QC Program Design: Westgard Rules, Six Sigma and SPC

What it is: a laboratory QC program is the written set of decisions that controls each test system: which control materials and levels are run, how often, what the target and acceptable range are for each lot, which statistical rules accept or reject a run, who reviews the results, and what happens when a run fails. Statistical process control (SPC), Westgard multirules and Sigma metrics are the tools used to make those decisions defensible.

Sound Lab Consultants (SLC) helps clinical labs design or rebuild QC programs that meet 42 CFR 493.1256 and the lab's accreditor, and that staff can follow on a busy shift. We work with whatever QC software or paper system you already use. The lab director makes the final decisions; we do the analysis and the writing.

Gloved hands attaching a pipette to a pipette aid in a clinical laboratory
Photo: Diane A. Reid, public domain, via Wikimedia Commons

The building blocks of a QC program

  • Control materials and levels. For quantitative tests CLIA requires at least two control materials of different concentrations at least once each day patient specimens are assayed, unless an alternative such as an IQCP applies or the manufacturer requires more.
  • Targets and SDs. Each control lot needs a mean and a standard deviation. Manufacturer ranges are a starting point; most labs replace them with their own data once enough results exist.
  • Rules. The statistical rules that decide whether a run is accepted.
  • Frequency and placement. How many control events per day or per run, and when, based on the method's stability and the risk of reporting a bad result.
  • Review. Who looks at QC before results go out, and who reviews trends weekly or monthly.
  • Response. What the bench does after a rejection, how patient results are evaluated, and how the event is documented.

What 42 CFR 493.1256 says, in plain language

42 CFR 493.1256 is the core QC section of CLIA. Its main points for program design:

  • Control procedures must detect immediate errors caused by test system failure, adverse environmental conditions and operator performance, and must monitor accuracy and precision over time.
  • The lab sets the number, type and frequency of control testing, within the minimums in paragraph (d), considering the method's performance specifications.
  • For each control material the lab must establish or verify acceptability criteria; for quantitative tests, a mean and SD must be defined and available for each lot.
  • Control results must meet the lab's criteria before patient results are reported.
  • All control procedures must be documented.

Related sections shape the program too: 42 CFR 493.1253 on verifying performance specifications, 42 CFR 493.1255 on calibration and calibration verification, and 42 CFR 493.1282 on corrective action.

SPC and Levey-Jennings charts

Statistical process control treats each test as a process with expected variation. A Levey-Jennings chart plots control results over time against the mean and lines at plus and minus 1, 2 and 3 SD. Patterns on the chart show different problems: a sudden jump suggests a calibration or reagent change, a slow drift suggests deterioration, and wide scatter suggests random error.

Charts only work if the mean and SD are right. If limits come from a wide manufacturer range, real shifts can stay inside the lines. We help labs set lab-specific targets from an adequate number of points over enough days and runs, document how they were set, and review them at each lot change. Our CLIA quality control requirements page covers the regulation behind this.

Westgard multirules: choosing, not defaulting

Westgard multirules combine several rules so the lab catches real errors without rejecting too many good runs. Common rules include:

  • 13s: one result beyond 3 SD. Detects large random error.
  • 22s: two consecutive results beyond 2 SD on the same side. Detects systematic error.
  • R4s: a range of more than 4 SD between two results in the same run. Detects random error.
  • 41s: four consecutive results beyond 1 SD on the same side. Detects smaller shifts.
  • 10x: ten consecutive results on one side of the mean. Detects shifts and trends.
  • 12s: one result beyond 2 SD, often used as a warning that triggers inspection of the other rules rather than as a rejection rule.

Using every rule on every test is not automatically safer. It raises false rejections, and repeated false rejections teach staff to rerun controls until they pass. The aim is to pick, for each test, the simplest set of rules that detects errors that matter. More on this in our Westgard rules under CLIA guide.

Six Sigma metrics: linking rules to a quality goal

Sigma metrics connect a method's performance to the quality the lab needs. A common form is:

Sigma = (TEa% โˆ’ |Bias%|) / CV%

  • TEa is the allowable total error, chosen by the lab director from a source the lab documents, such as CLIA proficiency testing criteria, a professional or clinical goal, or the lab's own policy.
  • Bias usually comes from proficiency testing, comparison with other labs or a reference method.
  • CV is the lab's own imprecision from QC data at the same level.

High-Sigma methods can often use simple rules and fewer controls. Low-Sigma methods need more rules, more control events or method improvement. Sigma is a design aid, not a CLIA requirement, and the result is only as good as the TEa source and the data behind bias and CV. We document every input so the reasoning can be reviewed later.

CLIA proficiency testing acceptance limits are set in Subpart I of 42 CFR 493. Many labs use them as one TEa source, but they are a minimum performance standard, not a clinical goal, so the director should decide whether a tighter goal is needed.

Lot changes and crossover

Many QC problems start at a control lot change. If a new lot goes into use with targets copied from the insert, or with a mean and SD set from a handful of points, the rules that looked well-chosen on paper stop working. A sound program writes the lot change down as a procedure:

  • Run the new lot alongside the current lot for a defined period before the current lot runs out.
  • Collect enough results across days, runs and, where relevant, calibrations to set a provisional mean and SD.
  • Have a qualified person review and approve the new targets before the new lot is used alone.
  • Recalculate the mean and SD once more data is available, and document any change.
  • Keep the crossover data and the approval with the QC records for the retention period.

