LDT Report Disclaimer Statement

Short answer: Labs that build an in-house test with analyte specific reagents must add this to the report: “This test was developed and its performance characteristics determined by (Laboratory Name). It has not been cleared or approved by the U.S. Food and Drug Administration” (21 CFR 809.30(e)). CAP also requires LDT reports to state the lab developed the assay.

Last reviewed: November 10, 2026. Primary sources: 21 CFR 809.30(e), CAP COM.40630 and 42 CFR 493.1291.

Book a free consultation

The FDA statement (ASR-based tests)

“The laboratory that develops an in-house test using the ASR shall inform the ordering person of the test result by appending to the test report the statement: ‘This test was developed and its performance characteristics determined by (Laboratory Name). It has not been cleared or approved by the U.S. Food and Drug Administration.’” The statement does not apply to results from the cleared or approved test reviewed with a class II or III ASR (21 CFR 809.30(e)).

CAP's LDT reporting requirement

CAP COM.40630 requires reports for LDTs to contain a statement that the lab developed the assay, except for traditional methods such as manual microscopy, conventional cultures, conventional cytogenetics, and manual hematology and immunology tests. CAP gives this commonly used example for US labs: “This test was developed and its performance characteristics determined by <insert laboratory/company name>. It has not been cleared or approved by the FDA. The laboratory is regulated under CLIA as qualified to perform high-complexity testing. This test is used for clinical purposes. It should not be regarded as investigational or for research.” CAP adds that the report should briefly describe the method and performance characteristics needed for clinical use, unless available to clinicians in an equivalent form such as a test catalog (CAP excerpt, 2017).

Other report elements CLIA requires

ElementRule
Patient identification; name and address of the performing lab; report date; test performed; specimen source when appropriate; result and units or interpretation; specimen condition issues493.1291(c)
Reference intervals available to the ordering provider493.1291(d)
Methods list and performance specifications available to clients on request; interference information on request493.1291(e)
Pertinent information required for interpretation493.1445(e)(8)

Building or fixing an LDT? We help labs plan validation studies, write the plan and report, and prepare for CLIA, CAP and state review.

Book a free consultation

Wording tips

  • Use your exact legal laboratory name, matching your CLIA certificate.
  • Do not imply FDA clearance or approval anywhere on the report or in marketing.
  • Avoid research-use language on clinical reports; CAP's example says the test “should not be regarded as investigational or for research.”
  • Keep clinical claims within your evidence (clinical validity).

How Sound Lab Consultants helps

  • Review report templates and LIS comments for LDT statements
  • Align test catalogs and websites with CAP claim rules

Talk through your test with a lab regulatory consultant.

Book a free consultation

Frequently asked questions

What is the LDT disclaimer statement?

For tests built with analyte specific reagents, 21 CFR 809.30(e) requires: This test was developed and its performance characteristics determined by (Laboratory Name). It has not been cleared or approved by the U.S. Food and Drug Administration.

Is the LDT disclaimer required for all LDTs?

The FDA rule applies to in-house tests using ASRs. CAP-accredited labs must include a statement that the lab developed the assay on all LDT reports, except traditional methods such as manual microscopy and conventional cultures.

Can the disclaimer be in a test catalog instead of the report?

The FDA ASR statement is appended to the test report. CAP allows the method description and performance characteristics to be provided in an equivalent form such as a test catalog, but the developed-by statement belongs on the report.

Does the disclaimer mean the test is experimental?

No. CAP's example wording states the test is used for clinical purposes and should not be regarded as investigational or for research.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the College of American Pathologists, the New York State Department of Health or any accrediting organization or Medicare contractor. This page is general regulatory information, not legal advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.