Proficiency Testing and Alternative Assessment for LDTs
Short answer: Most LDTs have no CMS-approved proficiency testing program, so 42 CFR 493.1236(c) requires the lab to verify the test's accuracy at least twice a year another way. CMS lists options including blind testing of samples with known values, other external assessment programs, split samples with another laboratory or method, and reference photomicrographs.
Last reviewed: November 10, 2026. Primary sources: 42 CFR 493.1236, SOM Appendix C and 493.1445(e)(4).
What the regulation says
42 CFR 493.1236(c): “At least twice annually, the laboratory must verify the accuracy of the following: (1) Any test or procedure it performs that is not included in subpart I of this part. (2) Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program.” CMS adds that this includes the accuracy of calculated results, and that “new or emerging technologies” may lack compatible PT (SOM Appendix C).
If your LDT measures a regulated analyte listed in Subpart I and a CMS-approved program offers compatible samples, you must enroll in PT.
Alternative assessment options
| Option (CMS survey probe) | How labs use it | Watch-outs |
|---|---|---|
| Blind testing of materials with known values | Characterized patient samples, spiked samples or reference materials inserted blind | Values must be truly known; keep the key away from testing staff |
| Other external assessment programs | Non-CMS-approved PT or interlaboratory comparison programs | No referral of samples or results before the event cut-off (CMS) |
| Split samples with another laboratory or method | Exchange samples with a lab running the same or a validated method | Choose a comparator with demonstrated accuracy; set agreement criteria |
| Comparison with reference photomicrographs | Morphology-based tests | Use a recognized reference source |
CAP checklists call this alternative performance assessment (CAP cites COM.01500 for it).
PT referral rules still apply
CMS states that a lab may voluntarily enroll in PT for tests not in Subpart I, but “under no circumstances may these PT samples be referred (or results communicated) to another laboratory for any reason prior to the PT testing event cut-off date,” and the referral consequences, loss of certificate and bar on the owner or operator, apply equally (SOM Appendix C). When multiple platforms run an analyte, choose one primary method per analyte for PT challenges. See PT referral guide.
Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.
Building a defensible program
- Write a procedure naming the method, samples, frequency (at least twice a year) and acceptance criteria for each LDT.
- Cover the reportable range and medical decision points, and include calculated results.
- Rotate testing among the staff who run patient samples.
- Review results, investigate failures and record corrective action; CMS says documentation must include review of all unsatisfactory scores (493.1236(d)).
- Have the director or designee review, consistent with 493.1445(e)(4).
How Sound Lab Consultants helps
- Design alternative assessment for each LDT on your menu
- Arrange split-sample exchanges and acceptance criteria
- Fix PT and alternative assessment citations
Talk through your test with a lab regulatory consultant.
Frequently asked questions
Do LDTs require proficiency testing?
Only when the LDT measures an analyte in Subpart I and a CMS-approved program offers compatible samples. Otherwise 42 CFR 493.1236(c) requires verifying accuracy at least twice a year by another method.
What counts as an alternative assessment?
CMS survey probes list blind testing of materials with known values, other external assessment programs, split samples with another lab or method, and comparison with reference photomicrographs.
How often is alternative assessment required?
At least twice annually under 42 CFR 493.1236(c).
Can I send my alternative assessment samples to another lab?
Split-sample exchange is an accepted alternative. But if you enroll in a PT program, even voluntarily, CMS prohibits referring those PT samples or results before the event cut-off date.
What documentation is required?
All PT evaluation and verification activities must be documented under 493.1236(d), including review of unsatisfactory results and corrective action.
Sources
- 42 CFR 493.1236
- 42 CFR 493 Subpart I
- CMS SOM Appendix C (PDF)
- 42 CFR 493.1445
- CAP Test Method Validation and Verification excerpt (PDF)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.