Clinical Validity for Laboratory Developed Tests
Short answer: Clinical validity is how accurately a test identifies or predicts a clinical condition; analytical validity is whether it measures the analyte correctly. CMS says CLIA does not address clinical validity. CAP requires labs to validate any clinical claims they make for an LDT, and MolDX reviews clinical validity and clinical utility for Medicare coverage.
Last reviewed: November 10, 2026. Primary sources: CMS LDT and CLIA FAQs, CAP COM.40640 and LCD L35025.
Three kinds of evidence
| Term | Question it answers | Source |
|---|---|---|
| Analytical validity | Does the test measure the analyte accurately and reliably in this lab? | CMS LDT FAQs; 493.1253(b)(2) |
| Clinical validity | “The accuracy with which the test identifies, measures, or predicts the presence or absence of a clinical condition or predisposition in a patient” | CMS LDT FAQs |
| Clinical utility | Does using the result improve patient management or outcomes? | MolDX LCD L35025 requires it for coverage |
What CLIA does and does not cover
CMS: “CMS' CLIA program does not address the clinical validity of any test.” It adds that the routine CLIA survey “does not include a review of the clinical validation of a LDT,” while FDA's premarket review includes clinical validity (CMS LDT FAQs). With the 2024 FDA rule vacated, FDA is not reviewing LDTs before use under that rule, so accreditors, states and payers are where clinical evidence is usually examined.
CAP: clinical claims must be validated
CAP COM.40640 requires validation of all clinical claims a lab makes for LDTs, and for FDA-cleared tests when the lab adds claims beyond the manufacturer's. Claims include diagnostic sensitivity and specificity, risk prediction, clinical usefulness or economic value, wherever they appear: reports, websites, test catalogs or advertisements. To support sensitivity, specificity or risk claims, the lab performs a clinical validation study “unless the clinical validity of the test is documented in peer-reviewed literature or textbooks.” The study must include at least 20 samples with positives and negatives; fewer needs a recorded director rationale (CAP excerpt, 2017). Labs are not required to make clinical claims, but any claim made must be validated.
Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.
MolDX and New York
- MolDX: LCD L35025 requires molecular tests to show analytical validity, clinical validity and clinical utility at a level that meets Medicare's reasonable and necessary standard, reviewed through technical assessment (L35025; MolDX evidence guide).
- New York: CLEP reviews LDT validation packages for tests on New York specimens; standard methods are those with “an established record of reliability and clinical validity” (Wadsworth; NYS CLEP guide).
Building clinical validity evidence
- Write a precise intended use: who is tested, why, and what the result means.
- Search peer-reviewed literature for the analyte, method and population; document why it applies to your assay.
- Where literature is thin or your claim is new, run a clinical study against a reference standard or clinical diagnosis.
- Report diagnostic sensitivity, specificity, and predictive values with the population's prevalence (APHL's 2×2 approach).
- Keep marketing and report language within what the evidence supports.
How Sound Lab Consultants helps
- Audit report, website and catalog claims against your evidence
- Plan clinical validation studies
- Assemble MolDX and New York evidence packages
Talk through your test with a lab regulatory consultant.
Frequently asked questions
Does CLIA require clinical validation of an LDT?
No. CMS states the CLIA program does not address the clinical validity of any test, and routine CLIA surveys do not review clinical validation of LDTs.
What is the difference between analytical and clinical validity?
Analytical validity is whether the test measures the analyte accurately and reliably. Clinical validity is how accurately the result identifies or predicts a clinical condition in patients.
Does CAP require clinical validation?
CAP COM.40640 requires labs to validate any clinical claims they make for LDTs, through a clinical study of at least 20 positive and negative samples or peer-reviewed literature.
What is clinical utility?
Whether using the test result improves patient care decisions or outcomes. MolDX LCD L35025 requires it, along with analytical and clinical validity, for coverage of molecular tests.
Sources
- CMS LDT and CLIA FAQs (PDF)
- CAP Test Method Validation and Verification excerpt (PDF)
- CMS LCD L35025
- NYS Wadsworth: Test Approval
- APHL V&V Toolkit: Qualitative Assays (PDF)
- Federal Register 2025-18239
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.