Joint Commission Laboratory Accreditation Requirements
To earn Joint Commission laboratory accreditation, a U.S. lab needs a non-waived CLIA certificate, any required state license and an ongoing quality review program. It must then meet the CAMLAB standards, including patient safety goals, proficiency testing and staff competency, and pass an on-site survey built around patient and specimen tracers. Labs are resurveyed every two years.
Last reviewed November 14, 2026. Main sources: The Joint Commission, Laboratory Services accreditation program, Accreditation Process, Laboratory Accreditation Survey Activity Guide, July 2026 (PDF), Laboratory National Patient Safety Goals, January 2026 (PDF), 42 CFR Part 493 (CLIA) and CMS CLIA program.
What the Joint Commission Laboratory Accreditation Program is
The Joint Commission is an independent nonprofit that accredits health care organizations. Its Laboratory Services program, also called the Laboratory and Point-of-Care Testing Accreditation Program, has accredited hospital labs since 1979 and freestanding labs since 1995. Today roughly 1,500 accredited organizations hold this accreditation, covering close to 2,000 labs with CLIA certificates (TJC Laboratory Services). Surveyors judge the lab against the laboratory standards and against relevant CMS and OSHA rules, with a focus on test quality and patient safety (TJC Accreditation Process). Because CMS has approved the program under CLIA, accreditation is also how a lab meets its federal CLIA obligations (TJC, Renewed Deeming Authority for Laboratories, 2024).
The program accredits these lab types:
- Hospital labs
- Reference and stand-alone labs
- Point-of-care testing sites
- IVF labs
- Physician office labs
- Ambulatory care labs
- Anatomic pathology labs
- Toxicology labs
- Histology labs
- Mohs surgery clinic labs
Core eligibility requirements: who can apply
The Joint Commission says a lab may apply only if it meets every condition below (source):
- It is located in the United States or a US territory, or it is run by the US government or under a congressional charter abroad.
- It holds a facility license or registration wherever the law requires one.
- It holds a non-waived CLIA certificate.
- It can show ongoing quality assessment and improvement, including review by clinicians or a qualified designee as CLIA defines them.
- It lists the services it provides directly, under contract or through other arrangements.
- Its services can be evaluated against Joint Commission standards.
CMS approval covers these CLIA specialties and subspecialties (source):
| Specialty | Subspecialties |
|---|---|
| Microbiology | Bacteriology, mycobacteriology, mycology, parasitology, virology |
| Diagnostic immunology | Syphilis serology, general immunology |
| Chemistry | Routine chemistry, toxicology, endocrinology |
| Hematology | Routine hematology, coagulation |
| Immunohematology | ABO group, D (Rho) typing, unexpected antibody detection, compatibility testing, antibody identification |
| Pathology | Histopathology, oral pathology, cytology |
CAMLAB standards: what the laboratory manual requires
Once a lab is eligible, it is measured against the Comprehensive Accreditation Manual for Laboratory and Point-of-Care Testing (CAMLAB). The 2026 edition was released October 15, 2025; a 2027 CAMLAB edition is listed for release on December 15, 2026. CAMLAB is the official manual of Joint Commission laboratory standards. Surveyors score those standards through Elements of Performance, the specific expectations described on The Joint Commission's Accreditation Process page. The manual also includes National Patient Safety Goals and Accreditation Participation Requirements. The sections below cover those contents, along with the quality control, proficiency testing, personnel, and competency rules that sit behind them.
Accreditation Participation Requirements (APR)
APRs are the ground rules for taking part in accreditation. In the APR chapter The Joint Commission posts publicly (PDF, written for its DMEPOS program), each APR is either met or not met, and a miss becomes a Requirement for Improvement like any other finding. Its items include:
- Submit required data and information on time (APR.01.01.01).
- Give accurate information at every stage; falsification leads to preliminary denial (APR.01.02.01).
- Report changes to the information in your application, including between surveys (APR.01.03.01).
- Allow the survey to take place; refusing leads to denial (APR.02.01.01).
- Tell staff they may report safety or quality concerns to The Joint Commission, and take no disciplinary or punitive action against anyone who does (APR.09.02.01).
- Keep services and the physical setting free of any condition that would count as an Immediate Threat to Health or Safety (APR.09.04.01).
