Lab Survey Preparation

Lab survey preparation is the work a clinical laboratory does before a CLIA, CAP, COLA or ACHC survey to show it meets 42 CFR Part 493 and its accreditor's standards. The core steps are: review the QC plan, risk assessment and QA logs; run a mock inspection; review the procedure manual; check calibration and maintenance records; review PT and corrective action; and prepare for the surveyor's facility tour and safety record review.

Last reviewed: Oct 31, 2026

Book a free consultation

Clipboard with a partly checked list beside a pen on a light background

Why survey preparation matters

Labs performing non-waived testing under a Certificate of Compliance are inspected by CMS or its state agent under 42 CFR Part 493 Subpart Q, and 493.1777 allows subsequent inspections on a biennial basis. Labs with a Certificate of Accreditation are inspected by their accrediting organization, such as CAP, COLA or ACHC. 42 CFR 493.1773 requires the lab to permit inspection, make records available and have staff available to answer questions. Deficiencies are written up and require a plan of correction, so a survey-prepared lab is one that can show its records on request rather than recreate them.

Core lab survey preparation steps

  1. QC Plan, Risk Assessment and QA Logs
  2. Mock Inspections
  3. Procedure Manual Review
  4. Calibration and Maintenance Records
  5. PT Review and Corrective Action
  6. Surveyor Facility Tour and Safety Records

QC Plan, Risk Assessment and QA Logs

Surveyors start with the quality system. 42 CFR 493.1256 requires control procedures for every test system, run at the frequency the regulation or your Individualized Quality Control Plan (IQCP) sets. CMS describes IQCP, which includes a risk assessment, a quality control plan and quality assessment, in the interpretive guidelines of SOM Appendix C. 42 CFR 493.1239 and 493.1289 require ongoing quality assessment of general laboratory systems and the analytic phase, with problems found, corrected and documented.

  • QC plan in place for each analyzer and kit, with frequency, levels and acceptance limits.
  • Risk assessment covering specimen, test system, reagent, environment and testing personnel when IQCP is used.
  • QC logs reviewed and signed at the interval your policy states; out-of-range QC tied to a corrective action.
  • Quality assessment (QA) reviews, meeting notes and indicators showing problems were followed to closure.
Rack of specimen tubes with colored caps on a laboratory bench

Mock Inspections

A mock inspection is an internal or outside walk-through that follows the same tasks a surveyor uses. SOM Appendix C lays out the survey procedures, and CAP publishes its accreditation checklists, so a mock inspection can use the same requirement language the real inspection will. Treat each finding as you would a real citation: record it, assign an owner and a due date, and keep evidence of correction.

  • Pull a sample of patient reports and trace each back to the specimen, QC, calibration, reagent lot, instrument and testing person.
  • Interview testing personnel about the procedures they follow.
  • Compare the test menu on your CLIA certificate with what is actually run.

Procedure Manual Review

42 CFR 493.1251 requires a written procedure manual for all tests, available to staff, that covers items such as specimen requirements, step-by-step performance, calibration and QC, reportable ranges, reference intervals, critical values and corrective action. The laboratory director must approve, sign and date new procedures and changes before use, and discontinued procedures are kept as 493.1105 requires.

  • Every test on the menu has a current procedure that matches the package insert or validated method.
  • Director signature and date on each new or changed procedure.
  • Retired procedures archived with the dates they were in use.

Calibration and Maintenance Records

42 CFR 493.1254 requires maintenance and function checks at least as often as the manufacturer specifies, documented. 42 CFR 493.1255 requires calibration and calibration verification at the frequency the regulation and manufacturer require, including at least every six months for calibration verification where it applies and when listed events occur such as reagent changes or major maintenance.

  • Maintenance logs complete for each instrument, with daily, weekly and monthly tasks initialed.
  • Calibration and calibration verification records with dates, materials and acceptance criteria.
  • Temperature, humidity and water quality logs where the manufacturer requires them.

PT Review and Corrective Action

Under 42 CFR 493.1236 the lab must review and evaluate the results of each proficiency testing (PT) event and take and document corrective action for unacceptable results. 42 CFR 493.1282 requires corrective action policies and procedures whenever errors are detected, and 493.1236(c) requires verifying the accuracy of tests not covered by PT at least twice a year.

  • PT enrollment matches every regulated analyte on the menu.
  • Each PT event reviewed, signed and dated, including passing events.
  • Unacceptable results investigated with root cause and corrective action recorded.
  • Twice-yearly accuracy checks for non-PT tests.

Surveyor Facility Tour and Safety Records

Surveyors walk through the testing areas. 42 CFR 493.1101 requires space, ventilation, utilities and conditions that support quality testing and safety, with contamination prevented and storage appropriate. Labs also keep the safety records that OSHA requires, such as the bloodborne pathogens exposure control plan (29 CFR 1910.1030) and the chemical hygiene plan for laboratories (29 CFR 1910.1450).

  • Benches clean and uncluttered; reagents labeled and in date; storage temperatures logged.
  • Exposure control plan, chemical hygiene plan, safety data sheets and training records current.
  • Eyewash, spill kits and personal protective equipment available and checked.

CLIA, CAP, COLA and ACHC survey preparation

SurveyStandard usedWhere to read it
CLIA (state agency)42 CFR Part 493 and the SOM Appendix C interpretive guidelineseCFR; CMS SOM Appendix C
CAPCAP accreditation checklistsCAP Laboratory Accreditation Program
COLACOLA accreditation criteriaCOLA
ACHCACHC clinical laboratory standardsACHC Clinical Laboratory

All four apply CLIA as the floor. See How to Prepare for a CLIA Survey, CAP Inspection Preparation, COLA Survey Preparation and ACHC Laboratory Survey Preparation.

How SLC helps

Sound Lab Consultants works with clinical labs on survey preparation: QC and IQCP documents, mock inspections against Appendix C and accreditor checklists, procedure manual reviews, record gap checks and plan of correction support. Start with the Survey-Prepared Lab Checklist or see Laboratory Survey Management.

Frequently asked questions

What is lab survey preparation?

It is the review of a lab's quality system, procedures, records and facility before a CLIA, CAP, COLA or ACHC survey so the lab can show compliance with 42 CFR Part 493 and its accreditor's standards. Source: eCFR 42 CFR Part 493 - Laboratory Requirements, CMS State Operations Manual Appendix C - Survey Procedures and Interpretive Guidelines for Laboratories (PDF).

How often are CLIA labs surveyed?

42 CFR 493.1777 allows CMS or its agent to conduct subsequent inspections on a biennial basis or at another frequency CMS sets. Accredited labs follow their accreditor's inspection cycle. Source: eCFR 42 CFR § 493.1777.

What records must a lab show a surveyor?

42 CFR 493.1773 requires the lab to make records and information available on request, including QC, PT, maintenance, calibration, personnel and test reports. Source: eCFR 42 CFR § 493.1773.

What happens if deficiencies are found?

Deficiencies are cited and the lab submits a plan of correction. Condition-level deficiencies can lead to principal or alternative sanctions under Subpart R. Source: CMS State Operations Manual Appendix C - Survey Procedures and Interpretive Guidelines for Laboratories (PDF), eCFR 42 CFR § 493.1816.

Sources

Related pages

Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.

Book a free consultation

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

📧 sales@soundlabconsult.com

📍 Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

© 2026 Sound Lab Consultants. All rights reserved.