Short answer: FDA CLIA categorization assigns each commercially marketed IVD to waived, moderate or high complexity. Non-waived tests are scored on seven criteria under 42 CFR 493.17; a total of 12 or less is moderate complexity and above 12 is high. Categorization follows clearance or a standalone CLIA Record request, and uncategorized tests default to high complexity.
Last reviewed: November 21, 2026. Primary sources: FDA, CLIA Categorizations, 42 CFR 493.17, test categorization, 42 CFR 493.5, categories of tests by complexity, FDA, CLIA test categorization database.
What FDA CLIA categorization is
FDA categorizes commercially marketed IVDs as waived, moderate complexity or high complexity. The category is set after FDA clears, approves or grants the marketing submission, or on request for legally marketed tests that need no new submission (FDA, CLIA Categorizations). It decides which CLIA certificate a lab needs to run the test and therefore who can buy it. Tests that have not been categorized default to high complexity (42 CFR 493.17).
The seven-criterion scorecard
For tests that are not waived, FDA reviews the package insert instructions and scores seven criteria from 1 (lowest) to 3 (highest); a 2 is assigned when a test falls between the descriptions. A total of 12 or less is moderate complexity and above 12 is high complexity (FDA, CLIA Categorizations; 42 CFR 493.17).
| Criterion | Scores 1 when | Scores 3 when |
|---|---|---|
| 1. Knowledge | Minimal scientific knowledge; on-the-job instruction is enough | Specialized scientific and technical knowledge is essential |
| 2. Training and experience | Minimal training and limited experience | Specialized training or substantial experience is needed |
| 3. Reagents and materials preparation | Stable, prepackaged or premeasured | Labile, or manual weighing or volumetric steps |
| 4. Operational steps | Automatic or easily controlled | Close monitoring, precise timing or temperature, accurate pipetting or calculations |
| 5. Calibration, QC and PT materials | Stable and readily available | Labile or not available |
| 6. Troubleshooting and maintenance | Automatic or simple; maintenance by the manufacturer | Requires judgment and special skills |
| 7. Interpretation and judgment | Minimal | Extensive independent interpretation |
Because the score is based on the instructions, labeling decisions during development (pre-measured reagents, automated timing, built-in QC, clear result display) directly affect the category.
Concurrent and standalone CLIA Records
| Type | When | FDA target |
|---|---|---|
| Concurrent CR | Created automatically when CDRH receives a 510(k), De Novo, PMA or HDE for an IVD | Notice within about two weeks of the marketing decision |
| Standalone CR | Trade, manufacturer or distributor name changes; new reagent-instrument combinations under FDA's Replacement Reagent and Instrument Family Policy; 510(k)-exempt IVDs; tests cleared by other FDA centers | Notice within about 30 days of receipt |
Source: FDA, CLIA Categorizations. There is no FDA user fee for a CR. The categorization is effective on the date of FDA's written notice and appears in the FDA CLIA database at the next weekly update.
Want to design toward a lower complexity category? Sound Lab Consultants can review your instructions against the scorecard.
What a standalone CR request includes
- A signed cover letter identifying the submission as a CLIA Record or CLIA categorization request.
- A contact who represents the marketing submission holder or 510(k)-exempt manufacturer, or an authorization letter from that holder for the correspondent.
- The reason for the request (new name, new distributor, RRP combination or exempt test).
- Each test system: manufacturer, trade and distributor names, submission number or product code and regulation, and current instructions for use.
Source: FDA, CLIA Categorizations. For an exempt device, confirm the exemption limits in the classification regulation first; a test outside them needs a premarket submission.
Moving from moderate complexity to waived
A moderate complexity test can be recategorized as waived only through a CLIA Waiver by Application; a new test can seek both at once with a Dual Submission. High complexity tests are not eligible for waiver without design changes that bring the score down first. Lab-side effects of each category are explained on CLIA certificate types and test complexity.
Why manufacturers should care about the lab side
- Moderate complexity tests can be run by labs with a Certificate of Compliance, Accreditation or Registration, with CLIA personnel, QC and proficiency testing rules.
- High complexity adds stricter personnel qualifications, which narrows the customer base.
- Waived tests reach physician offices, pharmacies and clinics, but those sites must still follow the manufacturer's instructions exactly.
- A lab that modifies a cleared or approved test must establish its own performance specifications under 42 CFR 493.1253(b)(2); see modified FDA-cleared tests.
- A request to change a category is reviewed only if it rests on new information, and not more than once a year for the same applicant (42 CFR 493.17(c)(3)).
How Sound Lab Consultants helps
- Scorecard reviews of draft instructions for use during design
- Standalone CLIA Record requests for name changes, new distributors, RRP combinations and exempt tests
- Waiver pathway planning when a test is near the moderate complexity line
- Customer-facing guidance for labs on certificate type and verification
Ask Sound Lab Consultants about categorization and waiver planning.
Frequently asked questions
Who decides the CLIA complexity of a test?
FDA categorizes commercially marketed tests under the CLIA criteria in 42 CFR 493.17, after clearance or approval or on request through a CLIA Record submission.
What score makes a test high complexity?
Each of seven criteria is scored 1, 2 or 3. A total of 12 or less is moderate complexity; a total above 12 is high complexity.
What is a standalone CLIA Record?
It is a request for FDA to categorize a legally marketed test that does not need a new marketing submission, such as a new trade or distributor name, a new reagent and instrument combination under FDA's replacement reagent policy, or a 510(k)-exempt IVD.
Is there a fee for CLIA categorization?
No. FDA does not charge a user fee for a CLIA Record submission.
What complexity applies before a test is categorized?
Under 42 CFR 493.17, a test system that has not been categorized is treated as high complexity until FDA categorizes it.
Sources
- FDA, CLIA Categorizations
- 42 CFR 493.17, test categorization
- 42 CFR 493.5, categories of tests by complexity
- FDA, CLIA test categorization database
- FDA, CLIA Waiver by Application
- FDA, Clinical Laboratory Improvement Amendments (CLIA)
- 42 CFR 493.1253, establishment and verification of performance specifications
- eCFR, 42 CFR Part 493, Laboratory Requirements
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the U.S. Department of Health and Human Services, any state survey agency, the College of American Pathologists, COLA, The Joint Commission, AABB, ACHC, ASHI, A2LA or any other accrediting organization or Medicare contractor. This page is general regulatory information, not legal, billing or medical advice. Confirm current requirements with FDA, CMS or the agency that oversees your product or laboratory. Only FDA decides clearance, authorization, approval, categorization and CLIA waiver; no consultant can guarantee an outcome. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.