Westgard Rules and QC Rule Selection in Unity Real Time

In Unity Real Time, Westgard (SPC) rules are set per test under Configure, Rules/Settings, and each rule can be set to warn or reject. Westgard Advisor online can suggest rules and the number of controls per run from your data, an allowable total error and a consensus group. Under CLIA, the lab still has to decide which rules to use, write them into the QC procedure, and apply them before reporting patients.

Sound Lab Consultants (SLC) is an independent lab compliance consulting firm that helps labs design and document QC rules. We are not Bio-Rad and do not support its software; we help you make rule choices that your director can defend and your staff can follow.

Gloved hands attaching a pipette to a pipette aid in a clinical laboratory
Photo: Diane A. Reid, public domain, via Wikimedia Commons

Why rule selection matters more than rule count

Every QC rule trades two risks. A tight rule catches more real errors but also rejects more good runs. A loose rule rarely false-rejects but can let a real shift reach patient reports. The right balance depends on how precise the method is compared with how much error the test can tolerate clinically.

CLIA does not name specific Westgard rules. 42 CFR 493.1256(c) says control procedures must detect immediate errors and monitor accuracy and precision over time. Paragraph (b) ties the number, type and frequency of controls to the performance specifications the lab verified or established under 42 CFR 493.1253. In other words, the rule set is supposed to follow from what the lab knows about the method, not from a default.

A common finding in our reviews is a single rule set applied to every analyte on an analyzer. It is simple to manage, but it means a very precise sodium method and a less precise troponin method are judged the same way. Neither is likely to be set well.

Where the rules live in Unity Real Time

Bio-Rad's online help lists the main locations in its Functions and Where to Find Them table:

  • Configure, Rules/Settings is where SPC rules and test settings are selected.
  • Configure, Analytical Goals holds quality goals such as allowable error.
  • Advisors, Design QC Rules is where Westgard Advisor online generates and applies suggested rules (see the Apply Rules help page).
  • Review, Bench Review and supervisor review show data points that break a rule; per the help pages, reject-level violations show in red and warn-level in yellow.
  • Reports, Audit Trail records changes.

The QCNet Help Center also has articles on setting SPC rules, selecting rules at the test level and setting rules from current floating statistics. Your staff should use those for the mechanics. The compliance question is what goes into those screens and how the choice is recorded.

How Westgard Advisor fits in

According to Bio-Rad's Westgard Advisor online overview, the feature uses the test's data, a selected TEa and a selected consensus group (a date range of the lab's own data, or a Unity Interlab peer, method or all-labs group) to recommend SPC rules and the number of control samples per run. It can display an OPSpecs chart.

Two points in that help text are worth copying into your procedure. First, Bio-Rad says the suggestions are recommendations only and that lab conditions may make some rules impossible to use. Second, when no multirule meets the quality goal, the tool reports a maximum QC condition and recommends all available multirules. That second case is a signal that the method itself may need attention, not just a tighter rule.

We encourage labs to treat Advisor output as an input to a director decision, not as the decision. Save the Advisor report or a screenshot with the date, the TEa source and the consensus group used, and file it with the signed rule selection.

A practical rule selection process

This is the sequence SLC typically helps labs put in writing:

  • Pick a quality goal per analyte. Record the TEa and its source, such as CLIA PT acceptance limits, a biological variation database or a clinical decision. Do not mix sources without noting it.
  • Use the lab's own mean and SD. Set targets from your data on the current lot, not from the control insert range. Insert ranges are usually wide because they cover many labs and instruments.
  • Estimate bias. Peer group data from Unity Interlab or Unity Next, or PT results, give a reasonable bias estimate.
  • Calculate sigma or use OPSpecs. High-performing methods can often use a simple rule with few controls; weaker methods need multirules and possibly more controls per run.
  • Decide warn versus reject. A 1-2s rule is often used as a warning only. If it is set to reject, expect many false rejections.
  • Write it down. List each analyte, its rules, warn or reject status, controls per run and the reason, and have the director sign and date it.
  • Match the software to the paper. Check that Unity Real Time settings exactly match the signed list. Recheck after upgrades, lot changes and new tests.
  • Review periodically. Revisit rules when performance changes, when a new lot behaves differently, or at least at the review interval your QC procedure sets.

Common Westgard rules, briefly

Labs use many combinations, but these are the rules most often seen in Unity rule settings and QC procedures:

  • 1-2s: one result beyond 2 SD. Usually a warning, since it false-rejects often when used alone.
  • 1-3s: one result beyond 3 SD. Commonly a rejection rule for random error.
  • 2-2s: two results in a row, or two levels in one run, beyond 2 SD on the same side. Points to systematic error.
  • R-4s: range between two controls in a run exceeds 4 SD. Points to random error.
  • 4-1s: four results in a row beyond 1 SD on the same side. Detects smaller shifts.
  • 10x: ten results in a row on the same side of the mean. Detects small, steady bias.