Read the laboratory wording of each APR in your CAMLAB or E-dition, since notes and time frames can differ by program.
National Patient Safety Goals for laboratories (effective January 2026)
The laboratory NPSG list effective January 2026 has three goals:
| Goal | What the lab must do |
|---|---|
| NPSG.01.01.01 Patient identification | Use at least two patient identifiers when collecting blood and other specimens, giving blood or blood components and doing other procedures such as bone marrow aspirates. A room number or location never counts as an identifier. Label specimen containers in front of the patient. |
| NPSG.02.03.01 Critical results | Work with leaders to write procedures that define critical results, say who reports them and who receives them, and set the acceptable time from result to report. Documentation is required, and the lab must measure how timely reporting actually is. |
| NPSG.07.01.01 Hand hygiene | Run a hand hygiene program built on current CDC and/or WHO guidelines, set compliance goals and improve against them. |
Hospitals and critical access hospitals moved from NPSGs to National Performance Goals on January 1, 2026. The laboratory program still uses NPSGs (source). Surveyors ask staff about the NPSGs during tracers and at the orientation session (SAG).
Quality control, proficiency testing and record retention
The Joint Commission confirms that the lab is enrolled in a CMS-approved PT program for every regulated analyte. It reviews two years (the last six events) of PT results, attestations and remedial action (SAG). CLIA sets the scoring:
| PT area (42 CFR Subpart H) | Satisfactory event score |
|---|---|
| ABO group and D (Rho) typing (493.859) | 100% |
| Unexpected antibody detection (493.861) | 100% |
| Compatibility testing (493.863) | 100% |
| Antibody identification (493.865) | 80% |
| Most other specialties, e.g. routine chemistry (493.841), hematology (493.851), general immunology (493.837) | 80% |
Two unsatisfactory events in a row for one analyte, or two out of any three in sequence, count as unsuccessful performance and can lead to CMS sanctions (493.803). Other QC and records items surveyors look for:
- IQCP for each test system that uses it: risk assessment, QC plan, quality assessment, implementation date and plan reviews (SAG).
- Method verification: before reporting patient results on an unmodified FDA-cleared test, show accuracy, precision and reportable range comparable to the manufacturer's, and verify reference intervals. Modified or lab-developed tests need full performance specifications (42 CFR 493.1253). The guide asks for validation studies for new instruments and tests from the past two years.
- Maintenance and function checks done and documented at least as often as the manufacturer specifies (42 CFR 493.1254), plus a list of critical equipment and its maintenance records.
- Record retention: CLIA requires PT records, QC and patient test records, and quality assessment records to be kept for at least two years. Pathology reports and histopathology slides are kept 10 years, and cytology slides 5 years (42 CFR 493.1105). The Joint Commission document list asks for 24 months of PT data for each CLIA number, on initial surveys and resurveys alike.
- Lab reports: each report must show the name and address of the lab that performed the test, including results from reference labs and contracted pathologists (TJC Standards FAQ, Document and Process Control).
Personnel: training, competency and qualifications
CLIA Subpart M sets the federal personnel qualifications and duties. On top of that, The Joint Commission has one expectation of its own: initial training and initial competency must be documented separately. Competency must cover all six CLIA procedures on the CLIA schedule. Both are explained in the training vs competency and six elements sections below.
What surveyors check on site
The July 2026 Survey Activity Guide (issued June 15, 2026, effective July 1, 2026) shows how surveyors test the CAMLAB standards on site. The main requirement groups are:
- CLIA compliance: the right certificate type for your test complexity, the specialties listed, director and staff licenses, PT enrollment for each regulated analyte, test volumes and IQCP documents. See regulatory review and IQCP.
- Documents on day one: most records cover 24 months. That includes PT, QC, calibration verification, correlations, validations, temperature charts, maintenance, competency, and a CMS-209 completed on site. See the document list.
- Proficiency testing: two years, or six events, of completed PT with attestations and remedial action. See PT review.
- People: qualifications, orientation, training, competency, staffing and contracted staff. See competency assessment.
- Patient tracers across every specialty (below).
- Patient safety systems and leadership, including safety culture and event reporting (below).
- Environment of care and emergency management. See EC and EM.
