Reportable Range and Reference Intervals for LDTs

Short answer: Reportable range is the span of results an LDT can report accurately, and reference intervals are the values expected in a healthy population. Labs must establish both for an LDT under 42 CFR 493.1253(b)(2)(v)–(vi), check the range at least every 6 months through calibration verification (493.1255), and make intervals available to ordering clinicians (493.1291(d)).

Last reviewed: November 10, 2026. Primary sources: 42 CFR 493.1253(b)(2)(v)–(vi), 493.1255, SOM Appendix C and CAP COM.40600 and COM.50000.

Book a free consultation

Reportable range and AMR

CMS: “The laboratory is responsible for establishing the upper and lower limits of the test system” (SOM Appendix C). CAP distinguishes the analytical measurement range (AMR), values the method measures directly without dilution or concentration, from the reportable range, the values the lab reports, which can extend through validated dilution or be narrowed to clinically relevant limits (COM.40600, CAP excerpt).

  • Test samples spanning low to high, including near the LoD and the top of the curve.
  • Validate any dilution protocol, including the diluent for each specimen type (CMS survey probes).
  • Results above the range are reported as greater than the upper limit, re-assayed after dilution, or sent to a reference lab (SOM).

Keeping the range valid: calibration verification

42 CFR 493.1255(b) requires calibration verification with at least a minimal (or zero), mid-point and maximum value, at least every 6 months and after a complete reagent change (unless shown not to affect the range), major maintenance or critical part replacement, or QC trends that other fixes do not resolve. For an LDT, the lab sets the number, type and concentration of materials and acceptable limits from its validation (493.1253(b)(3)).

Establishing reference intervals

CMS: the lab “must establish a reference range that is appropriate for the laboratory's patient population (i.e., a normal range that reflects the type of specimen and demographic variables such as age and sex, as applicable)” (SOM).

ApproachWhen it fitsSource
Full reference interval study in healthy individualsNew analyte, new matrix, or no suitable published intervalCAP COM.50000 cites CLSI EP28-A3c
Transfer and verify a published or manufacturer intervalComparable method and population; CAP example: test 20 healthy representative individuals, interval verified if no more than 2 fall outsideCAP COM.50000
Documented literature evaluationFormal study not practical (e.g., therapeutic drugs)CAP COM.50000
Expected result statement (qualitative tests)Results are positive/negative; the expected healthy result is stated in the plan, report and SOPAPHL toolkit

Re-evaluate intervals when you add an analyte, change methodology or your patient population changes (CAP COM.50100). Body fluids without established intervals need a report comment (CAP COM.40620).

Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.

Book a free consultation

Reporting

Reference intervals determined by the performing lab must be available to the person who ordered the test (493.1291(d)), and the procedure manual must include the reportable range and reference intervals (493.1251(b)(6), (b)(10)).

How Sound Lab Consultants helps

  • Design linearity and dilution studies
  • Choose between establishing and transferring reference intervals
  • Write calibration verification procedures for LDTs

Talk through your test with a lab regulatory consultant.

Book a free consultation

Frequently asked questions

What is the difference between reportable range and AMR?

CAP describes the AMR as what the method measures directly without dilution or concentration, and the reportable range as the values the lab reports, which may extend through validated dilution or be narrower for clinical reasons.

How often must the reportable range be verified?

Calibration verification under 42 CFR 493.1255(b)(3) is required at least every 6 months and after events such as complete reagent changes, major maintenance or unresolved QC shifts.

How many samples verify a reference interval?

CAP's published example tests 20 healthy representative individuals; if no more than two fall outside the proposed interval, it is considered verified, citing CLSI EP28-A3c.

Does a qualitative LDT need a reference interval?

It needs a stated expected result, such as negative or not detected, documented in the validation plan, report and procedure, as APHL's toolkit describes.

Can I use literature reference intervals for an LDT?

Yes, if you evaluate them for your method and population and keep a record of that evaluation, per CAP COM.50000.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.