Levey-Jennings Chart and Westgard Rules: Guide + Rule Checker

Levey-Jennings Charts and Westgard Rules

Last reviewed Nov 1, 2026 · Sound Lab Consultants

A Levey-Jennings chart plots each quality control (QC) result in run order (x-axis) against its concentration (y-axis), with horizontal lines at the control's established mean and at ±1, ±2 and ±3 standard deviations (SD). Westgard rules are the decision criteria read from that chart: a single point beyond 2SD (1-2s) is a warning, while 1-3s, 2-2s, R-4s, 4-1s and 10x violations mean the run is out of control and patient results should be held until the cause is found and corrected.

What is a Levey-Jennings chart?

The chart is a time-ordered picture of how a measurement procedure performs on a stable control material. Because an in-control process produces results that scatter randomly around a central value, the chart lets a technologist see at a glance whether today's control is behaving like the last month's controls or whether something has changed. It is named for S. Levey and E. R. Jennings, who adapted industrial Shewhart charts to the clinical laboratory in 1950.

Chart structure

  • X-axis: run number, date or time of each control measurement.
  • Y-axis: control value in reporting units, scaled to cover roughly mean ±4SD.
  • Center line: the laboratory's own established mean for that control lot.
  • Limit lines: mean ±1SD, ±2SD and ±3SD.
  • One chart per analyte, per control level, per lot, per instrument.
-3SD-2SD-1SDMean+1SD+2SD+3SD68.3%95.4% inside ±2SD · 99.7% inside ±3SD
About 68.3% of in-control results fall within ±1 SD, 95.4% within ±2 SD and 99.7% within ±3 SD.

How to build a Levey-Jennings chart, step by step

  1. Run the new control lot for at least 20 data points across at least 10 operating days (20 days is better) so normal day-to-day variation is captured. See Setting Up a New QC Lot Mean and SD.
  2. Calculate the mean, SD and CV% (use the QC Mean, SD and CV% Calculator).
  3. Compute the limits: mean ± 1SD, 2SD, 3SD.
  4. Draw the center line and limit lines; label analyte, level, lot, units, instrument and dates.
  5. Plot each new control result as it is run and connect points in order.
  6. Apply the laboratory's selected Westgard rules before releasing patient results.

Worked example: cholesterol control

Twenty results for a control lot (mg/dL): 200, 202, 198, 201, 199, 203, 197, 200, 204, 196, 200, 201, 199, 202, 198, 200, 203, 197, 201, 199.

Mean = Σx / n = 4000 / 20 = 200.0 mg/dL SD = √[Σ(x − mean)² / (n − 1)] = √(90 / 19) = 2.176 mg/dL CV = SD / mean × 100 = 1.09% ±1SD: 197.82 to 202.18 ±2SD: 195.65 to 204.35 ±3SD: 193.47 to 206.53

A next-day result of 205 mg/dL gives z = (205 − 200) / 2.176 = 2.30, beyond +2SD but inside +3SD: a 1-2s warning. If the following run is 206 (z = 2.76), the two consecutive points beyond +2SD violate 2-2s and the run is rejected.

Westgard rules explained

Each rule is written as NL: how many control results (N) must exceed which limit (L). Rules are combined into a multirule so the false-rejection rate stays low while error detection stays high.

RuleMeaningError it signalsUseExample
1-2sOne control result outside mean ±2SDRandom or systematic (screen only)Warning in classic multirule; ~5% of in-control points exceed it by chance
1-3sOne result outside mean ±3SDRandom error (or a large shift)Reject
2-2sTwo consecutive results beyond the same 2SD limit (same level across runs, or both levels in one run)Systematic errorReject
R-4sWithin one run, one result above +2SD and another below -2SD (range over 4SD)Random errorReject; apply within a run only
4-1sFour consecutive results beyond the same 1SD limitSystematic error (shift)Reject or warn, by sigma
2of3-2sTwo of three results beyond the same 2SD limitSystematic errorReject, 3-level QC
3-1sThree consecutive results beyond the same 1SD limitSystematic errorReject, 3-level QC
6x / 8x / 9x / 10x / 12x6, 8, 9, 10 or 12 consecutive results on one side of the meanSystematic error (bias)Reject; choose N to fit control count (6x/9x for 3 levels; 8x/10x/12x for 2 or 4)