- Infection prevention and control: an assessment tool added for labs in July 2025, built on IC.04.01.01, IC.06.01.01 and IC.07.01.01. See IC requirements.
| Survey activity | Suggested time in the guide |
|---|---|
| Arrival, planning and safety briefing | 15-30 min |
| Opening conference / orientation to the organization | 15-30 / 30-45 min |
| Regulatory review | 30-45 min (longer with IQCP) |
| PT validation and PI data review | 90-180 min |
| Individual tracer | 60-120 min each |
| HR and competence assessment | 60-120 min |
| Environment of care and emergency management | 45-90 min |
| Exit conference | 30 min |
For a lab in a mid-sized acute care hospital, one surveyor is typical. Since July 1, 2026, health systems can ask for a system survey (Joint Commission News, July 2026), which reviews several labs in one region within two weeks and still awards accreditation to each applicant separately. To qualify, the system must meet at least five of nine integration criteria, such as shared policies and common training, competency and QC programs (source).
CLIA rules behind the standards: waived and non-waived testing
CLIA (42 CFR Part 493) sets rules by test complexity. Under 42 CFR 493.15(e), labs that do only waived tests must follow the manufacturer's instructions and meet the certificate of waiver rules. Moderate and high complexity (non-waived) testing brings in much more:
- Proficiency testing (Subpart H)
- Quality systems and QC (Subpart K), including general quality assessment in 493.1239
- Personnel qualifications and duties (Subpart M)
- Written competency policies for employees and consultants (493.1235)
A Joint Commission applicant needs a non-waived certificate. Waived testing in the organization, such as point-of-care glucose, is still traced. The guide's waived-testing checks include watching a fingerstick, internal and external control records, age-specific reference ranges in the patient record, and reagent lot numbers and expiration dates.
Training vs competency: The Joint Commission documents them separately
Training and competency are easy to blur, but they answer different questions. Training records show the employee was taught to do the task your lab's way. Competency records show the employee then did it correctly on their own. A Joint Commission laboratory field director, writing in MLO, Dec 2025, says the agency expects initial training and initial competency in separate records. She calls this a Joint Commission-specific expectation and says one combined form is a frequent finding.
Surveyors pull HR files for staff who came up in tracers. In very small labs (five or six people), they may look at nearly every file. They also confirm that job descriptions match the minimum education, experience and credential rules. Surveyors don't interpret your paperwork for you. Expect to show them the process and explain it.
In practice, keep these records for each person and each test system:
- A dated training checklist signed by the trainer
- A separate initial competency record covering all six CLIA elements, finished before the person reports patient results on their own
- Later competency records on the CLIA schedule
The six CLIA competency assessment elements
For non-waived testing, competency evaluations must include at least these six procedures (42 CFR 493.1413(b)(8) for moderate complexity, 493.1451(b)(8) for high complexity):
- Direct observation of routine patient testing, including patient preparation if it applies, plus specimen handling, processing and testing.
- Monitoring how test results are recorded and reported.
- Review of intermediate results or worksheets, QC records, PT results and preventive maintenance records.
- Direct observation of instrument maintenance and function checks.
- Testing of previously analyzed specimens, internal blind samples or external PT samples.
- Assessment of problem-solving skills.
Timing: evaluate each person at least twice during their first year of patient testing, then at least yearly. If the method or instrument changes, reassess them before they report patient results on it (493.1413(b)(9), 493.1451(b)(9)). The Joint Commission's HR session also asks how you assess contracted staff (SAG). More: CLIA competency assessment six elements.
Patient Safety Systems (PS) chapter
The laboratory manual opens with a Patient Safety Systems chapter (CAMLAB Update 1, July 2026, PDF). It explains how existing standards fit together into one system for reducing harm. Its three guiding principles are:
- Link the standards to daily work.
- Help labs become learning organizations.
- Promote proactive safety methods that build trust and reduce fear of blame.
Topics a lab leader should be able to speak to:
- Leaders and safety culture: leaders build a culture of transparency, accountability and respect, and report and address behaviors that undermine safety.
- Fair and just culture: staff report events through a standard process without fear of punishment for system flaws, while still being accountable for their own choices.
