Short answer: A CLIA compliance checklist confirms five things under 42 CFR Part 493: the certificate and CMS-116 match your highest test complexity, staff are qualified with competency checked twice in year one and yearly after, procedures and safety are documented, QC and proficiency testing run on schedule, and specimens, reports and records are handled and kept at least two years.
Last reviewed: November 21, 2026. Primary sources: eCFR, 42 CFR Part 493, Laboratory Requirements, CMS State Operations Manual, Appendix C, Survey Procedures and Interpretive Guidelines for Laboratories (PDF), 42 CFR 493.51, Notification requirements for laboratories issued a certificate of compliance, 42 CFR 493.17, Test categorization.
What this CLIA compliance checklist is built from
CMS does not publish one mandatory lab checklist. Surveyors work from the regulations in 42 CFR Part 493 and the D-tags in State Operations Manual Appendix C. Several official tools turn those rules into self-checks, and this page draws on all of them in its own wording: the CMS CLIA Certification Quick Start Guide, the APHL guide for CLIA internal audits of high complexity testing, the Illinois pre-inspection check list and CDC's Ready? Set? Test! self-assessment for waived sites. Work through the five sections below, then use the 14-point summary as a final pass.
1. Certification and administration
| Check | Evidence to have on hand |
|---|---|
| Certificate type matches the most complex test on the menu and is not expired (2-year term) | Current certificate (download from QCOR) |
| Test menu and specialties match what the certificate lists | Current test list with FDA complexity for each test |
| Form CMS-116 complete, on the revised 04/2026 form for any update from November 9, 2026, and signed by the director or owner | Copy of the filed application (Form CMS-116) |
| Ownership disclosure and Form CMS-209 Laboratory Personnel Report ready for nonwaived surveys, where the State Agency requests them | Completed forms (Illinois list) |
| Changes reported on time: 30 days for owner, name, address or director; 6 months for new specialties (CoC) | State Agency letters (493.51) |
| Fees paid online and invoices tracked in a shared inbox | pay.gov receipts |
| Last survey report reviewed and every past deficiency closed | Prior Form CMS-2567 and plan of correction |
2. Personnel and competency
| Check | Evidence to have on hand |
|---|---|
| Director qualifies for the complexity and visits onsite at least every 6 months, 4+ months apart | CV, license, board or degree records, visit log (493.1445) |
| Every required role filled: TC and CC for moderate; TS, GS and CC for high complexity | Signed delegation of duties (Subpart M) |
| Testing personnel qualify | Diploma, transcript or equivalency for each person |
| Job descriptions and the specific tests each person may perform | Signed test authorization list |
| Training finished before anyone tests patients | Dated training checklists |
| Competency done twice in the first year, then yearly, using all six procedures for each test | Competency forms (CMS competency brochure) |
3. Manuals, procedures and safety
| Check | Evidence to have on hand |
|---|---|
| Procedure manual for every test, approved, signed and dated by the director; retired procedures kept 2 years | Manual and review sign-offs (493.1251) |
| Current package inserts and equipment manuals at the bench; inserts re-checked with each new lot | Insert file with dates |
| Space, ventilation and utilities fit the testing; molecular amplification has one-way workflow when not in a closed system | Floor plan (493.1101) |
| Temperatures logged for reagent storage, refrigerators, freezers and testing rooms | Daily temperature logs with corrective action |
| Safety manual covers PPE, hand hygiene, disinfectant, spills, and biohazard and sharps disposal | Safety manual; bloodborne pathogen plan (OSHA) |
| Reagents stored as labeled, expiration dates checked, lots not mixed | Reagent log |
Want this checklist run against your actual records?
