Proficiency Testing and Alternative Assessment for LDTs

Short answer: Most LDTs have no CMS-approved proficiency testing program, so 42 CFR 493.1236(c) requires the lab to verify the test's accuracy at least twice a year another way. CMS lists options including blind testing of samples with known values, other external assessment programs, split samples with another laboratory or method, and reference photomicrographs.

Last reviewed: November 10, 2026. Primary sources: 42 CFR 493.1236, SOM Appendix C and 493.1445(e)(4).

Book a free consultation

What the regulation says

42 CFR 493.1236(c): “At least twice annually, the laboratory must verify the accuracy of the following: (1) Any test or procedure it performs that is not included in subpart I of this part. (2) Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program.” CMS adds that this includes the accuracy of calculated results, and that “new or emerging technologies” may lack compatible PT (SOM Appendix C).

If your LDT measures a regulated analyte listed in Subpart I and a CMS-approved program offers compatible samples, you must enroll in PT.

Alternative assessment options

Option (CMS survey probe)How labs use itWatch-outs
Blind testing of materials with known valuesCharacterized patient samples, spiked samples or reference materials inserted blindValues must be truly known; keep the key away from testing staff
Other external assessment programsNon-CMS-approved PT or interlaboratory comparison programsNo referral of samples or results before the event cut-off (CMS)
Split samples with another laboratory or methodExchange samples with a lab running the same or a validated methodChoose a comparator with demonstrated accuracy; set agreement criteria
Comparison with reference photomicrographsMorphology-based testsUse a recognized reference source

CAP checklists call this alternative performance assessment (CAP cites COM.01500 for it).

PT referral rules still apply

CMS states that a lab may voluntarily enroll in PT for tests not in Subpart I, but “under no circumstances may these PT samples be referred (or results communicated) to another laboratory for any reason prior to the PT testing event cut-off date,” and the referral consequences, loss of certificate and bar on the owner or operator, apply equally (SOM Appendix C). When multiple platforms run an analyte, choose one primary method per analyte for PT challenges. See PT referral guide.

Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.

Book a free consultation

Building a defensible program

  1. Write a procedure naming the method, samples, frequency (at least twice a year) and acceptance criteria for each LDT.
  2. Cover the reportable range and medical decision points, and include calculated results.
  3. Rotate testing among the staff who run patient samples.
  4. Review results, investigate failures and record corrective action; CMS says documentation must include review of all unsatisfactory scores (493.1236(d)).
  5. Have the director or designee review, consistent with 493.1445(e)(4).

How Sound Lab Consultants helps

  • Design alternative assessment for each LDT on your menu
  • Arrange split-sample exchanges and acceptance criteria
  • Fix PT and alternative assessment citations

Talk through your test with a lab regulatory consultant.

Book a free consultation

Frequently asked questions

Do LDTs require proficiency testing?

Only when the LDT measures an analyte in Subpart I and a CMS-approved program offers compatible samples. Otherwise 42 CFR 493.1236(c) requires verifying accuracy at least twice a year by another method.

What counts as an alternative assessment?

CMS survey probes list blind testing of materials with known values, other external assessment programs, split samples with another lab or method, and comparison with reference photomicrographs.

How often is alternative assessment required?

At least twice annually under 42 CFR 493.1236(c).

Can I send my alternative assessment samples to another lab?

Split-sample exchange is an accepted alternative. But if you enroll in a PT program, even voluntarily, CMS prohibits referring those PT samples or results before the event cut-off date.

What documentation is required?

All PT evaluation and verification activities must be documented under 493.1236(d), including review of unsatisfactory results and corrective action.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.