- Trust, report, improve: trust leads to reporting, reporting leads to fixes, and visible fixes build more trust.
- Reporting and data: capture adverse events, close calls and hazardous conditions, analyze them, change the process and share the lessons. Give reporters timely feedback.
- Proactive risk assessment: study high-risk, high-volume processes before they fail. Failure Modes and Effects Analysis (FMEA) is the best-known tool.
- Sentinel events: handled under the Sentinel Event Policy (SE) chapter, with Joint Commission support for the systematic analysis.
- Patient activation: patients and families get information and take part in decisions, and unexpected outcomes are disclosed.
In the July 2026 survey guide, these themes come up in the orientation session as Leadership Safety Culture, culture of safety data on the document list, and performance improvement examples (SAG).
Survey process: tracer methodology, specimen tracers and documents reviewed
The on-site survey checks the standards in practice. It starts with document review and spends most of its time on tracers. On day one, surveyors expect the document list items ready, including:
- CLIA certificates with specialties and subspecialties, state licenses, the test menu and major instruments, and a CMS-209 completed on site
- 24 months of PT data with investigations, worksheets and attestations
- IQCP documents, QC records with dates and times, calibration verifications and correlations
- Personnel files: lab director credentials and contract, staff licenses, highest education, and 24 months of training and competency records
- Performance improvement data, safety inspections, hazardous waste manifests, the emergency operations plan and culture of safety results
Initial surveys need at least four months of data, except PT. If documents are not ready, the surveyor goes straight to a tracer.
Tracers are where most survey time goes. The surveyor picks a patient and follows that patient's testing through the pre-analytic, analytic and post-analytic phases, all the way to the result in the chart (TJC Laboratory Services, Accreditation Process). Patients are picked so that every specialty and subspecialty on your CLIA certificate is covered. The guide sets three look-back windows: at least one record from 13 to 24 months ago, one from 6 to 12 months ago and one from the last six months (SAG).
Specimen tracers follow a sample from collection and labeling through processing, testing, QC and maintenance on the day of testing, lot numbers, and the final report in the chart. Blood bank tracers can follow a unit to the patient. Starting July 1, 2026, lab surveys focus on transfusion steps under lab control and review critical results only up to lab reporting (SAG, 2026 changes).
Along the way, surveyors:
- Watch infection control practice, the environment, QC, IQCP, maintenance and test performance
- Ask staff about hand-offs, use of data, National Patient Safety Goals, orientation and competency
- Ask about downtime and backup procedures, and how staff handle cyber risks
The guide also has checklists by specialty. Examples: next-generation sequencing pipeline validation in molecular, annual FDA verification of tissue suppliers, and screener workload and reviews in cytology. See tracer methodology by specialty.
Environment of Care and emergency management on survey
The guide's activity list allows 45 to 90 minutes for Environment of Care (EC) and emergency management (EM), and the session page estimates about 30 to 60. The work can happen during tracers, in a meeting, or both (SAG).
- EC tour and interviews: lab safety, infection control, security, safety equipment, space, chemical storage, hazardous materials and waste disposal, and information security. Point-of-care sites are checked when surveyors visit them.
- Biosafety and infection control: the IC tool checks for a qualified infection control leader, standard precautions, an annual infection risk assessment, PPE use, OSHA bloodborne pathogens and PPE rules, staff exposure follow-up, and protocols for high-consequence pathogens.
- Emergency management for independent labs: review the hazard vulnerability analysis, emergency operations plan and policies, and communications plan at least every two years. Be ready to discuss real events and the one required annual exercise from the past 12 to 36 months, along with staff training and lessons learned.
- Labs inside an accredited hospital: show access to the hospital's emergency plans and explain the lab's role, training and drills.
- Equipment: critical equipment lists, maintenance records, temperature charts, and alarms and backup for refrigerators and freezers in tissue and IVF areas.
Details: EC and EM and infection prevention and control.
Health equity standard LD.04.03.08: does it apply to labs?
Search summaries often list LD.04.03.08 as a lab requirement, so it helps to be precise. The Joint Commission added LD.04.03.08 to its Leadership chapter, effective January 1, 2023, with six elements of performance (ASCP, Oct 28, 2022). In summary, the organization must:
- Name a leader for the work.
