Total Allowable Error and Bias Calculator
Last reviewed Nov 1, 2026 · Sound Lab Consultants
Bias % = (your mean − target) / target × 100. Total error combines bias and imprecision: TE = |bias| + 2 × CV (some laboratories use 1.65 × CV for a one-sided 95% estimate). A method is acceptable when TE is less than or equal to the allowable total error (TEa), such as a CLIA proficiency testing limit.
Total error and bias calculator
Formulas
Worked examples
Total cholesterol (CLIA TEa ±10%): lab mean 206 mg/dL vs target 200 → bias = 6/200 × 100 = 3.0%. CV = 2.0%. TE = 3.0 + 2 × 2.0 = 7.0% ≤ 10% → pass. With k = 1.65, TE = 6.3%. Sigma = (10 − 3)/2 = 3.5.
Potassium (CLIA ±0.3 mmol/L): at a target of 4.0 mmol/L, TEa = 0.3/4.0 = 7.5%. Lab mean 4.12 → bias 3.0%; CV 2.5% → TE = 3.0 + 5.0 = 8.0% > 7.5% → fail.
Bias vs imprecision
Bias is a consistent offset from the true value (systematic error); CV is scatter (random error). Total error estimates how far a single patient result could be from the truth for about 95% of results. A method can have tiny CV but fail because of bias, or vice versa.
Getting the target value
- Proficiency testing target or peer group mean.
- Reference method or reference material value.
- Comparison of methods regression at the decision level (bias = (a + b·Xc) − Xc).
TEa sources
| Analyte | CLIA acceptance limit (target value ±) |
|---|---|
| Albumin | ±8% |
| Calcium, total | ±1.0 mg/dL |
| Cholesterol, total | ±10% |
| Creatinine | ±10% or ±0.2 mg/dL (greater) |
| Glucose (excluding FDA-cleared home-use devices) | ±8% or ±6 mg/dL (greater) |
| Hemoglobin A1c | ±8% |
| Potassium | ±0.3 mmol/L |
| Sodium | ±4 mmol/L |
| Triglycerides | ±15% |
| Urea nitrogen | ±9% or ±2 mg/dL (greater) |
Verified Nov 1, 2026 against 42 CFR §493.931 (routine chemistry) in 42 CFR Part 493, Subpart I. For limits written as "X% or Y units (greater)", compute both at your decision level and use the larger.
Convert limits to the method's sigma with the Sigma Metric Calculator for the Clinical Lab and choose QC rules accordingly in the Westgard Rules Calculator (Two Control Levels).
Frequently asked questions
How do you calculate total error?
TE = |bias| + 2 × CV, with both in percent at the same concentration. Some laboratories use 1.65 instead of 2.
What is the difference between TE and TEa?
TE is the error your method actually shows. TEa is the allowable limit, the quality goal TE must not exceed.
When should I use 1.65 instead of 2?
1.65 corresponds to a one-sided 95% estimate and is used by some guidance documents; 2 is a more conservative and common choice. Document which you use.
How do I calculate bias percent?
Bias% = (lab mean − target) / target × 100. Use the absolute value when adding it to TE.
What if my method fails TE vs TEa?
Investigate calibration and bias, look for ways to reduce imprecision, consider method replacement, and increase QC while the issue is resolved.
Are CLIA PT limits appropriate as TEa?
They are a widely used regulatory minimum. Many laboratories use tighter clinically based goals for key analytes.
How is total error related to sigma?
Sigma measures how many CVs fit in TEa after bias; a TE that just equals TEa with k = 2 corresponds to 2 sigma.
Want a second set of eyes on your QC rules, limits or survey preparation?
Book a free consultationRelated tools and guides
- Levey-Jennings Charts and Westgard Rules
- QC Mean, SD and CV% Calculator
- Westgard Rules Calculator (Two Control Levels)
- Sigma Metric Calculator for the Clinical Lab
- Corrective Action for Westgard Rule Violations
- Setting Up a New QC Lot Mean and SD
- Bio-Rad Unity QC consulting
- Unity Real Time Westgard rules setup
- IQCP using Unity QC data
Sources and standards
- 42 CFR §493.1256 Standard: Control procedures (eCFR)
- 42 CFR §493.1282 Standard: Corrective actions (eCFR)
- 42 CFR Part 493, Subpart I, Proficiency testing acceptance limits (eCFR)
- CMS: Clinical Laboratory Improvement Amendments (CLIA)
- CDC: CLIA program
- CLSI C24, Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions (consult the current edition)
Sound Lab Consultants is an independent consulting firm and is not affiliated with Westgard QC, Bio-Rad, CLSI, CMS, CDC or any accrediting organization. Rule names are used descriptively. Calculators are educational aids; confirm results against your laboratory's approved QC procedures.