Several instruments running the same test

Labs with more than one analyzer or method for the same test have an extra duty. Under 42 CFR 493.1281, the lab must have a system to evaluate and define the relationship between test results using different methodologies, instruments or testing sites, twice a year. QC design should account for this: whether each instrument has its own targets, how differences between instruments are tracked, and how the twice-yearly comparison is documented and reviewed. Shared targets across instruments can hide a bias on one of them, so we usually recommend instrument-specific means and SDs, with comparison records kept alongside the QC program.

Review, response and documentation

A well-designed rule set fails if the response is weak. We write the response steps into the procedure:

  • Stop and hold patient results when a rejection rule is broken.
  • Look for the cause: controls, reagents, calibration, instrument, environment, operator.
  • Fix and verify with acceptable QC.
  • Evaluate patient results back to the last acceptable run, as 42 CFR 493.1282 requires, and decide whether any need correction.
  • Document cause, action, patient evaluation and reviewer.
  • Review trends and repeat failures monthly, and feed them into quality assessment.

Our QC corrective action documentation guide shows what a complete record looks like.

Where IQCP fits

An Individualized Quality Control Plan lets a lab use a QC frequency other than the CLIA default for eligible tests, based on a documented risk assessment, a QC plan and ongoing quality assessment. CMS explains the process in its IQCP workbook. QC program design and IQCP overlap: the same performance data and Sigma analysis that inform rule choice can support the risk assessment. See our IQCP consulting page.

How a QC design project runs

  • Free consultation to understand your menu, methods, accreditor and concerns.
  • Data review. We review recent QC data, current rules, limits, lot change records, proficiency testing results and corrective actions.
  • Analysis. We calculate imprecision and, where you want it, Sigma metrics per test and level, and flag tests where limits or rules do not fit performance.
  • Proposal. A per-test table of controls, frequency, targets, SDs and rules, with reasons, for director review.
  • Procedures and training. Revised QC and corrective action procedures, and short training for staff.
  • Follow-up. A check of a later month of QC to confirm the program works as intended.

Frequently asked questions

What is a lab QC program?

The written decisions that control each test: control materials and levels, frequency, targets and SDs, acceptance rules, review steps and corrective action, as required by 42 CFR 493.1256 and 493.1282.

Does CLIA require Westgard rules?

No. CLIA requires the lab to set acceptability criteria and control procedures that detect errors. Westgard multirules are a common, well-documented way to do that, but CLIA does not name them.

Is Six Sigma required for lab QC?

No. Sigma metrics are an optional design tool that links rule choice to a quality goal. They are useful for choosing rules and frequency, and for supporting an IQCP risk assessment.

Should we use manufacturer ranges or our own mean and SD?

Manufacturer ranges are a starting point. Most labs replace them with lab-established means and SDs once they have enough data, because insert ranges are often wide and can hide shifts.

Is 1-2s a rejection rule?

In most multirule designs it is a warning that triggers a check of the other rules. Using it as a rejection rule with two controls causes frequent false rejections.

What TEa should we use?

The lab director chooses and documents it. Common sources include CLIA proficiency testing criteria in 42 CFR 493 Subpart I, professional or clinical goals, or lab policy.

Can SLC work with our current QC software?

Yes. We are vendor-neutral and work with your existing software or paper records.

Sources: 42 CFR 493.1256 Control procedures; 42 CFR 493.1282 Corrective actions; 42 CFR 493.1253 Establishment and verification of performance specifications; 42 CFR 493.1255 Calibration and calibration verification; 42 CFR 493.1281 Comparison of test results; 42 CFR 493 Subpart I Proficiency testing; CMS IQCP workbook (PDF); CMS CLIA program.

Related SLC pages: QC software selection and implementation, LIS and LIMS QC module setup and validation, Westgard rules and CLIA QC, CLIA quality control requirements, IQCP consulting, QC corrective action documentation.

Want a second look at your QC rules, limits and frequency before your next survey?

Book a free consultation

Sound Lab Consultants is an independent consulting firm. We do not sell software, and this page does not compare or recommend any vendor. This page is general information, not legal advice.

Sound Lab Consultants ยท PO Box 2135, Chandler, AZ 85244

Sound Lab Consultants Logo

Clinical Lab Growth & Compliance Experts helping labs launch, scale, and pass inspections with confidence.

๐Ÿ“ง sales@soundlabconsult.com

๐Ÿ“ Headquarters: 25 Arizona Pl, Chandler, AZ 85225
Serving laboratories nationwide in all 50 states

Services

  • CLIA, COLA & CAP Consulting
  • Lab Startup & Buildout
  • Inspection Preparation
  • Lab Billing & Revenue Cycle
  • Staff Training & Compliance

Get Started

Ready to build or scale your lab?

Schedule a Consultation

Trusted by labs nationwide

ยฉ 2026 Sound Lab Consultants. All rights reserved.