- Assess patients' health-related social needs. A representative sample is allowed.
- Break quality and safety data down by sociodemographic factors to find disparities.
- Write and carry out a plan for at least one disparity.
- Act when the plan's goals are not met or not sustained.
- Report progress to leaders, practitioners and staff at least once a year.
Accreditation cycle: two years for labs, about three for other programs
The Joint Commission says an accreditation decision takes effect on the last day of the survey and lasts about three years for most programs. The laboratory program is the exception, with a two-year cycle (source). The reason is CLIA. The Joint Commission notes that CLIA requires non-waived moderate and high complexity labs to be surveyed every two years by CMS or an approved accrediting organization (source). CMS's own rule for certificate-of-compliance labs also sets biennial inspections (42 CFR 493.1777(b)). Health systems on the system survey option can line up the two-year surveys for several labs into one concurrent survey.
Unannounced surveys and notice
- Survey Activity Guide: most surveys happen with no notice or short notice, and every CMS deemed survey is unannounced. Exceptions are listed in the Accreditation Process chapter of the manual (SAG).
- Accreditation Process page: most surveys are unannounced, except non-deemed initial surveys. The page also says a 14-day notice applies to full laboratory surveys (source).
These two statements read differently, so confirm your lab's terms with your account executive. Either way, plan as if surveyors could arrive any day within the window.
When they arrive, check their identity on your Joint Commission Connect extranet site. Hand over the document list items quickly; if documents aren't ready, the surveyor starts with a tracer. Surveyors usually arrive around 7:45 a.m. unless your hours start later.
CMS deemed status for accredited labs
Under 42 CFR 493.551, CMS can deem a lab to meet CLIA when a CMS-approved private nonprofit organization accredits it. To use that route, the lab must:
- Get a CLIA certificate of accreditation
- Pay CLIA fees
- Meet the Subpart H proficiency testing rules
- Let its PT provider send results to the accrediting organization
- Allow unsuccessful PT results to be released to CMS
Deemed status doesn't end CMS oversight. CMS or its agent can run a validation or complaint inspection at any time during operating hours, and serious findings lead to a full CMS review (42 CFR 493.1780).
The Joint Commission's current CMS deeming authority for its laboratory program runs from May 24, 2024 to May 24, 2030. Six years is the longest term CMS grants (source). CMS's overview of the CLIA program is on its CLIA page.
What changed for Joint Commission lab accreditation in 2026
These are the changes The Joint Commission has published for its laboratory program in 2026:
- July 1, 2026 survey changes: minor revisions on blood administration and critical result reporting to align with CLIA and reduce duplication. Surveys now focus on transfusion processes the lab controls and limit critical result review to lab reporting (SAG July 2026).
- Lab system survey option, July 1, 2026: one coordinated survey within two weeks for a system's labs in one area, if the system meets at least five of nine criteria (JC News, July 2026).
- NPSGs, January 2026: the lab program's NPSG list has three goals. National Performance Goals replaced NPSGs only for hospitals and critical access hospitals (source).
- Manuals: the 2026 CAMLAB was released October 15, 2025, and the Patient Safety Systems chapter was updated in July 2026 (PDF). The 2027 CAMLAB is listed for release December 15, 2026.
For changes approved but not yet in effect, check The Joint Commission's prepublication standards page.
Where to read the standards: CAMLAB, E-dition and public standards
- CAMLAB: the Comprehensive Accreditation Manual for Laboratory and Point-of-Care Testing, which holds the standards and elements of performance (EPs).
- E-dition Laboratory: the online version of the manual. Accredited labs get one free single-user license through Joint Commission Connect, and more licenses can be bought. E-dition adds CMS crosswalks, filters for new or changed EPs and printing. Organizations that haven't applied can ask for a free 90-day trial; the lab services page describes it as a trial of the hospital standards, so confirm lab content is included (source, source).
- Public Standards: free, searchable, current-only standards for every program, with no login. They can't be printed or downloaded.
- Survey Activity Guide: posted on your extranet site and published as a public PDF.
- First-timer help: The Joint Commission's on-demand First-Timer's Roadmap to Laboratory Accreditation webinar explains who on its staff supports you before, during and after the survey.
