister.gov/documents/2026/07/16/2026-14358/request-for-information-clinical-laboratory-improvement-amendments-of-1988-clia-regulations)

Related: What is a laboratory developed test? (/what-is-a-laboratory-developed-test) · LDT validation (/ldt-validation) · CAP LDT requirements (/cap-ldt-requirements)

Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225. Independent consulting firm, not affiliated with FDA, CMS, CDC, CAP, the New York State Department of Health or any accrediting organization. General regulatory information, not legal advice.

Gloved hand reviewing a laboratory validation binder beside a pipette and sample rack to check CLIA
Gloved hand reviewing a laboratory validation binder beside a pipette and sample rack to check CLIA

CLIA Requirements for Laboratory Developed Tests (LDTs)

Short answer: Under CLIA, a laboratory developed test (LDT) is high complexity. The lab needs a CLIA certificate covering high complexity testing, qualified high complexity personnel, and documented establishment of accuracy, precision, analytical sensitivity, specificity, reportable range and reference intervals under 42 CFR 493.1253(b)(2) before reporting results, plus QC, PT or alternative assessment and records.

Last reviewed: November 18, 2026. Primary sources: CMS LDT and CLIA FAQs, 42 CFR 493.1253, 42 CFR 493.25 and CMS State Operations Manual Appendix C.

Book a free consultation

Gloved hand reviewing a laboratory validation binder beside a pipette and sample rack to check CLIA requirements for an LDT
CLIA reviews LDT validation records during the routine survey.

CLIA requirements for LDTs at a glance

CLIA requirements that apply to every LDT
RequirementWhat CLIA requiresCitation
CertificateCertificate of compliance or accreditation covering high complexity testing in the specialtyCMS LDT FAQs; 493.25
ComplexityLDTs are considered high complexity testsCMS LDT FAQs
Performance specificationsEstablish accuracy, precision, analytical sensitivity, analytical specificity incl. interfering substances, reportable range, reference intervals and other needed characteristics before reporting493.1253(b)(2)
Calibration and QCCalibration and control procedures based on the established specifications; QC under 493.1256493.1253(b)(3); 493.1256
DocumentationDocument all activities under 493.1253; keep records while the test is in use and at least 2 years493.1253(c); 493.1105(a)(3)(i)
Procedure manualWritten procedure approved, signed and dated by the director before use493.1251
Proficiency testingEnroll where CMS-approved PT exists; otherwise verify accuracy at least twice annually493.1236(c)
Calibration verificationAt least every 6 months and when specified events occur493.1255
Comparison of resultsWhen a test is run on more than one method or site, evaluate the relationship twice a year493.1281
PersonnelHigh complexity director, technical supervisor, clinical consultant, general supervisor and testing personnelSubpart M (493.1441–493.1495)
Test reportsRequired report elements493.1291
SurveyAnalytical validation is reviewed during the routine biennial survey, after testing has startedCMS LDT FAQs

Why every LDT is high complexity

CMS states that “LDTs are considered high complexity tests. Therefore, the laboratory must meet all applicable CLIA requirements for high complexity testing” (CMS). The high complexity laboratory standard in 42 CFR 493.25 pulls in the facility, PT, QC, personnel and quality system subparts. CMS guidance also says a modified moderate complexity test becomes high complexity (SOM Appendix C; modified FDA-cleared tests).

Establishing performance specifications (493.1253(b)(2))

CMS says CLIA “prohibits the release of any test results prior to the laboratory establishing certain performance characteristics relating to analytical validity” and that this validation is “limited to the specific conditions, staff, equipment and patient population of the particular laboratory” (CMS). The rule applies when a lab introduces a test it developed, modifies an FDA-cleared or approved test, or uses a test system with no manufacturer performance specifications (493.1253(b)(2)). Study-by-study guidance: LDT validation requirements.

What CLIA does not cover: clinical validity

CMS states its CLIA program “does not address the clinical validity of any test” (CMS). CAP and New York do look at clinical claims, so plan for them if you are accredited or test New York specimens (clinical validity; CAP; NYS CLEP).

Building, validating or fixing an LDT? We help CLIA labs plan and document LDT development and validation and prepare for CLIA, CAP and New York State review.

Book a free consultation

PT and alternative assessment for LDTs

Most LDTs have no CMS-approved PT program. In that case the lab must verify the accuracy of the test “at least twice annually” (493.1236(c)). Options include split-sample comparison with another lab, testing assayed material, or chart review where appropriate. See LDT PT and alternative assessment.

What surveyors ask for

  • The LDT validation plan, raw data, statistics and the signed summary.
  • Evidence the study came before the first patient report.
  • QC and calibration procedures derived from the data.
  • The approved procedure manual entry.
  • Alternative assessment records twice a year.
  • Change records and revalidation for modifications since the last survey (revalidation).

Could CLIA's LDT rules change?

On July 16, 2026, CMS and CDC published a CLIA request for information (CMS-3485-NC) asking how labs establish performance specifications for tests that are not FDA-cleared or approved, which methods (for example toxicology and NGS) have characteristics not addressed in current rules, and how labs prepare in-house reagents. Comments closed September 14, 2026 (91 FR 43586). An RFI is not a rule; current 493.1253 requirements apply until a final rule changes them.

How Sound Lab Consultants helps

  • Gap review of your certificate, specialties and personnel for high complexity LDT testing.
  • Validation plan and report review against 493.1253(b)(2).
  • Alternative assessment program design under 493.1236(c).
  • Mock LDT file review before a CLIA survey.

Talk through your test with an LDT regulatory consultant. Bring your intended use, platform and target launch date.

Book a free consultation

Frequently asked questions

What CLIA certificate do I need to run an LDT?

A CLIA certificate of compliance or certificate of accreditation that covers high complexity testing in the relevant specialty. A certificate of waiver or PPM certificate does not allow LDT testing.

What does CLIA require to validate an LDT?

Before reporting patient results, the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals and any other needed performance characteristic under 42 CFR 493.1253(b)(2).

Does CLIA require clinical validation of LDTs?

No. CMS states its CLIA program does not address clinical validity. CAP and New York State review clinical claims for labs they oversee.

How long must LDT validation records be kept?

While the test is in use and for at least 2 years, under 42 CFR 493.1105(a)(3)(i). CAP-accredited labs keep them while in use plus 2 years after retirement.

What if there is no proficiency testing for my LDT?

Verify the accuracy of the test at least twice annually under 42 CFR 493.1236(c), using methods such as split-sample exchange with another laboratory.

Did the 2025 FDA ruling change CLIA requirements for LDTs?

No. The March 31, 2025 ruling vacated FDA's LDT rule. CLIA requirements in 42 CFR Part 493 were not changed.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.