CLIA Compliance Checklist for Clinical Laboratories

Short answer: A CLIA compliance checklist confirms five things under 42 CFR Part 493: the certificate and CMS-116 match your highest test complexity, staff are qualified with competency checked twice in year one and yearly after, procedures and safety are documented, QC and proficiency testing run on schedule, and specimens, reports and records are handled and kept at least two years.

Last reviewed: November 21, 2026. Primary sources: eCFR, 42 CFR Part 493, Laboratory Requirements, CMS State Operations Manual, Appendix C, Survey Procedures and Interpretive Guidelines for Laboratories (PDF), 42 CFR 493.51, Notification requirements for laboratories issued a certificate of compliance, 42 CFR 493.17, Test categorization.

Book a free consultation

What this CLIA compliance checklist is built from

CMS does not publish one mandatory lab checklist. Surveyors work from the regulations in 42 CFR Part 493 and the D-tags in State Operations Manual Appendix C. Several official tools turn those rules into self-checks, and this page draws on all of them in its own wording: the CMS CLIA Certification Quick Start Guide, the APHL guide for CLIA internal audits of high complexity testing, the Illinois pre-inspection check list and CDC's Ready? Set? Test! self-assessment for waived sites. Work through the five sections below, then use the 14-point summary as a final pass.

1. Certification and administration

CheckEvidence to have on hand
Certificate type matches the most complex test on the menu and is not expired (2-year term)Current certificate (download from QCOR)
Test menu and specialties match what the certificate listsCurrent test list with FDA complexity for each test
Form CMS-116 complete, on the revised 04/2026 form for any update from November 9, 2026, and signed by the director or ownerCopy of the filed application (Form CMS-116)
Ownership disclosure and Form CMS-209 Laboratory Personnel Report ready for nonwaived surveys, where the State Agency requests themCompleted forms (Illinois list)
Changes reported on time: 30 days for owner, name, address or director; 6 months for new specialties (CoC)State Agency letters (493.51)
Fees paid online and invoices tracked in a shared inboxpay.gov receipts
Last survey report reviewed and every past deficiency closedPrior Form CMS-2567 and plan of correction

2. Personnel and competency

CheckEvidence to have on hand
Director qualifies for the complexity and visits onsite at least every 6 months, 4+ months apartCV, license, board or degree records, visit log (493.1445)
Every required role filled: TC and CC for moderate; TS, GS and CC for high complexitySigned delegation of duties (Subpart M)
Testing personnel qualifyDiploma, transcript or equivalency for each person
Job descriptions and the specific tests each person may performSigned test authorization list
Training finished before anyone tests patientsDated training checklists
Competency done twice in the first year, then yearly, using all six procedures for each testCompetency forms (CMS competency brochure)

3. Manuals, procedures and safety

CheckEvidence to have on hand
Procedure manual for every test, approved, signed and dated by the director; retired procedures kept 2 yearsManual and review sign-offs (493.1251)
Current package inserts and equipment manuals at the bench; inserts re-checked with each new lotInsert file with dates
Space, ventilation and utilities fit the testing; molecular amplification has one-way workflow when not in a closed systemFloor plan (493.1101)
Temperatures logged for reagent storage, refrigerators, freezers and testing roomsDaily temperature logs with corrective action
Safety manual covers PPE, hand hygiene, disinfectant, spills, and biohazard and sharps disposalSafety manual; bloodborne pathogen plan (OSHA)
Reagents stored as labeled, expiration dates checked, lots not mixedReagent log

Want this checklist run against your actual records?

Book a free consultation

4. Quality control, proficiency testing and quality assessment

CheckEvidence to have on hand
Performance specifications verified for every nonwaived test before patient useVerification studies with director approval (493.1253)
Controls run each day of testing (two levels quantitative; positive and negative qualitative) or an approved IQCPQC logs, Levey-Jennings charts, IQCP (493.1256)
QC reviewed by the director or designee, with corrective action written upSigned QC review
Calibration verification at least every 6 months; same test on two methods or sites compared twice a yearCalibration and comparison records (493.1255; 493.1281)
Enrolled in approved PT for every regulated analyte; PT run with the regular workload by staff who routinely test; never referredPT enrollment, attestations (493.801)
PT results reviewed and signed; every unsatisfactory result investigated, including whether patient results were affectedPT review and corrective action
Accuracy checked twice a year for tests with no required PTSplit-sample records (493.1236)
Quality assessment plan monitors all phases and documents fixesQA minutes and indicator reviews (493.1200)

5. Specimen handling, reporting and record retention

CheckEvidence to have on hand
Written or electronic test orders from an authorized person; oral orders backed by written authorization within 30 daysRequisitions (493.1241)
Positive patient and specimen identification from collection through reporting (CDC recommends at least two patient-specific identifiers)Labeling policy (493.1232; CDC)
Specimen acceptance and rejection criteria written and followedRejection log
Reports carry required elements; corrected reports traceableSample reports (493.1291)
Critical or panic values defined; ordering provider alerted at once; who, when and by whom recordedCritical value log
Records kept for the required periods (most 2 years; pathology 10; cytology slides 5)Retention policy (493.1105)
CLIA record retention minimums, 42 CFR 493.1105
RecordMinimum time kept
Requisitions and test authorizations2 years
Discontinued test procedures2 years after the procedure is retired
QC, patient test records, instrument printouts and other analytic records2 years
Performance specifications you verified or establishedAs long as the test system is in use, and at least 2 years
Proficiency testing and quality assessment records2 years
Final, preliminary and corrected reports2 years after reporting
Pathology reports; histopathology slides10 years
Cytology slides5 years
Pathology blocks2 years

The 14-point CLIA compliance summary

Use this checklist for moderate and high complexity testing. Waived-only sites can use the short list further down. Each line points to the governing rule in 42 CFR Part 493; surveyors cite the matching D-tags in Appendix C.