How to apply for Joint Commission laboratory accreditation: steps
- Confirm eligibility. Check your non-waived CLIA certificate, state license and the specialties you test against the lists above.
- Read the standards. Use the public standards or an E-dition trial, plus the PS chapter and the Survey Activity Guide.
- Close gaps against CLIA first. Cover PT enrollment, QC, IQCP, personnel files and the six-element competency records.
- Submit the e-application in Joint Commission Connect. It records your services and volumes, which set the number of survey days, the team and the services reviewed. Fees are based on the services you provide and your average daily census (source, source).
- Work with your account executive. This person is your main contact for application and survey questions and for changes to your details.
- Prepare. Assemble the document list, run mock tracers in every specialty and assign an owner for each survey session.
- Host the on-site survey.
- Respond to findings. Submit Evidence of Standards Compliance within 60 days.
- Receive the decision.
- Stay compliant between surveys with Intracycle Monitoring, then expect the next survey within the two-year cycle.
After the survey: findings, ESC and decisions
At the exit conference, surveyors review a preliminary Summary of Survey Findings report. The report is posted on Joint Commission Connect. Findings are listed as Requirements for Improvement (RFIs) and plotted on the SAFER Matrix by how likely each is to cause harm and how widespread it is. For each RFI, the lab submits Evidence of Standards Compliance (ESC) within 60 days (source).
The possible decisions are:
- Limited Temporary Accreditation
- Accreditation
- Accreditation with Follow-up Survey
- Preliminary Denial of Accreditation
- Denial of Accreditation
Between surveys, the Intracycle Monitoring profile and the Focused Standards Assessment self-scoring tool help you keep up with high-risk topics. If you need help writing corrective action, see failed survey support.
Preparation checklist
- Make record review a daily habit. Pick one critical value a day and follow it through the chart, related PT, QC and temperature logs (MLO).
- Focus on the three areas labs most often struggle with: PT, competency and procedures (MLO).
- Keep six completed PT events per CLIA number, or a provider scorecard. Investigate every unacceptable result. A root cause analysis isn't required, but it helps prevent repeats.
- Match practice to written procedure. Review procedures each year with staff input, version control and one central location.
- Keep training and initial competency in separate records. Confirm all six CLIA elements and the CLIA schedule.
- Run a full mock survey, from opening to exit conference. Include nursing, respiratory therapy and quality where they touch lab testing.
- Build the document list folder. Use the 24-month look-back and the three tracer windows.
- Review the PS chapter themes (event reporting, close calls, proactive risk assessment), plus the EC/EM and infection control items.
Related guides:
- Survey preparation
- CLIA mock inspection
- How to prepare for a lab inspection
- COLA vs CAP vs Joint Commission
How Sound Lab Consultants helps
Sound Lab Consultants (SLC) works with clinical labs remotely on Joint Commission, CAP, COLA and state CLIA surveys. Typical work includes:
- Eligibility and CLIA certificate checks
- Gap reviews against the Survey Activity Guide document list
- Training and competency record redesign
- Mock tracers in every specialty
- PT and QC record review
- IQCP review
- Patient safety and event reporting process setup
- Help writing Evidence of Standards Compliance
SLC supports all CLIA specialties and subspecialties. See our CLIA consultant and clinical laboratory consulting pages.
Frequently asked questions
What does it mean to be accredited by The Joint Commission?
It means Joint Commission surveyors visited the lab, measured its work against the published laboratory standards and relevant CMS and OSHA rules, and granted an accreditation decision once any Requirements for Improvement were corrected. For a CLIA lab, accreditation by a CMS-approved organization is also the route to a CLIA certificate of accreditation, so CMS treats the lab as meeting CLIA.
What are the requirements of Joint Commission accreditation for a laboratory?
The Joint Commission lists six eligibility conditions. The lab is in the United States or its territories, or is run by the US government abroad. It holds any facility license the law requires and a non-waived CLIA certificate. It continually assesses and improves quality, with review by qualified clinicians. It identifies its direct and contracted services, and its services can be evaluated against the standards. After that, it must meet the CAMLAB standards and pass the on-site survey.
Is Joint Commission accreditation required for a laboratory?
