LDT Quality Management System and Change Control

LDT quality management is the set of systems that keep a laboratory developed test performing as validated after it goes live: quality control, calibration and calibration verification, proficiency testing or alternative assessment, personnel competency, document control, corrective action and change control. CLIA's quality system requirements for nonwaived testing are in 42 CFR Part 493 Subpart K, and they apply to LDTs the same way they apply to any other test. Change control is the part that matters most for LDTs, because the laboratory itself owns the design and every change to reagents, instruments, software or procedure can affect the validated performance. Sound Lab Consultants works with labs to build LDT quality plans and change control procedures that hold up at survey.

Last reviewed: Oct 30, 2026

Book a free consultation

The CLIA rules behind LDT quality management

ElementCLIA sectionKey point
Performance specifications493.1253Validation sets the baseline; calibration and control procedures are set from it
Calibration and calibration verification493.1255Calibration verification at least once every 6 months and when specified events occur, where applicable
Control procedures493.1256QC appropriate to the test, documented and reviewed
Proficiency testing for unregulated analytes493.1236(c)Verify accuracy at least twice annually for tests without required PT
Competency493.1235Policies to assess employee competency
Corrective action493.1282Documented corrective action when problems are found
Records493.1105Retention of procedures, QC and test records
Procedure manual493.1251Written procedures for each test

Ongoing oversight checklist

  • QC procedures set from the validation results (493.1253(b)(3)) and reviewed on a defined schedule.
  • Calibration and calibration verification schedules where the method requires them.
  • Proficiency testing enrollment, or a documented alternative assessment at least twice a year.
  • Personnel training and competency assessment for the test.
  • Document control for procedures, forms and software versions.
  • Review of QC trends, PT results, complaints and nonconformances with corrective action.
  • Periodic director review of the test, its performance and its literature.
  • A change control log for every LDT.

Change control for LDTs

Any change to a validated LDT should be assessed before it is used for patient testing. A good change record states the change, the reason, the performance characteristics it could affect, the studies needed, the acceptance criteria, the results and the approval.

ChangeTypical assessment
New reagent lotLot-to-lot comparison against defined criteria
New reagent supplier or critical componentRisk assessment; often partial revalidation of affected characteristics
New instrument or relocationComparison and precision checks; calibration verification
Software or bioinformatics pipeline versionRe-run a reference data set; revalidate affected outputs
New specimen type or collection deviceTreat as a new claim and validate under 493.1253(b)(2)
Cutoff or algorithm changeRevalidate affected characteristics and update reports
Procedure wording onlyDocument control; confirm no performance impact

Records to keep

  • Validation file and summary with director approval.
  • QC records and QC review sign-offs.
  • Calibration and calibration verification records.
  • PT results or alternative assessment records with any investigation.
  • Competency assessments.
  • Corrective action records.
  • Change control log and every revalidation study.
  • Superseded procedures, kept at least 2 years after discontinuation (493.1105).

Common findings, framed generally

  • A reagent supplier or software version changed with no record of assessment.
  • Alternative assessment done once a year instead of at least twice.
  • QC rules not derived from the validation data.
  • Corrective action closed without evidence the fix worked.
  • Competency records missing for staff who run the LDT.

Worked example: a bioinformatics update

A lab running a sequencing LDT upgrades its variant-calling software. Under change control, the quality lead logs the change, the director identifies accuracy, precision and analytical sensitivity as at risk, and the team re-runs a stored set of previously characterized samples through the new version. Results are compared with the original calls against criteria written before the run. The pipeline goes live only after the director signs the summary, and the procedure and version record are updated. Patient testing continues on the old version until then.

Related pages

See CAP Checklist Requirements for Laboratory Developed Tests, modified FDA-cleared tests, LDT validation and the LDT consulting overview. See also our clinical laboratory consulting, CLIA consultant and full-service laboratory consulting services.

Frequently asked questions

What is LDT change control?

A documented process to assess, test and approve changes to a validated LDT before patient use.

When does an LDT need revalidation?

When a change could affect a performance characteristic, such as a new specimen type, instrument, critical reagent, software version, cutoff or algorithm. The extent depends on the risk.

What CLIA rules cover the LDT quality system?

42 CFR Part 493 Subpart K covers the quality system for nonwaived testing, including 493.1253, 493.1255, 493.1256 and 493.1282.

What if there is no PT program for my LDT?

493.1236(c) requires the lab to verify accuracy at least twice annually for tests without required PT, for example by split-sample or blinded sample testing.

How often is calibration verification required?

Where it applies, 493.1255 requires calibration verification at least once every 6 months and whenever specified events occur.

How long must records be kept?

493.1105 sets minimum retention periods, including at least 2 years after a procedure is discontinued. Accreditors and states may require longer.

Who signs off on LDT changes?

The laboratory director or delegated qualified personnel, per your quality plan.

Can SLC write our LDT change control procedure?

Yes. SLC drafts and reviews quality and change control procedures for your lab to adopt.

Sources

Related LDT pages

Book a free consultation

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.