4. Quality control, proficiency testing and quality assessment
| Check | Evidence to have on hand |
|---|---|
| Performance specifications verified for every nonwaived test before patient use | Verification studies with director approval (493.1253) |
| Controls run each day of testing (two levels quantitative; positive and negative qualitative) or an approved IQCP | QC logs, Levey-Jennings charts, IQCP (493.1256) |
| QC reviewed by the director or designee, with corrective action written up | Signed QC review |
| Calibration verification at least every 6 months; same test on two methods or sites compared twice a year | Calibration and comparison records (493.1255; 493.1281) |
| Enrolled in approved PT for every regulated analyte; PT run with the regular workload by staff who routinely test; never referred | PT enrollment, attestations (493.801) |
| PT results reviewed and signed; every unsatisfactory result investigated, including whether patient results were affected | PT review and corrective action |
| Accuracy checked twice a year for tests with no required PT | Split-sample records (493.1236) |
| Quality assessment plan monitors all phases and documents fixes | QA minutes and indicator reviews (493.1200) |
5. Specimen handling, reporting and record retention
| Check | Evidence to have on hand |
|---|---|
| Written or electronic test orders from an authorized person; oral orders backed by written authorization within 30 days | Requisitions (493.1241) |
| Positive patient and specimen identification from collection through reporting (CDC recommends at least two patient-specific identifiers) | Labeling policy (493.1232; CDC) |
| Specimen acceptance and rejection criteria written and followed | Rejection log |
| Reports carry required elements; corrected reports traceable | Sample reports (493.1291) |
| Critical or panic values defined; ordering provider alerted at once; who, when and by whom recorded | Critical value log |
| Records kept for the required periods (most 2 years; pathology 10; cytology slides 5) | Retention policy (493.1105) |
| Record | Minimum time kept |
|---|---|
| Requisitions and test authorizations | 2 years |
| Discontinued test procedures | 2 years after the procedure is retired |
| QC, patient test records, instrument printouts and other analytic records | 2 years |
| Performance specifications you verified or established | As long as the test system is in use, and at least 2 years |
| Proficiency testing and quality assessment records | 2 years |
| Final, preliminary and corrected reports | 2 years after reporting |
| Pathology reports; histopathology slides | 10 years |
| Cytology slides | 5 years |
| Pathology blocks | 2 years |
The 14-point CLIA compliance summary
Use this checklist for moderate and high complexity testing. Waived-only sites can use the short list further down. Each line points to the governing rule in 42 CFR Part 493; surveyors cite the matching D-tags in Appendix C.
| # | Check | Rule | Evidence a surveyor will ask for |
|---|---|---|---|
| 1 | Certificate type matches highest complexity; changes reported in 30 days | 493.51 | Certificate, CMS-116, change letters |
| 2 | Every test categorized; modified tests treated as high complexity | 493.17 | Test menu with FDA categories |
| 3 | Director qualified and onsite every 6 months (4+ months apart) | 493.1445 | CV, license, visit log |
| 4 | All roles filled and qualified for the complexity | Subpart M | Personnel files, transcripts, delegation |
| 5 | Competency: 6 elements, semiannual in year one, then annual | CMS brochure | Competency records per test |
| 6 | Procedure manual approved, signed, dated | 493.1251 | Manual, review signatures |
| 7 | Performance specifications verified before patient testing | 493.1253 | Verification studies, director approval |
| 8 | Calibration verification at least every 6 months | 493.1255 | Calibration records |
| 9 | QC each day of testing (2 levels quantitative; positive and negative qualitative) or approved IQCP | 493.1256 | QC logs, reviews, IQCP |
| 10 | PT enrolled for regulated analytes; no referral; failures investigated | 493.801 | PT results, attestations, corrective action |
| 11 | Accuracy verified twice a year where PT is not required | 493.1236 | Split-sample or alternative assessment records |
| 12 | Test reports complete; critical values and corrections documented | 493.1291 | Sample reports, call logs |
| 13 | Records kept 2 years (pathology reports and histology slides 10, cytology slides 5) | 493.1105 | Archive, retention policy |
| 14 | Facility and safety appropriate | 493.1101 | Temperature logs, safety plans |
Short checklist for waived-only sites
- Current Certificate of Waiver for each testing location, renewed every two years with the fee paid
- Every test and specimen type confirmed as waived in the FDA database
- Current manufacturer instructions followed step by step, including QC, timing and sample type; inserts re-checked for each new lot
- QC results acceptable before patient testing, with fixes recorded when they are not
- Patients identified with at least two identifiers before collection and against the order (CDC good practice)
- Storage and testing room temperatures logged; expired reagents discarded
- Staff trained and documented before testing; PPE, hand hygiene and sharps disposal in place under OSHA rules
- Critical results defined, reported promptly and the call documented
- Changes in owner, name, location or director reported within 30 days
- Ready to allow announced or unannounced inspection. CDC's Ready? Set? Test! checklist covers more good practices. Details: waived testing compliance
How to use the checklist
Work through it at least yearly and before every survey. Assign one owner per line, attach evidence, and re-check closed items a month later. Pair it with a record trace from a CLIA compliance audit and the survey-day steps in how to prepare for a CLIA survey. For accredited labs, run your accreditor's checklist as well; it may be stricter than CLIA.