No federal rule requires Joint Commission accreditation in particular. CLIA requires a lab to hold a CLIA certificate that fits the tests it performs, unless it is CLIA-exempt (42 CFR 493.3 and 493.5). A non-waived lab can be inspected by CMS or its state agency under a certificate of compliance, or it can choose a CMS-approved accrediting organization such as The Joint Commission and hold a certificate of accreditation (42 CFR 493.551). Check your state rules and contracts for any added expectations.
What are the accreditation standards for laboratories?
Joint Commission lab standards are published in the Comprehensive Accreditation Manual for Laboratory and Point-of-Care Testing (CAMLAB); the 2026 edition was released October 15, 2025. Chapters named in Joint Commission lab sources include Patient Safety Systems, National Patient Safety Goals, Sentinel Event Policy, Infection Prevention and Control, Document and Process Control and The Accreditation Process. Joint Commission manuals also carry Accreditation Participation Requirements. Underneath sit the federal CLIA rules in 42 CFR Part 493.
What are the new Joint Commission standards for 2026?
For laboratories, the July 2026 Survey Activity Guide (effective July 1, 2026) made minor revisions on blood administration and critical result reporting to align with CLIA: lab surveys now focus on transfusion steps the lab controls and review critical results only as far as lab reporting. A lab system survey option also began July 1, 2026. The lab NPSG list effective January 2026 has three goals; the new National Performance Goals replaced NPSGs only for hospitals and critical access hospitals. The 2027 CAMLAB is scheduled for release December 15, 2026.
What proficiency testing score does a lab need under CLIA?
For most regulated analytes, CLIA counts a testing event as satisfactory at 80 percent or higher. ABO group and D (Rho) typing, unexpected antibody detection and compatibility testing require 100 percent, while antibody identification requires 80 percent. Two unsatisfactory events in a row for one analyte, or two out of any three in sequence, count as unsuccessful performance (42 CFR Subpart H).
How long does Joint Commission laboratory accreditation last?
Two years. The Joint Commission says most programs are on a cycle of about three years, but the laboratory program is a two-year cycle, which matches the CLIA requirement that accredited non-waived labs be surveyed every two years.
Are Joint Commission laboratory surveys unannounced?
The July 2026 Laboratory Survey Activity Guide says most lab surveys come with no notice or only short notice, and that every CMS deemed survey is unannounced. The Joint Commission's general accreditation process page also says a 14-day notice applies to full laboratory surveys. Because the two pages read differently, confirm your lab's notice terms in the Accreditation Process chapter of your manual or with your account executive.
Does Joint Commission accreditation replace a CLIA certificate?
No. The lab still needs a CLIA certificate. Under 42 CFR 493.551, a lab that uses accreditation to meet CLIA obtains a certificate of accreditation, pays CLIA fees, meets the proficiency testing rules and lets its PT results be shared. CMS can still run validation and complaint inspections at any time (42 CFR 493.1780).
What are the six CLIA competency assessment elements?
Under 42 CFR 493.1413(b)(8) and 493.1451(b)(8): direct observation of routine patient testing, including patient preparation and specimen handling; monitoring how results are recorded and reported; reviewing worksheets, QC, PT and preventive maintenance records; direct observation of instrument maintenance and function checks; testing previously analyzed specimens, blind samples or PT samples; and assessing problem-solving skills.
Does The Joint Commission require separate training and competency records?
Yes. A Joint Commission laboratory field director, writing in MLO, says the agency expects labs to keep initial training and initial competency in separate records and that using one form for both is a common mistake. Training shows the person was taught your method; competency shows they can perform it correctly.
Does the health equity standard LD.04.03.08 apply to laboratories?
LD.04.03.08 took effect January 1, 2023 with six elements of performance. In its October 2022 webinar, The Joint Commission said the standard applies to hospitals, critical access hospitals, primary care groups in ambulatory care and some behavioral health services. The Laboratory program was not on that list. A lab inside an accredited hospital should ask its hospital quality team how lab data feeds the hospital's equity work.
Where can I read the Joint Commission laboratory standards?
The standards are in the Comprehensive Accreditation Manual for Laboratory and Point-of-Care Testing (CAMLAB). Accredited labs get a free single-user E-dition license through Joint Commission Connect. Organizations that haven't applied can ask for a free 90-day E-dition trial (the lab services page describes it as a trial of the hospital standards, so ask whether lab content is included) or view the free public standards online. The public standards can't be printed or downloaded.
