CLIA Waiver by Application for IVD Manufacturers

Short answer: A CLIA Waiver by Application (CW) is an FDA submission asking that a cleared or approved moderate complexity test be recategorized as waived. The manufacturer must show the test is simple and has an insignificant risk of an erroneous result, using risk analysis, flex studies and studies with untrained operators at intended use sites. FDA's decision goal is 150 FDA days, with no user fee.

Last reviewed: November 21, 2026. Primary sources: FDA, CLIA Waiver by Application, FDA guidance, Recommendations for CLIA Waiver Applications for Manufacturers of IVDs (Feb 2020, PDF), FDA guidance, Recommendations for Dual 510(k) and CLIA Waiver by Application Studies (Feb 2020), 42 U.S.C. 263a, Certification of laboratories (Cornell LII).

Book a free consultation

What a CLIA Waiver by Application is

Under CLIA, FDA places each commercially marketed test in one of three categories: waived, moderate complexity or high complexity. Tests waived by regulation, and tests cleared or approved for home use, are waived automatically. Everything else is scored as moderate or high complexity (FDA, CLIA Waiver by Application; 42 CFR 493.17). A manufacturer whose test was categorized as moderate complexity can ask FDA to recategorize it as waived by filing a CLIA Waiver by Application (CW).

Waived status matters commercially because waived tests can be run by sites holding a Certificate of Waiver or a PPM certificate, such as physician offices, pharmacies and urgent care clinics, in addition to labs with higher-level certificates (FDA, CLIA Waiver by Application; 42 CFR 493.15).

The CLIA statute allows waiver for tests that are simple and have an insignificant risk of an erroneous result, either because the method makes errors by the user negligible or because an incorrect result would pose no unreasonable risk of harm (42 U.S.C. 263a(d)(3)). FDA's February 2020 waiver guidance turns that into evidence a manufacturer can produce (FDA guidance, Recommendations for CLIA Waiver Applications for Manufacturers of IVDs (Feb 2020, PDF)):

What you must showHow it is usually shownGuidance section
The test is simpleFew steps, no specimen manipulation beyond simple handling, unit-use or automated steps, no technical judgment to read results, simple labelingIII. Demonstrating Simple
Errors are controlled (Tier 1)A systematic risk analysis of every source of user, environmental and device error, plus flex studies that stress operating limitsIV.A Risk Analysis and Flex Studies
Failure alerts and fail-safes workValidation of lockouts, internal controls and error messages under stress conditionsIV. Failure Alerts and Fail-Safe Mechanisms
Accuracy in real handsComparison studies with untrained operators at intended use sites against a comparator, plus reproducibilityV. Accuracy
Labeling is simpleInstructions and a quick reference guide written for the intended waived userVI. Labeling for Waived Devices

Study design basics for a waiver

FDA's guidance describes the clinical study for waiver as one run at a minimum of three intended use sites, with one to three untrained operators per site and at least nine untrained operators overall. Operators should have limited or no laboratory testing experience and should receive only the instructions and training materials intended for waived users (FDA guidance, Recommendations for CLIA Waiver Applications for Manufacturers of IVDs (Feb 2020, PDF)). Comparator testing is done by trained operators or a reference method, depending on the study option chosen.

  • Write the risk analysis first; it decides which flex conditions you test (temperature, humidity, specimen volume, timing, lighting, reading position and similar).
  • Choose specimens that cover the measuring range or both sides of the cutoff, including samples near the cutoff where errors are most likely.
  • Pre-specify acceptance criteria and the statistical plan before enrollment.
  • Plan IRB review and informed consent under 21 CFR Part 50 and Part 56, and confirm whether the IVD exemption in 21 CFR 812.2(c)(3) applies.
  • Ask FDA to comment on protocols through a Pre-Submission before you start.

Planning a waiver for a point-of-care test? Sound Lab Consultants can help design the risk analysis and studies.

Book a free consultation

What to include in the CW submission

  1. A signed cover letter that identifies the submission as a CW, references the cleared or approved marketing submission and names a contact person.
  2. A device description that shows the test is simple to use.
  3. Risk analysis results, including all identified sources of error.
  4. Risk evaluation and control results, with verification and validation of failure alerts and fail-safe mechanisms, including under stress.
  5. Flex study results showing the test is insensitive to environmental and usage variation.
  6. Design and results of the clinical studies with intended users at intended use sites.
  7. Proposed labeling consistent with a simple device.

The list follows FDA's CW page (FDA, CLIA Waiver by Application). CW files go to CDRH's Office of In Vitro Diagnostics through the CDRH Portal or the Document Control Center.

FDA review steps and timeline

StepFDA goalNotes
AcknowledgmentWithin about one weekAssigns a CW number
Substantive Interaction90 FDA daysApproval, proceed interactively, or an Additional Information request
Additional Information responseDue within 180 daysOtherwise the CW is considered withdrawn
MDUFA decision150 FDA days (320 with an advisory panel)Waiver approval or denial; database updated weekly

Source: FDA, CLIA Waiver by Application. There is no FDA user fee for a standalone CW.

Standalone CW or Dual Submission?

If the test is already cleared and you now want waived status, the standalone CW is the route. If the test is new and you want waived status at launch, consider the Dual 510(k) and CLIA Waiver by Application, which uses one set of comparison and reproducibility studies with untrained operators to support both decisions (FDA guidance, Recommendations for Dual 510(k) and CLIA Waiver by Application Studies (Feb 2020)). Labs that will run the test after waiver can find site-level rules on our waived testing compliance and Certificate of Waiver pages.

How Sound Lab Consultants helps

  • Waiver feasibility review of the test design, steps and labeling
  • Risk analysis and flex study protocols aligned with ISO 14971 practice
  • Untrained-operator clinical study protocols, site selection support and statistical plans
  • Quick reference instructions and waived labeling review
  • CW assembly, Pre-Submission questions and response management

Talk with Sound Lab Consultants about your CW strategy.

Book a free consultation

Frequently asked questions

Who can file a CLIA Waiver by Application?

The manufacturer of a test that FDA has cleared or approved and categorized as moderate complexity. The submission references the cleared or approved marketing application.

How many sites and operators does a CLIA waiver study need?

FDA's 2020 waiver guidance recommends at least three intended use sites, one to three untrained operators per site and at least nine untrained operators in total, using only the instructions intended for waived users.

How long does FDA take to decide a CLIA waiver?

FDA aims for a Substantive Interaction within 90 FDA days and a decision within 150 FDA days, or 320 FDA days if an advisory panel is needed.

Is there a user fee for a CLIA waiver?

No. FDA does not charge a user fee for a standalone CLIA Waiver by Application. A Dual 510(k) and CLIA Waiver submission does carry the 510(k) fee.

What are flex studies?

Flex studies stress the test's operating limits, such as temperature, humidity, specimen volume and timing, to show that results stay accurate when users or conditions vary in ways the risk analysis identified.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the U.S. Department of Health and Human Services, any state survey agency, the College of American Pathologists, COLA, The Joint Commission, AABB, ACHC, ASHI, A2LA or any other accrediting organization or Medicare contractor. This page is general regulatory information, not legal, billing or medical advice. Confirm current requirements with FDA, CMS or the agency that oversees your product or laboratory. Only FDA decides clearance, authorization, approval, categorization and CLIA waiver; no consultant can guarantee an outcome. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.