How Sound Lab Consultants helps
- Works through the checklist with your team and attaches evidence line by line.
- Fixes gaps in personnel files, competency, QC and PT documentation.
- Runs a mock survey once the checklist is closed.
Close every checklist gap before your survey.
Frequently asked questions
What should a CLIA compliance checklist include?
Certificate and test menu, test complexity, director and personnel qualifications, competency, procedure manual, method verification, calibration, QC, PT, reports, quality assessment, record retention and facilities.
Is there an official CLIA checklist?
CMS does not publish a single lab checklist, but surveyors use State Operations Manual Appendix C, which lists every requirement and D-tag. Accreditors publish their own checklists.
How long must CLIA records be kept?
Most records for at least two years; pathology reports and histopathology slides for 10 years; cytology slides for 5 years; blocks for 2 years.
Do waived labs need a CLIA checklist?
A shorter one: current certificate, waived tests only, manufacturer instructions followed, changes reported in 30 days and basic records kept.
How often must CLIA competency be assessed?
Twice in the first year a person tests patient specimens, then at least once a year, covering all six required procedures for every test they perform.
What documents does a CLIA surveyor ask for first?
Typically the certificate and CMS-116, personnel records and competency files, the procedure manual, QC and calibration records, PT results with corrective action, and sample patient reports.
Is there an official CLIA checklist for waived testing?
CDC publishes a voluntary Ready? Set? Test! self-assessment checklist for Certificate of Waiver sites. Some items go beyond the legal minimum but reflect good testing practice.
Sources
- eCFR, 42 CFR Part 493, Laboratory Requirements
- CMS State Operations Manual, Appendix C, Survey Procedures and Interpretive Guidelines for Laboratories (PDF)
- 42 CFR 493.51, Notification requirements for laboratories issued a certificate of compliance
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1445, Laboratory director responsibilities (high complexity)
- 42 CFR Part 493, Subpart M, Personnel for Nonwaived Testing
- CMS, CLIA: Assessing Personnel Competency brochure (PDF)
- 42 CFR 493.1251, Standard: Procedure manual
- 42 CFR 493.1253, Standard: Establishment and verification of performance specifications
- 42 CFR 493.1255, Standard: Calibration and calibration verification procedures
- 42 CFR 493.1256, Standard: Control procedures
- 42 CFR 493.801, Condition: Enrollment and testing of samples
- 42 CFR 493.1236, Standard: Evaluation of proficiency testing performance
- 42 CFR 493.1291, Standard: Test report
- 42 CFR 493.1105, Standard: Retention requirements
- 42 CFR 493.1101, Standard: Facilities
- CMS, CLIA Certification Quick Start Guide, November 2026 (PDF)
- CMS, Form CMS-116 CLIA Application for Certification (04/2026, PDF)
- APHL, Guide for CLIA Internal Audits Related to High Complexity Testing (2022, PDF)
- Illinois Department of Public Health, CLIA Pre-Inspection Check List (PDF)
- CDC, Ready? Set? Test! Self-Assessment Checklist for Good Testing Practices (PDF)
- CMS QCOR, CLIA Laboratory Lookup
- 42 CFR 493.1200, Introduction to the quality system
- 42 CFR 493.1232, Specimen identification and integrity
- 42 CFR 493.1241, Test request
- 42 CFR 493.1281, Comparison of test results
- OSHA, Healthcare standards (bloodborne pathogens)
Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the U.S. Department of Health and Human Services, any state survey agency, the College of American Pathologists, COLA, The Joint Commission, AABB, ACHC, ASHI, A2LA or any other accrediting organization or Medicare contractor. This page is general regulatory information, not legal, billing or medical advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.