How far back do Joint Commission lab surveyors look?
Most items on the Survey Activity Guide document list cover the past 24 months. Initial surveys need at least four months of data, except proficiency data. For tracers, surveyors pick at least one patient record from 13 to 24 months ago, one from 6 to 12 months ago and one from the last six months.
What happens after the survey?
Surveyors post a preliminary Summary of Survey Findings on Joint Commission Connect and review it at the exit conference. Findings are called Requirements for Improvement and are plotted on the SAFER Matrix. The lab submits Evidence of Standards Compliance within 60 days, and the final decision is then issued, effective the last day of the survey.
Can a waived-only lab get Joint Commission laboratory accreditation?
Not as a stand-alone applicant. The Joint Commission's eligibility list requires a non-waived CLIA certificate. Waived testing done by an accredited organization, such as point-of-care testing, is still reviewed during tracers. Under CLIA, waived labs must follow the manufacturer's instructions and the certificate of waiver rules in 42 CFR 493.15(e).
Is Sound Lab Consultants affiliated with The Joint Commission?
No. Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or accredited by The Joint Commission, and nothing on this page is an official interpretation of Joint Commission standards.
Sources
- The Joint Commission, Laboratory Services Accreditation Program
- The Joint Commission, Accreditation Process
- The Joint Commission, Laboratory Accreditation Organization Survey Activity Guide, July 2026 (PDF)
- The Joint Commission, National Patient Safety Goals Effective January 2026 for the Laboratory Program (PDF)
- The Joint Commission, National Patient Safety Goals
- The Joint Commission News, July 2026 (lab system survey)
- The Joint Commission, 2026 CAMLAB
- The Joint Commission, 2027 CAMLAB
- The Joint Commission, Accreditation Participation Requirements chapter (public PDF, DMEPOS)
- The Joint Commission, Standards FAQ: Laboratory Reports, Testing Facility Name and Address
- The Joint Commission, Prepublication Standards
- The Joint Commission, Patient Safety Systems (PS) chapter, CAMLAB Update 1, July 2026 (PDF)
- The Joint Commission, Renewed Deeming Authority for Laboratories (Jun 27, 2024)
- The Joint Commission, E-dition Laboratory
- The Joint Commission, Public Standards
- The Joint Commission, First-Timer's Roadmap to Laboratory Accreditation (webinar)
- The Joint Commission, Improving Health Care Equity Standards webinar transcript (Oct 2022, PDF)
- ASCP, New Joint Commission Health Equity Standard Takes Effect January 1, 2023 (Oct 28, 2022)
- MLO, How to prepare for a Joint Commission survey (Dec 22, 2025)
- 42 CFR Part 493 (eCFR)
- 42 CFR 493.3 applicability
- 42 CFR 493.551 deemed status
- 42 CFR 493.803 successful PT participation
- 42 CFR 493.859 ABO and D typing PT
- 42 CFR 493.841 routine chemistry PT
- 42 CFR 493.1105 record retention
- 42 CFR 493.1253 performance specifications
- 42 CFR 493.1254 maintenance and function checks
- 42 CFR 493.1413 moderate complexity technical consultant duties
- 42 CFR 493.1451 high complexity technical supervisor duties
- 42 CFR 493.1780 inspection of accredited labs
- CMS, Clinical Laboratory Improvement Amendments (CLIA)
Related pages
- Joint Commission Laboratory Survey Preparation
- Joint Commission Laboratory Document List
- Joint Commission Laboratory Tracer Methodology
- Joint Commission Laboratory Proficiency Testing Review
- Joint Commission Laboratory Competency Assessment
- Joint Commission Laboratory Infection Prevention and Control
- Joint Commission Laboratory Emergency Management and Environment of Care
- Joint Commission Laboratory Regulatory Review and IQCP
- CLIA competency assessment six elements
- COLA vs CAP vs Joint Commission
- CLIA, CAP, COLA and TJC regulations
Independence note: Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or accredited by The Joint Commission, and nothing here is an official interpretation of Joint Commission standards. Accreditation decisions are made only by The Joint Commission and, where applicable, CMS.