#CheckRuleEvidence a surveyor will ask for
1Certificate type matches highest complexity; changes reported in 30 days493.51Certificate, CMS-116, change letters
2Every test categorized; modified tests treated as high complexity493.17Test menu with FDA categories
3Director qualified and onsite every 6 months (4+ months apart)493.1445CV, license, visit log
4All roles filled and qualified for the complexitySubpart MPersonnel files, transcripts, delegation
5Competency: 6 elements, semiannual in year one, then annualCMS brochureCompetency records per test
6Procedure manual approved, signed, dated493.1251Manual, review signatures
7Performance specifications verified before patient testing493.1253Verification studies, director approval
8Calibration verification at least every 6 months493.1255Calibration records
9QC each day of testing (2 levels quantitative; positive and negative qualitative) or approved IQCP493.1256QC logs, reviews, IQCP
10PT enrolled for regulated analytes; no referral; failures investigated493.801PT results, attestations, corrective action
11Accuracy verified twice a year where PT is not required493.1236Split-sample or alternative assessment records
12Test reports complete; critical values and corrections documented493.1291Sample reports, call logs
13Records kept 2 years (pathology reports and histology slides 10, cytology slides 5)493.1105Archive, retention policy
14Facility and safety appropriate493.1101Temperature logs, safety plans

Short checklist for waived-only sites

  • Current Certificate of Waiver for each testing location, renewed every two years with the fee paid
  • Every test and specimen type confirmed as waived in the FDA database
  • Current manufacturer instructions followed step by step, including QC, timing and sample type; inserts re-checked for each new lot
  • QC results acceptable before patient testing, with fixes recorded when they are not
  • Patients identified with at least two identifiers before collection and against the order (CDC good practice)
  • Storage and testing room temperatures logged; expired reagents discarded
  • Staff trained and documented before testing; PPE, hand hygiene and sharps disposal in place under OSHA rules
  • Critical results defined, reported promptly and the call documented
  • Changes in owner, name, location or director reported within 30 days
  • Ready to allow announced or unannounced inspection. CDC's Ready? Set? Test! checklist covers more good practices. Details: waived testing compliance

How to use the checklist

Work through it at least yearly and before every survey. Assign one owner per line, attach evidence, and re-check closed items a month later. Pair it with a record trace from a CLIA compliance audit and the survey-day steps in how to prepare for a CLIA survey. For accredited labs, run your accreditor's checklist as well; it may be stricter than CLIA.

How Sound Lab Consultants helps

  • Works through the checklist with your team and attaches evidence line by line.
  • Fixes gaps in personnel files, competency, QC and PT documentation.
  • Runs a mock survey once the checklist is closed.

Close every checklist gap before your survey.

Book a free consultation

Frequently asked questions

What should a CLIA compliance checklist include?

Certificate and test menu, test complexity, director and personnel qualifications, competency, procedure manual, method verification, calibration, QC, PT, reports, quality assessment, record retention and facilities.

Is there an official CLIA checklist?

CMS does not publish a single lab checklist, but surveyors use State Operations Manual Appendix C, which lists every requirement and D-tag. Accreditors publish their own checklists.

How long must CLIA records be kept?

Most records for at least two years; pathology reports and histopathology slides for 10 years; cytology slides for 5 years; blocks for 2 years.

Do waived labs need a CLIA checklist?

A shorter one: current certificate, waived tests only, manufacturer instructions followed, changes reported in 30 days and basic records kept.

How often must CLIA competency be assessed?

Twice in the first year a person tests patient specimens, then at least once a year, covering all six required procedures for every test they perform.

What documents does a CLIA surveyor ask for first?

Typically the certificate and CMS-116, personnel records and competency files, the procedure manual, QC and calibration records, PT results with corrective action, and sample patient reports.

Is there an official CLIA checklist for waived testing?

CDC publishes a voluntary Ready? Set? Test! self-assessment checklist for Certificate of Waiver sites. Some items go beyond the legal minimum but reflect good testing practice.

Sources

Sound Lab Consultants is an independent consulting firm. It is not affiliated with, endorsed by or acting for CMS, CDC, FDA, the U.S. Department of Health and Human Services, any state survey agency, the College of American Pathologists, COLA, The Joint Commission, AABB, ACHC, ASHI, A2LA or any other accrediting organization or Medicare contractor. This page is general regulatory information, not legal, billing or medical advice. Confirm current requirements with the agency or accreditor that oversees your laboratory. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.