Short answer: Since February 2, 2026, FDA's Quality Management System Regulation (QMSR) in 21 CFR Part 820 incorporates ISO 13485:2016 by reference. IVD manufacturers must run a quality system that meets ISO 13485, including product realization, design controls and risk management, plus remaining FDA provisions. FDA now inspects under Compliance Program 7382.850.
Last reviewed: November 21, 2026. Primary sources: FDA, Quality Management System Regulation (QMSR), FDA, QMSR Frequently Asked Questions, eCFR, 21 CFR Part 820, Quality Management System Regulation, eCFR, 21 CFR Part 807 (registration, listing, premarket notification).
What changed on February 2, 2026
FDA's device good manufacturing practice rule, 21 CFR Part 820, became the Quality Management System Regulation (QMSR) on February 2, 2026. It incorporates ISO 13485:2016 by reference, along with clause 3 of ISO 9000:2015 for terms, and FDA now inspects under Compliance Program 7382.850 instead of the older QSIT approach (FDA, Quality Management System Regulation (QMSR); eCFR, 21 CFR Part 820, Quality Management System Regulation). Where ISO 13485 conflicts with the FD&C Act or FDA regulations, FDA law controls.
QMSR applies to finished device manufacturers that intend to commercially distribute devices, including IVD kits, reagents sold as finished devices and accessories. FDA's QMSR page states that devices made under an IDE are not exempt from design and development requirements (FDA, Quality Management System Regulation (QMSR)). FDA will not require or issue ISO 13485 certificates, and a certificate does not exempt a site from FDA inspection (FDA, QMSR Frequently Asked Questions).
Product realization and design controls for IVDs
ISO 13485 clause 7, product realization, covers planning, customer requirements, design and development, purchasing, production and service provision, and control of monitoring and measuring equipment. For an IVD maker, design and development is where FDA submissions and quality records meet.
| Design control element | IVD example of the record | Ties to the submission |
|---|---|---|
| Design and development planning | Plan with stages, reviews, responsibilities and the regulatory pathway | Pathway and Pre-Sub strategy |
| Design inputs | Intended use, user needs, performance targets (LoD, precision, agreement), specimen types, stability, labeling and regulatory requirements | Intended use and acceptance criteria |
| Design outputs | Reagent formulations, specifications, instrument and software specs, IFU, QC and release criteria | Device description and labeling |
| Design review | Documented reviews at stage gates with independent reviewers | Traceability of decisions |
| Design verification | Analytical studies proving outputs meet inputs | Analytical performance section |
| Design validation | Clinical and usability studies on production-equivalent lots under defined conditions | Clinical performance and human factors sections |
| Design transfer | Manufacturing procedures, lot release, supplier controls | Manufacturing information and lot consistency |
| Design changes | Change control with impact assessment, including regulatory impact | When a change needs a new 510(k) |
| Design and development files | The complete record for each device type | Evidence an FDA investigator can follow |
Risk management is now explicit
FDA notes that incorporating ISO 13485 now specifically requires risk management across the quality system (FDA, Quality Management System Regulation (QMSR)). For IVDs, keep one risk file that links hazards such as false negatives, false positives, delayed results and user error to design controls, labeling, verification data and post-market signals. The same analysis feeds the CLIA waiver risk analysis and De Novo risk-to-mitigation tables.
Moving your quality system to QMSR? Sound Lab Consultants can run the gap assessment with you.
Other FDA rules that sit alongside QMSR
| Requirement | What it covers | Source |
|---|---|---|
| Registration and listing | Establishment registration, device listing, annual fee; U.S. agent for foreign firms | 21 CFR Part 807 |
| IVD labeling | Label and package insert content | 21 CFR 809.10 |
| Medical Device Reporting | Reporting deaths, serious injuries and certain malfunctions | 21 CFR Part 803 |
| Change assessment | Whether a change needs a new 510(k) | FDA change guidance |
Building a QMS that fits a diagnostic company
- Map current procedures to ISO 13485 clauses and the remaining Part 820 provisions; list the gaps.
- Set up document control, records, training and management review first; everything else depends on them.
- Write design control and risk management procedures, then retrofit the design file for products already in development.
- Add supplier controls for critical raw materials such as antibodies, enzymes, primers and plastics.
- Build complaint handling, CAPA, nonconformance and MDR procedures before launch.
- Run an internal audit against the QMSR inspection approach, fix findings, and hold management review.
If you also run a CLIA lab
A company that runs a CLIA lab and also sells kits answers to two systems: CLIA for the lab and QMSR for the products. Keep the boundaries clear in procedures and records. Laboratory quality under ISO 15189 is a different standard from ISO 13485; see ISO 15189 accreditation consulting.
How Sound Lab Consultants helps
- QMSR and ISO 13485 gap assessments for diagnostic companies
- Design control, risk management, CAPA and complaint procedures
- Design history and development file builds for products in development
- Supplier controls for critical reagents and components
- Internal audits against FDA's QMSR inspection approach
Talk about design controls and QMSR with Sound Lab Consultants.
Frequently asked questions
When did the QMSR take effect?
FDA's Quality Management System Regulation took effect on February 2, 2026, amending 21 CFR Part 820 to incorporate ISO 13485:2016 by reference.
Does FDA require ISO 13485 certification?
No. FDA requires compliance with the QMSR, which incorporates ISO 13485 requirements. FDA says it will not require or issue ISO 13485 certificates, and a certificate does not exempt a manufacturer from FDA inspection.
Do design controls apply during development under an IDE?
Yes. FDA states that devices manufactured under an investigational device exemption are not exempt from the QMSR design and development requirements.
What is product realization in ISO 13485?
Product realization is clause 7 of ISO 13485. It covers planning, customer requirements, design and development, purchasing, production and service provision, and control of monitoring and measuring equipment.
Sources
- FDA, Quality Management System Regulation (QMSR)
- FDA, QMSR Frequently Asked Questions
- eCFR, 21 CFR Part 820, Quality Management System Regulation
- eCFR, 21 CFR Part 807 (registration, listing, premarket notification)
- 21 CFR 809.10, labeling for IVD products
- eCFR, 21 CFR Part 803, Medical Device Reporting
- FDA guidance, Deciding When to Submit a 510(k) for a Change to an Existing Device
- FDA, Medical Device Reporting (MDR)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the U.S. Department of Health and Human Services, any state survey agency, the College of American Pathologists, COLA, The Joint Commission, AABB, ACHC, ASHI, A2LA or any other accrediting organization or Medicare contractor. This page is general regulatory information, not legal, billing or medical advice. Confirm current requirements with FDA, CMS or the agency that oversees your product or laboratory. Only FDA decides clearance, authorization, approval, categorization and CLIA waiver; no consultant can guarantee an outcome. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.