Rules that look across runs, such as 4-1s and 10x, depend on consistent targets. If targets change with every lot without a crossover, those rules lose their meaning. Our page on Westgard rules and CLIA quality control covers rule logic in more depth.

Rules across multiple instruments and sites

Labs with two or more analyzers running the same test, or several testing sites, face an extra decision: one rule set for all, or rules per instrument. Unity lets you set rules at the test level, so each instrument can carry its own rules if their performance differs. The trade-off is training. Staff moving between benches must know which rules apply where, and the written rule table has to show it clearly.

Separately, 42 CFR 493.1281 requires labs that run the same test on different instruments, methods or sites to evaluate and define the relationship between results twice a year. QC data is not a substitute for that comparison, but consistent rules and targets make it easier to spot an instrument drifting away from the other instruments before the twice-yearly study does. When we review multi-instrument labs, we check that the comparison study, the rule table and the Unity settings all name the same instruments and that changes to one are reflected in the others.

If one instrument repeatedly needs tighter rules or more controls than its twin, treat that as a performance finding. Document it, look for a cause such as maintenance, reagent handling or calibration, and decide whether the instrument should stay in service for that test.

Documentation that holds up in a survey

Surveyors rarely argue with a lab's choice of rules. They look for whether the lab made a choice, wrote it down and followed it. For each test system, keep:

  • The QC procedure naming rules, warn or reject status and the action for each, approved under 42 CFR 493.1251(d).
  • The signed analyte rule table with quality goals and the date.
  • Westgard Advisor or OPSpecs output used to support the choice, if any.
  • Evidence that the software settings match, such as a dated screenshot or rule listing.
  • Records of rule changes and why they were made.
  • Corrective action records for every rejection, meeting 42 CFR 493.1282.

If you use a rule as a warning only, say what staff must do when it fires, even if the answer is "note and continue to evaluate with the reject rules." Silence in the procedure tends to produce inconsistent behavior at the bench.

Training staff on rules

Rules are only as good as the person reading them at 3 a.m. Competency assessment under 42 CFR 493.1235 and the Subpart M personnel sections should include QC interpretation. Practical checks include asking staff to explain a recent red flag in Unity, what they did, and where the record is. SLC can help build those questions into your competency program and our six elements of competency assessment guide explains the framework.

Frequently asked questions

Where do I set Westgard rules in Unity Real Time?

Bio-Rad's online help lists SPC rule selection under Configure, Rules/Settings, with rules selectable at the test level. Westgard Advisor suggestions are applied from Advisors, Design QC Rules. Use Bio-Rad's help center for step-by-step instructions.

Do I have to accept Westgard Advisor suggestions?

No. Bio-Rad's help says the suggested rules are recommendations only. The lab director decides which rules to use and the lab documents that decision.

Does CLIA require Westgard rules?

No. CLIA requires control procedures that detect immediate errors and monitor accuracy and precision over time, with defined acceptability criteria (42 CFR 493.1256). Westgard multirules are a widely used way to meet that.

Should 1-2s be a reject rule?

Usually not on its own, because it rejects many runs that are actually in control. Many labs use it as a warning that triggers review against other rules. Whatever you choose, write it in the procedure.

What TEa should I use?

Pick a source and record it for each analyte. Options include CLIA proficiency testing acceptance limits, biological variation goals or a clinically defined limit. The director should approve the choice.

How often should QC rules be reviewed?

Set an interval in your QC procedure and also review rules when a method's performance changes, a new control lot behaves differently, or a new instrument or test is added.

Can SLC set the rules inside our Unity software for us?

Your staff or Bio-Rad support should make changes in the software. SLC reviews the settings, recommends rule choices and helps document them.

Sources: 42 CFR 493.1256; 42 CFR 493.1253; 42 CFR 493.1251; 42 CFR 493.1282; 42 CFR 493.1235; QCNet help: Westgard Advisor online overview; QCNet help: Apply Rules with Westgard Advisor; QCNet help: Functions and Where to Find Them; QCNet help: Bench and Supervisor Review; QCNet Help Center: Unity Real Time; Unity Real Time 2 User Guide (PDF).

Related SLC pages: Bio-Rad Unity QC consulting, Westgard rules and CLIA QC, CLIA QC requirements, Quality control log, Competency assessment elements, PT and QC consulting.

Want a second set of eyes on your Unity Real Time rule settings?

Book a free consultation

Bio-Rad, Unity, Unity Real Time, UnityWeb, Unity Next, QCNet, InstantQC, Westgard Advisor, EQAS, Lyphochek, Liquichek and InteliQ are trademarks of Bio-Rad Laboratories, Inc. Sound Lab Consultants is an independent consulting firm and is not affiliated with or endorsed by Bio-Rad. Product details on this page come from Bio-Rad's public pages and user documentation, linked below; confirm current features with Bio-Rad. This page is general information, not legal advice.

Sound Lab Consultants · 25 Arizona Pl, Chandler, AZ 85225

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.