Setting Up a New QC Lot Mean and SD
Last reviewed Nov 1, 2026 · Sound Lab Consultants
To set up a new QC lot, run the new control alongside the current lot (crossover) before the old lot expires, collect at least 20 results over at least 10 operating days, and calculate the new lot's own mean and SD. While data accumulate, use provisional limits: the new lot's early mean with an SD estimated from the established CV of the old lot (SD = old CV% × new mean / 100), then replace them with the 20-point values.
New lot mean and SD calculator
Step-by-step lot change
- Order enough of the new lot to overlap the old lot by several weeks.
- Confirm the new lot's storage, open-vial stability and reconstitution steps.
- Run the new lot alongside the old lot in the same runs; judge acceptability on the old lot only.
- After a few days, calculate a provisional mean; set provisional SD from historical CV.
- At 20 points over at least 10 days, calculate the final mean and SD and update limits.
- Document the crossover data, calculations and approval.
Worked example
Old lot CV = 1.09%. New lot 20 results: mean = 201.925, SD = 1.360, CV = 0.67%.
A higher target on the new lot is expected (different material); what matters is that the SD and CV are consistent with the method's historical imprecision.
Why not use the package insert range?
Manufacturer ranges typically reflect many laboratories and instruments, so they are wider than a single laboratory's performance and miss real shifts. 42 CFR §493.1256(d)(10) allows a stated value for an assayed control only when the laboratory verifies it; most laboratories set their own mean and SD.
Common pitfalls
- Setting final limits from a few days of data (SD too small, false rejections).
- Starting the new lot right after a calibration or reagent change, baking a bias into the mean.
- Changing reagent and control lots at the same time.
- Keeping the old SD when the new lot is at a different concentration (use CV to scale).
After setup, use the QC Mean, SD and CV% Calculator for monthly cumulative updates and the Westgard Rules Calculator (Two Control Levels) for daily review. Unity users: Bio-Rad Unity QC consulting, Unity Real Time Westgard rules setup and building an IQCP from Unity QC data.
Frequently asked questions
How many days should I overlap QC lots?
Long enough to collect at least 20 results on the new lot, preferably across 20 operating days; at minimum 10 days.
What are provisional QC limits?
Temporary limits used while new-lot data accumulate, typically the early mean with an SD derived from the historical CV.
Can I use the old lot SD for the new lot?
If the concentrations are similar the old SD may be close, but scaling by CV (SD = CV × new mean / 100) is more robust.
Should I run the new lot before the old lot expires?
Yes. Crossover lets you collect data while patient results are still judged by the established lot.
When do I recalculate the new lot mean and SD?
After 20 points, then periodically (often monthly) as cumulative statistics.
Does CLIA require a lab-established mean and SD?
CLIA requires statistical parameters for each control lot be defined; assayed control stated values may be used only if verified.
What if the new lot mean is far from the insert value?
Investigate calibration and reagent status first; a valid laboratory mean can differ from the insert mean, but a large difference warrants review.
Want a second set of eyes on your QC rules, limits or survey preparation?
Book a free consultationRelated tools and guides
- Levey-Jennings Charts and Westgard Rules
- QC Mean, SD and CV% Calculator
- Westgard Rules Calculator (Two Control Levels)
- Sigma Metric Calculator for the Clinical Lab
- Total Allowable Error and Bias Calculator
- Corrective Action for Westgard Rule Violations
- Bio-Rad Unity QC consulting
- Unity Real Time Westgard rules setup
- IQCP using Unity QC data
Sources and standards
- 42 CFR §493.1256 Standard: Control procedures (eCFR)
- 42 CFR §493.1282 Standard: Corrective actions (eCFR)
- 42 CFR Part 493, Subpart I, Proficiency testing acceptance limits (eCFR)
- CMS: Clinical Laboratory Improvement Amendments (CLIA)
- CDC: CLIA program
- CLSI C24, Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions (consult the current edition)
Sound Lab Consultants is an independent consulting firm and is not affiliated with Westgard QC, Bio-Rad, CLSI, CMS, CDC or any accrediting organization. Rule names are used descriptively. Calculators are educational aids; confirm results against your laboratory's approved QC procedures.