Random error vs systematic error

Random errorSystematic error
What it looks likeWider scatter; isolated far-out points; levels move in opposite directionsA shift or trend; points pile up on one side of the mean
Rules that catch it1-3s, R-4s2-2s, 4-1s, 2of3-2s, 3-1s, 6x to 12x
Typical causesBubbles, clots, pipetting or sampling faults, unstable power or temperature, poorly mixed reagent, wrong control reconstitutionCalibration change, new reagent or calibrator lot, deteriorating reagent, lamp aging, incorrect temperature setting, control material degradation

Westgard multirule flowchart

Control data1-2s?NoIn control: acceptYes1-3s?YesNo2-2s?YesNoR-4s?YesNo4-1s?YesNo10x?YesOut of control: reject run, hold patient resultsNo to all: accept (1-2s was a warning)
Classic Westgard multirule logic for two control levels. A 1-2s warning triggers inspection by the rejection rules; any rejection rule violated means reject the run.

Laboratories running three control levels typically substitute 2of3-2s for 2-2s, 3-1s for 4-1s and 6x or 9x for 10x so the rules fit the number of measurements per run.

Try it: Levey-Jennings chart and rule checker

Interactive Levey-Jennings chart and Westgard rule checker

Enter each control level's established mean and SD, then paste results in run order (one value per run, separated by spaces, commas or new lines). Run n on each level is treated as the same analytical run. Leave a level blank to skip it.

The sample data contain a 1-2s warning, a within-level 2-2s, an R-4s and a 4-1s pattern so you can see each flag. A standalone two-level version is at Westgard Rules Calculator (Two Control Levels).

When to reject a run

  • Any enabled rejection rule is violated on any level, within the run or across recent runs.
  • A 1-2s alone is not a rejection in the classic multirule, but it should trigger a check of all other rules.
  • Rule choice should match the method's sigma performance; high-sigma tests may use 1-3s alone while low-sigma tests need more rules and more controls. Use the Sigma Metric Calculator for the Clinical Lab.
  • Do not simply repeat the control until it passes. Repeating without a reason hides real error.

Troubleshooting out-of-control QC

  1. Identify the rule and the error type (random or systematic).
  2. Check what changed: reagent, calibrator or control lot, calibration date, maintenance, operator, temperature.
  3. Inspect the control material: expiration, open-vial stability, reconstitution, storage.
  4. Fix the root cause, then rerun controls and confirm they are in control.
  5. Evaluate patient results since the last acceptable run and document everything. See Corrective Action for Westgard Rule Violations.

If your rules are configured in QC software, see Bio-Rad Unity QC consulting, Unity Real Time Westgard rules setup and building an IQCP from Unity QC data.

Frequently asked questions

What is the difference between a Levey-Jennings chart and Westgard rules?

The Levey-Jennings chart is the graph: control values over time against the mean and ±1/2/3 SD lines. Westgard rules are the decision criteria applied to the points on that graph to decide whether a run is accepted or rejected.

What does the 1-2s rule mean?

One control result falls outside mean ±2SD. In the classic Westgard multirule it is a warning that prompts checking the rejection rules, not an automatic rejection, because about 1 in 20 in-control results will exceed 2SD by chance.

Why not reject every run that is outside 2SD?

With two controls per run, a 1-2s rejection rule falsely rejects roughly 9% of good runs, and with three controls about 13%. The multirule keeps false rejection low while still catching real error.

Which Westgard rules detect systematic error?

2-2s, 4-1s, 2of3-2s, 3-1s and the mean rules (6x, 8x, 9x, 10x, 12x). They look for results clustering on one side of the mean.

Which Westgard rules detect random error?

1-3s and R-4s. R-4s should be applied only within a single run.

How many data points are needed to set up a Levey-Jennings chart?

A widely used rule of thumb is at least 20 results over at least 10 working days, preferably 20 working days, before setting final limits.

Can I use the manufacturer insert mean and SD?

The insert ranges are usually wide because they reflect many laboratories. CLIA at 42 CFR §493.1256(d)(10) allows a stated value for assayed controls only if the laboratory verifies it; most laboratories establish their own mean and SD.

Are Westgard rules required by CLIA?

CLIA requires that control results meet the laboratory's criteria for acceptability before patient results are reported, but it does not mandate specific Westgard rules. The laboratory selects and documents its rules.

Want a second set of eyes on your QC rules, limits or survey preparation?

Book a free consultation

Related tools and guides

Sources and standards

Sound Lab Consultants is an independent consulting firm and is not affiliated with Westgard QC, Bio-Rad, CLSI, CMS, CDC or any accrediting organization. Rule names are used descriptively. Calculators are educational aids; confirm results against your laboratory's approved QC procedures.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.